A Sensitivity Study on Modelling a Flow-Diverting Stent as a Porous Medium using Computational Fluid Dynamics

被引:0
|
作者
Li, Yujie [1 ,2 ]
Zhang, Mingzi [1 ,2 ]
Verrelli, David I. [1 ]
Yang, William [3 ]
Chong, Winston [4 ]
Ohta, Makoto [5 ]
Qian, Yi [1 ]
机构
[1] Macquarie Univ, Fac Med & Hlth Sci, Sydney, NSW 2109, Australia
[2] Tohoku Univ, Grad Sch Engn, Sendai, Miyagi, Japan
[3] CSIRO, Mineral Resources, Clayton, Vic, Australia
[4] Monash Med Ctr, Dept Neuroradiol, Melbourne, Vic, Australia
[5] Tohoku Univ, Inst Fluid Sci, Sendai, Miyagi, Japan
基金
澳大利亚研究理事会;
关键词
ANEURYSM HEMODYNAMICS; CEREBRAL ANEURYSMS; DIVERSION;
D O I
暂无
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The flow-diverting (FD) stent has become a commonly used endovascular device to treat cerebral aneurysms. This discourages blood from entering the aneurysm, thereby reducing the likelihood of aneurysm rupture. Using computational fluid dynamics (CFD) to simulate the aneurysmal haemodynamics after FD treatment could help clinicians predict the stent effectiveness prior to the real procedure in the patient. As an alternative to modelling the stent as a fine wire mesh, modelling the FD stent as a porous medium was established to save computational time, and has also been proved capable of predicting the same haemodynamics as obtained using the real FD stent geometry. The flow resistance effect of a porous-medium stent may differ with respect to its morphology or permeability; however, the flow resistance effect after adjusting these parameters had not been clarified. In this study, we analysed the haemodynamic changes caused by alterations of porous-medium thickness and permeability, thereby providing future porous-medium stent simulations with important information on the respective parametric sensitivities. We found significant sensitivity to permeability. Results were insensitive to thickness when permeability was adjusted beforehand to compensate. We also compared our results with observations from an in-vitro model, and found good agreement. This supports adoption of porous-medium models in future work.
引用
收藏
页码:3389 / 3392
页数:4
相关论文
共 50 条
  • [1] A COMPUTATIONAL MODELLING FOR HEMODYNAMIC CONDITIONS FOLLOWING FLOW-DIVERTING TREATMENT IN CEREBRAL ANEURYSMS
    Fu, Gen
    Untaroiu, Alexandrina
    PROCEEDINGS OF THE ASME FLUIDS ENGINEERING DIVISION SUMMER MEETING, 2018, VOL 2, 2018,
  • [2] Interventional Exclusion of Iliac Artery Aneurysms Using the Flow-Diverting Multilayer Stent
    Pieper, Claus Christian
    Meyer, Carsten
    Rudolph, Jens
    Verrel, Frauke
    Schild, Hans Heinz
    Wilhelm, Kai E.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 36 (04) : 917 - 925
  • [3] Interventional Exclusion of Iliac Artery Aneurysms Using the Flow-Diverting Multilayer Stent
    Claus Christian Pieper
    Carsten Meyer
    Jens Rudolph
    Frauke Verrel
    Hans Heinz Schild
    Kai E. Wilhelm
    CardioVascular and Interventional Radiology, 2013, 36 : 917 - 925
  • [4] Internal carotid artery "donut" aneurysm treated using DERIVO flow-diverting stent
    Nestorovic, Dragoslav
    Nikolic, Igor
    Milosevic-Medenica, Svetlana
    Janicijevic, Aleksandar
    Tasic, Goran
    SRPSKI ARHIV ZA CELOKUPNO LEKARSTVO, 2022, 150 (1-2) : 96 - 99
  • [5] Numerical simulation of aneurysmal haemodynamics with calibrated porous-medium models of flow-diverting stents
    Li, Yujie
    Zhang, Mingzi
    Verrelli, David I.
    Chong, Winston
    Ohta, Makoto
    Qian, Yi
    JOURNAL OF BIOMECHANICS, 2018, 80 : 88 - 94
  • [6] Computational fluid dynamics analysis of flow reduction induced by flow-diverting stents in intracranial aneurysms: a patient-unspecific hemodynamics change perspective
    Ouared, Rafik
    Larrabide, Ignacio
    Brina, Olivier
    Bouillot, Pierre
    Erceg, Gorislav
    Yilmaz, Hasan
    Lovblad, Karl-Olof
    Pereira, Vitor Mendes
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (12) : 1288 - 1293
  • [7] Fluid dynamics in intracranial aneurysms treated with flow-diverting stents: effect of multiple geometrical parameters
    Chassagne, Fanette
    Barbour, Michael C.
    Levitt, Michael R.
    Aliseda, Alberto
    JOURNAL OF FLUID MECHANICS, 2023, 973
  • [8] Prospective Study of Early MRI Appearances following Flow-Diverting Stent Placement for Intracranial Aneurysms
    McGuinness, B. J.
    Memon, S.
    Hope, J. K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (05) : 943 - 948
  • [9] Prediction of Cerebral Aneurysm Hemodynamics With Porous-Medium Models of Flow-Diverting Stents via Deep Learning
    Li, Gaoyang
    Song, Xiaorui
    Wang, Haoran
    Liu, Siwei
    Ji, Jiayuan
    Guo, Yuting
    Qiao, Aike
    Liu, Youjun
    Wang, Xuezheng
    FRONTIERS IN PHYSIOLOGY, 2021, 12
  • [10] Computational aspects in modelling of the convection flow in porous medium
    Chirita, Aurel
    Ene, Horia
    Nicolescu, Bogdan
    Carstea, Ion
    MATHEMATICAL METHODS AND APPLIED COMPUTING, VOL 1, 2009, : 390 - +