Evaluation of the Innovance D-DIMER analytical performance

被引:28
|
作者
Herak, Desiree Coen
Milos, Marija
Zadro, Renata
机构
[1] Univ Zagreb, Sch Med, Zagreb, Croatia
[2] Univ Zagreb, Clin Hosp Ctr Zagreb, Inst Clin, Lab Diag, Zagreb, Croatia
关键词
coagulation; D-dimer; evaluation; particle-enhanced immunoturbidimetric assay; VENOUS THROMBOEMBOLISM; ASSAYS; EXCLUSION; AGREEMENT; DIAGNOSIS;
D O I
10.1515/CCLM.2009.219
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Widespread use of D-dimer in recent years has led to the development of a number of new fully automated quantitative D-dimer assays. Methods: We evaluated the analytical performance of the particle-enhanced immunoturbidimetric assay Innovance D-DIMER (Siemens Medical Solutions) on the Behring Coagulation System (BCS) analyzer. Results: Within-run coefficients of variation (CVs) for samples with low, borderline, slightly, and extremely increased D-dimer concentrations were 2.1%-5.5%, whereas between-run CVs for control samples with low and extremely increased D-dimer were 5.5%-8.4%. The assay exhibited good linearity in the working range between 0.17 mg/L and 5.45 mg/L fibrinogen equivalent units (FEU), with the lower limit of detection of 0.099 mg/L FEU. The upper reference value determined in 40 plasma samples from healthy volunteers was 0.495 mg/L FEU. The results obtained in 457 fresh plasma samples were compared with results obtained with VIDAS D-Dimer Exclusion. Passing and Bablok regression analysis demonstrated highly significant correlation (y=1.370x-0.108, r=0.952, p<0.001). Bland and Altman difference plots demonstrated slightly higher results obtained with Innovance D-DIMER that was more pronounced with increasing values. Very good agreement between both assays was observed (kappa=0.860; 95% confidence interval (CI), 0.811-0.908). Conclusions: This study demonstrates that Innovance D-DIMER fulfills all analytical requirements for daily routine use. Clin Chem Lab Med 2009;47:945-51.
引用
收藏
页码:945 / 951
页数:7
相关论文
共 50 条
  • [1] Evaluation of the innovance D-dimer analytical performance
    Herak, Coen D.
    Milos, M.
    Zadro, R.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2009, 7 : 392 - 393
  • [2] Analytical comparison of AxSYM, HemosIL DD HS and Innovance D-dimer immunoassays with the Vidas D-dimer
    Salvagno, G. L.
    Lippi, G.
    Manzato, F.
    Giavarina, D.
    Montagnana, M.
    Poli, G.
    Guidi, G. C.
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2009, 31 (04) : 475 - 477
  • [3] INNOVANCE D-DIMER IS SURPERIOR TO PLUS D-DIMER TO EXCLUDE FALSE NEGATIVE
    Guo Ye
    Shou Wl
    Wu Wei
    Hao Xp
    Cui Wei
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2013, 35 : 80 - 80
  • [4] Performance of two relatively new quantitative D-dimer assays (Innovance D-dimer and AxSYM D-dimer) for the exclusion of deep vein thrombosis
    Elf, J. L.
    Strandberg, K.
    Svensson, P. J.
    [J]. THROMBOSIS RESEARCH, 2009, 124 (06) : 701 - 705
  • [5] Evaluation of the innovance® D-dimer assay on the sysmex® ca-7000 system
    Park, S.
    Chi, H.
    Chun, S.
    Jang, S.
    Park, C.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2009, 7 : 1103 - 1103
  • [6] Evaluation of Performance Including Influence by Interfering Substances of the Innovance D-Dimer Assay on the Sysmex Coagulation Analyzer
    Park, Seo-Jin
    Chi, Hyun-Sook
    Chun, Soh Hyun
    Jang, Seongsoo
    Park, Chan-Jeoung
    [J]. ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2011, 41 (01): : 20 - 24
  • [7] Long- and short-term in vitro D-dimer stability measured with INNOVANCE D-Dimer
    Boehm-Weigert, Martina
    Wissel, Thomas
    Muth, Heidrun
    Kemkes-Matthes, Bettina
    Peetz, Dirk
    [J]. THROMBOSIS AND HAEMOSTASIS, 2010, 103 (02) : 461 - 465
  • [8] Evaluation of the Innovance D-Dimer Assay for the Diagnosis of Disseminated Intravascular Coagulopathy in Different Clinical Settings
    Khalafallah, Alhossain
    Jarvis, Chris
    Morse, Michael
    Albarzan, Abdul-Majeed
    Stewart, Phoebe
    Bates, Gerald
    Hayes, Robert
    Robertson, Iain
    Seaton, David
    Brain, Terry
    [J]. CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2014, 20 (01) : 91 - 97
  • [9] Innovance D-Dimer实验室检测应用体会
    禹华玮
    梁冰
    王金乐
    胡峰
    杨银荣
    [J]. 国际检验医学杂志, 2012, 33 (21) : 2644 - 2645
  • [10] Analytical and Clinical Evaluation of Four D-dimer Methods
    Padurean, A. M.
    Lund, B.
    Gifford, G.
    Hanson, G.
    Kirchick, H. J.
    [J]. ANNALS OF EMERGENCY MEDICINE, 2008, 52 (04) : S152 - S153