Design and performance of a low noise, 128-channel ASIC preamplifier for readout of active matrix flat-panel imaging arrays

被引:8
|
作者
Maolinbay, M
Zimmerman, T
Yarema, RJ
Antonuk, LE
El-Mohri, Y
Yeakey, M
机构
[1] Univ Michigan, Med Ctr, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
[2] Fermilab Natl Accelerator Lab, Batavia, IL 60510 USA
基金
美国国家卫生研究院;
关键词
low noise charge sensitive preamplifier; active matrix flat-panel imager; amorphous silicon; thin-film transistors;
D O I
10.1016/S0168-9002(01)02129-5
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Design architecture and performance measurements of a low noise, 128-channel application-specific-integrated-circuit (ASIC) preamplifier are reported. The ASIC was designed for readout of active matrix flat-panel imager (AMFPI) arrays. Such arrays, which presently can be made as large as 41 cm x 41 cm and with pixel-to-pixel pitches down to similar to70 mum, require large numbers of low noise, high density, custom integrated readout circuits. The design of this new chip is specifically tailored for research and development of active matrix flat-panel arrays for various medical imaging applications. The design architecture includes the following features: (1) Programmable signal gain which allows acquisition of a wide range of signal sizes from various array designs so as to optimize the signal-to-noise ratio; (2) Correlated double sampling (CDS) which significantly reduces certain noise components; (3) Pipelined readout (simultaneously sampling and multiplexing signals) which reduces image acquisition time; (4) Programmable bandwidth controls which balance noise and acquisition speed; and (5) Two selectable modes of output multiplexing (64:1, 16:1) for slow or fast readout. In this paper, detailed measurements of various performance parameters are presented. These measurements include noise characteristics, the relationship between bandwidth and noise, signal response linearity, channel-to-channel and pipeline cross-talk, signal gain and gain variation across channels, and the effect of sampling methods on noise. These characterizations indicate that the performance of the ASIC has achieved the original design goals. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:661 / 675
页数:15
相关论文
共 49 条
  • [1] A 128-Channel CMOS Charge Readout ASIC for Flat-Panel X-Ray Detectors
    Adachi, Susumu
    Nishimura, Akihiro
    Yoshimlita, Toshinori
    Tanabe, Koichi
    Okamura, Shoichi
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (06) : 3673 - 3683
  • [2] High performance, low-noise, 128-channel readout integrated circuit for flat panel x-ray detector systems
    Beuville, E
    Belding, M
    Costello, A
    Hansen, R
    Petronio, S
    [J]. MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 714 - 725
  • [3] VERITAS: A 128-Channel ASIC for the Readout of pnCCDs and DEPFET Arrays for X-Ray Imaging, Spectroscopy and XFEL Applications
    Porro, Matteo
    Bianchi, Davide
    De Vita, Giulio
    Hartmann, Robert
    Hauser, Guenter
    Herrmann, Sven
    Strueder, Lothar
    Wassatsch, Andreas
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (01) : 446 - 455
  • [4] Additive noise properties of active matrix flat-panel imagers
    Maolinbay, M
    El-Mohri, Y
    Antonuk, LE
    Jee, KW
    Nassif, S
    Rong, X
    Zhao, Q
    [J]. MEDICAL PHYSICS, 2000, 27 (08) : 1841 - 1854
  • [5] Enhancement of signal performance through innovative pixel design for indirect detection active matrix flat-panel arrays
    Antonuk, L.
    Zhao, Q.
    El-Mohri, Y.
    Du, H.
    Behravan, M.
    Wang, Y.
    Weisfield, R.
    Yao, B.
    Street, R.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2525 - 2526
  • [6] Flat-panel digital radiology with amorphous selenium and active-matrix readout
    Rowlands, JA
    Zhao, W
    Blevis, IM
    Waechter, DF
    Huang, ZS
    [J]. RADIOGRAPHICS, 1997, 17 (03) : 753 - 760
  • [7] A high performance, low-noise 128-channel readout integrated circuit for instrumentation and X-ray applications
    Beuville, E
    Belding, M
    Costello, A
    Hansen, R
    Petronio, S
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 142 - 146
  • [8] Active matrix flat-panel imagers (AMFPIs) for electronic portal imaging
    Antonuk, LE
    El-Mohri, Y
    Jee, KW
    [J]. IMAGING IN RADIATION THERAPY, 1998, (24): : 371 - 392
  • [9] Design of a Monolithic Multi-Channel Front-End Readout ASIC for LYSO/SiPM-based Small-Animal Flat-Panel PET Imaging
    Gao, W.
    Gao, D.
    Wei, T.
    Zeng, H.
    Duan, Y.
    Lu, S.
    Shen, L.
    Xu, W.
    Xie, Q.
    Hu, Y.
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2447 - 2450
  • [10] A theoretical investigation of advanced design active matrix flat-panel imagers for mammography
    Jee, KW
    Antonuk, LE
    El-Mohri, Y
    Cionca, C
    Maolinbay, M
    Nassif, S
    Zhao, QH
    [J]. MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 : 250 - 256