proficiency testing; PFA-100; external quality assessment; PFA-200; EQA; quality control; platelet function analyzer; IQC; PLATELET-FUNCTION ANALYZER; LIGHT TRANSMISSION AGGREGOMETRY; TREATED CARDIOVASCULAR PATIENTS; ANTIPHOSPHOLIPID ANTIBODIES; CLOSURE TIME; WHOLE-BLOOD; NEW-ZEALAND; ASSURANCE; COAGULATION; HEMOSTASIS;
D O I
10.1055/s-0034-1365844
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Platelet function testing is an essential component of comprehensive hemostasis evaluation within the framework of bleeding and/or bruising investigations, and it may also be performed to evaluate antiplatelet medication effects. Globally, the platelet function analyzer (PFA)-100 (Siemens Healthcare, Marburg, Germany) is the most used primary hemostasis-screening instrument and has also been recently remodeled/upgraded to the PFA-200. The PFA-100 is sensitive to a wide range of associated disorders, including platelet function defects and von Willebrand disease (VWD), as well as to various antiplatelet medications. The PFA-100 is also useful in therapy monitoring, especially in VWD. External quality assessment (EQA) (or proficiency testing) and internal quality control (IQC) are critical to ensuring quality of test practice, inclusive of all hemostasis tests. However, both EQA and IQC for platelet function testing, including the PFA-100, is logistically challenging, given theoretical requirements for production, storage, and shipment of large volumes of "stabilized" normal and pathological blood/platelets covering both normal function plus a wide variety of potential defects. We accordingly describe the development and testing of novel feasible approaches to both EQA and IQC of PFA-100/PFA-200 instruments, whereby a range of formulated "platelet function antagonist" materials are utilized. For EQA purposes, these are distributed to participants, and citrated normal whole blood collected on site is then added locally, thereby creating test material that can be locally evaluated. Several exercises have been conducted by the Royal College of Pathologists of Australasia Quality Assurance Program (RCPAQAP) over the past 6 years. A total of 26 challenges, with most designed to mimic moderate to severe primary hemostasis defects, have been tested in 26 to 50 laboratories depending on the year of dispatch. Numerical results for PFA-100/PFA-200 closure times (CTs) and interpretive comments supplied by participants are analyzed by the RCPAQAP. During this period, reported CTs for each challenge were within limits of expectation and good reproducibility was evidenced by repeated challenges. Coefficients of variation (CVs) generated for challenges using the two major PFA-100/PFA-200 cartridge types (collagen/adenosine diphosphate and collagen/epinephrine) are always similar to those obtained using native whole blood, and in general range from 15 to 25%. Interpretations are also in general consistent with expectations and test data provided by laboratories. The EQA created material has also been assessed within the context of possible IQC material. In conclusion, EQA and IQC processes for the PFA-100/PFA-200 have been developed that include highly reproducible test challenge processes, not only supporting the concept that EQA/IQC is possible for platelet function testing but also providing a valuable mechanism for monitoring and improving laboratory performance in this area.
引用
收藏
页码:239 / 253
页数:15
相关论文
共 50 条
  • [1] An update on quality control for the PFA-100/PFA-200
    Favaloro, Emmanuel J.
    Bonar, Roslyn
    [J]. PLATELETS, 2018, 29 (06) : 622 - 627
  • [2] Proficiency testing/external quality assurance for the PFA-100®
    Favaloro, Emmanuel J.
    Bonar, Roslyn
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (08) : 1393 - 1401
  • [3] An external quality assessment program for the PFA-100
    Favaloro, E. J.
    Bonar, R.
    Marsden, K.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2009, 7 : 568 - 568
  • [4] External quality assurance for the PFA-100®
    Favaloro, E. J.
    Bonar, R.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 (04) : 878 - 880
  • [5] External quality assessment of platelet function by the PFA-100: an update from the RCPAQAP Haematology
    Bonar, R. A.
    Favaloro, E. J.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2013, 11 : 471 - 471
  • [6] Description of an In Vitro Platelet Function Analyzer (PFA-100/PFA-200) 30 Years in the Making
    Favaloro, Emmanuel J.
    [J]. SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2024, 50 (02): : 320 - 324
  • [7] Quality control of platelets during storage by the PFA-100®:: A comparison to platelet aggregation
    Beck, KH
    [J]. TRANSFUSION AND APHERESIS SCIENCE, 2002, 27 (03) : 247 - 253
  • [8] PFA-100和INNOVANCE PFA-200血小板功能分析仪的临床应用及研究进展
    田奎朋
    关杰
    邓新立
    刘洁静
    丛玉隆
    [J]. 中国医学装备, 2017, 14 (04) : 170 - 174
  • [9] Time for a conceptual shift in assessment of internal quality control for whole blood or cell-based testing systems? An evaluation using platelet function and the PFA-100 as a case example
    Favaloro, Emmanuel J.
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (04) : 767 - 774
  • [10] An approach to an external quality assurance program (EQA) for the platelet function analyzer PFA-100: application to a systematic screening of platelet disorders
    Escolar, G.
    Escolar, G.
    Pino, M.
    Molina, P.
    Galan, A. M.
    Barraso, A.
    Arellano-Rodrigo, E.
    Reverter, J. C.
    Gutierrez, G.
    Diaz-Ricart, M.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 : 946 - 946