Processor quality control in laser imaging systems

被引:8
|
作者
Bogucki, TM
Murphy, WR
Baker, CW
Piazza, SS
Haus, AG
机构
[1] Eastman Kodak Company, Rochester
关键词
laser camera; quality control;
D O I
10.1118/1.597940
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Sensitometric techniques for performing processor quality control in laser imaging systems are analyzed in this study. The sensitivity of conventional x-ray films using simulated screen-light sensitometry is compared with helium-neon (HeNe) laser film exposed with a simulated red-light sensitometer, a standalone (reference) laser sensitometer, an experimental (unstable) laser sensitometer, and laser printers. Infrared (IR) laser film exposed with an IR laser diode and a simulated IR sensitometer are also evaluated. It is demonstrated that laser-generated step tablets provide an easy and reliable method of performing laser film processor quality control. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:581 / 584
页数:4
相关论文
共 50 条
  • [1] An automated system for film processor quality control
    Shearer, RDC
    Shearer, DR
    JOURNAL OF THE AMERICAN MEDICAL INFORMATICS ASSOCIATION, 1997, : 1005 - 1005
  • [2] Guidelines for Quality Control Testing of Molecular Breast Imaging Systems
    Nardinger, Sara M.
    Tran, Thuy D.
    Swanson, Tiffinee N.
    Ellingson, Lacey R.
    Solberg, Courtney M.
    O'Connor, Michael K.
    Hruska, Carrie B.
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2018, 46 (04) : 349 - 354
  • [3] TL dosimetry for quality control of CR mammography imaging systems
    Gaona, E.
    Nieto, J. A.
    Gongora, J. A. I. D.
    Arreola, M.
    Enriquez, J. G. F.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2007, 162 (10-12): : 759 - 763
  • [4] Phantom for quality control of electronic-portal-imaging-systems
    Röser, A
    Divoux, S
    Rachinskiy, I
    Schilcher, RB
    STRAHLENTHERAPIE UND ONKOLOGIE, 2005, 181 : 112 - 113
  • [5] Quality control testing for dedicated molecular breast imaging systems
    Nardinger, Sara
    Thuy Tran
    Swanson, Tiffinee
    Ellingson, Lacey
    Solberg, Courtney
    O'Connor, Michael
    Hruska, Carrie
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [6] FDOCT imaging processor for portable OCT systems with high imaging rate
    Tang, Song-Nien
    Hsiang, Chih-Yu
    Huang, Sheng-Jie
    Chen, Wan-Wei
    IEICE ELECTRONICS EXPRESS, 2018, 15 (03):
  • [7] A ROVING MONITORING PROCESSOR FOR DETECTION OF CONTROL FLOW ERRORS IN MULTIPLE PROCESSOR SYSTEMS
    SHEN, JP
    TOMAS, SP
    MICROPROCESSING AND MICROPROGRAMMING, 1987, 20 (4-5): : 249 - 269
  • [8] QUALITY-CONTROL AND YOUR SILVER MICROFILM PROCESSOR
    BAILER, R
    INTERNATIONAL JOURNAL OF MICROGRAPHICS & OPTICAL TECHNOLOGY, 1995, 13 (03): : 151 - 153
  • [9] DIGITAL QUALITY CONTROL PROCESSOR AND BATCH COUNTER.
    Jawalekar, S.R..
    Gulanikar, A.C.
    Journal of the Institution of Engineers (India), Part ET: Electronics & Telecommunication Engineering Division, 1977, 58 (pt ET 1): : 16 - 17
  • [10] QUALITY-CONTROL OF WHOLE-BODY IMAGING-SYSTEMS
    BLOKLAND, JAK
    CAMPS, JAJ
    PAUWELS, EKJ
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1992, 19 (08): : 638 - 638