Lab-on-a-chip for multiplexed biosensing of residual antibiotics in milk

被引:43
|
作者
Suarez, Guillaume [1 ]
Jin, Young-Hyun [2 ]
Auerswald, Janko [1 ]
Berchtold, Stefan [1 ]
Knapp, Helmut F. [1 ]
Diserens, Jean-Marc [3 ]
Leterrier, Yves [2 ]
Manson, Jan-Anders E. [2 ]
Voirin, Guy [1 ]
机构
[1] CSEM Ctr Suisse Elect & Microtech SA, CH-2002 Neuchatel, Switzerland
[2] Ecole Polytech Fed Lausanne, Lab Technol Composites & Polymeres LTC, Stn 12, CH-1015 Lausanne, Switzerland
[3] Nestec Ltd, Nestle Res Ctr, CH-1000 Lausanne, Switzerland
关键词
BETA-LACTAM ANTIBIOTICS; MICROFLUIDIC DEVICES; SYSTEM; FLOW; FABRICATION;
D O I
10.1039/b819688e
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A multiplexed immunoassay-based antibiotic sensing device integrated in a lab-on-a-chip format is described. The approach is multidisciplinary and involves the convergent development of a multi-antibiotic competitive immunoassay based on sensitive wavelength interrogated optical sensor (WIOS) technology and a polymer-based self-contained microfluidic cartridge. Immunoassay solutions are pressure-driven through external and concerted actuation of a single syringe pump and multiposition valve. Moreover, the use of a novel photosensitive material in a 'one step' fabrication process allowed the rapid fabrication of microfluidic components and interconnection port simultaneously. Pre-filled microfluidic cartridges were used as binary response rapid tests for the simultaneous detection of three antibiotic families - sulfonamides, fluoroquinolones and tetracyclines - in raw milk. For test interpretation, any signal lower than the threshold value obtained for the corresponding Maximum Residue Limit (MRL) concentration (100 mu g L(-1)) was considered negative for a given antibiotic. The reliability of the multiplexed detection system was assessed by way of a validation test carried out on a series of six blind milk samples. A test accuracy of 95% was calculated from this experiment. The whole immunoassay procedure is fast (less than 10 minutes) and easy to handle (automated actuation).
引用
收藏
页码:1625 / 1630
页数:6
相关论文
共 50 条
  • [1] A fluorescence spectroscopy biosensor for lab-on-a-chip detection of antibiotics in milk
    Bosma, Rick
    Devasagayam, Jasen
    Singh, Ashutosh
    Collier, Christopher M.
    [J]. INTEGRATED OPTICS: DEVICES, MATERIALS, AND TECHNOLOGIES XXIV, 2020, 11283
  • [2] Recent advances in lab-on-a-chip for biosensing applications
    Lafleur, Josiane P.
    Joensson, Alexander
    Senkbeil, Silja
    Kutter, Joerg P.
    [J]. BIOSENSORS & BIOELECTRONICS, 2016, 76 : 213 - 233
  • [3] Integrated optical devices for lab-on-a-chip biosensing applications
    Estevez, M-Carmen
    Alvarez, Mar
    Lechuga, Laura M.
    [J]. LASER & PHOTONICS REVIEWS, 2012, 6 (04) : 463 - 487
  • [4] Lab-on-a-Chip Systems for Aptamer-Based Biosensing
    Khan, Niazul, I
    Song, Edward
    [J]. MICROMACHINES, 2020, 11 (02)
  • [5] Multiplexed optical pathogen detection with lab-on-a-chip devices
    Schulze, Holger
    Giraud, Gerard
    Crain, Jason
    Bachmann, Till T.
    [J]. JOURNAL OF BIOPHOTONICS, 2009, 2 (04) : 199 - 211
  • [6] Lab-on-a-chip for monitoring the quality of raw milk
    Choi, Jeong-Woo
    Kim, Young-Kee
    Kim, Hee-Joo
    Lee, Woochang
    Seong, Gi Hun
    [J]. JOURNAL OF MICROBIOLOGY AND BIOTECHNOLOGY, 2006, 16 (08) : 1229 - 1235
  • [7] Got milk? Lab-on-a-chip tests for contamination
    不详
    [J]. INTECH, 2004, 51 (07) : 14 - 14
  • [8] Editorial for the Special Issue on Microfluidics and Lab-on-a-Chip Applications for Biosensing
    Cerqueira, Laura
    Miranda, Joao Mario
    [J]. MICROMACHINES, 2022, 13 (12)
  • [9] Towards autonomous lab-on-a-chip devices for cell phone biosensing
    Comina, German
    Suska, Anke
    Filippini, Daniel
    [J]. BIOSENSORS & BIOELECTRONICS, 2016, 77 : 1153 - 1167
  • [10] Lab-on-a-chip Systems for Point-of-Care Biosensing Applications
    Im, Hyungsoon