X-ray tube voltage and image quality in adult and pediatric CT

被引:0
|
作者
Huda, W. [1 ]
Ogden, K. M. [1 ]
Scalzetti, E. M. [1 ]
Lavallee, R. L. [1 ]
Samei, E. [1 ]
机构
[1] SUNY Upstate Med Univ, Dept Radiol, Syracuse, NY 13210 USA
关键词
CT; image quality; contrast to noise ratio; pediatric patient; x-ray tube voltage; optimization;
D O I
10.1117/12.649279
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate how tissue x-ray attenuation coefficients, and their uncertainties, vary with x-ray tube voltage in different sized patients. Anthropomorphic phantoms (newborn, 10 year old, adult) were scanned a GE LightSpeed scanner at four x-ray tube voltages. Measurements were made of tissue attenuation in the head, chest and abdomen regions, as well as the corresponding noise values. Tissue signal to noise ratios (SNR) were obtained by dividing the average attenuation coefficient by the corresponding standard deviation. Soft tissue attenuation coefficients, relative to water, showed little variation with patient location or x-ray voltage (< 0.5%), but increasing the x-ray tube voltage from 80 to 140 kV reduced bone x-ray attenuation by similar to 14%. All tissues except adult bone showed a reduction of noise with increasing x-ray tube voltage (kV); the noise was found to be proportional to kV(n) and the average value of n for all tissues was -1.19 +/- 0.57. In pediatric patients at a constant x-ray tube voltage, SNR values were approximately independent of the body region, but the adult abdomen soft tissue SNR values were similar to 40% lower than the adult head. SNR values in the newborn were more than double the corresponding SNR soft tissue values in adults. SNR values for lung and bone were generally lower than those for soft tissues. For soft tissues, increasing the x-ray tube voltage from 80 to 140 kV increased the SNR by an average of similar to 90%. Data in this paper can be used to help design CT imaging protocols that take into account patient size and diagnostic imaging task.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Optimizing abdominal CT dose and image quality with respect to x-ray tube voltage
    Huda, W
    Ogden, KM
    [J]. MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 499 - 507
  • [2] Effect of X-ray Tube Parameters and Iodine Concentration on Image Quality and Radiation Dose in Cerebral Pediatric and Adult CT Angiography A Phantom Study
    Papadakis, Antonios E.
    Perisinakis, Kostas
    Raissaki, Maria
    Damilakis, John
    [J]. INVESTIGATIVE RADIOLOGY, 2013, 48 (04) : 192 - 199
  • [3] CT x-ray tube voltage optimisation and image reconstruction evaluation using visual grading analysis
    Zheng, Xiaoming
    Kim, Ted Myeongsoo
    Davidson, Rob
    Lee, Seongju
    Shin, Cheongil
    Yang, Sook
    [J]. MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [4] VARIATION OF IMAGE QUALITY WITH X-RAY TUBE POTENTIAL IN MAMMOGRAPHY
    LAW, J
    FAULKNER, K
    SMITH, S
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (734): : 192 - 192
  • [5] INFLUENCE OF X-RAY TUBE VOLTAGE ON EXPOSURE OF X-RAY FILMS
    ANGERSTEIN, W
    WOLF, M
    [J]. RADIOLOGIA DIAGNOSTICA, 1975, 16 (03) : 437 - 439
  • [6] X-RAY TUBE LIFETIME AND IMAGE QUALITY IN COMPUTED-TOMOGRAPHY
    SMITH, CB
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1981, 54 (645): : 820 - 820
  • [7] THE STABILIZATION OF X-RAY TUBE CURRENT AND VOLTAGE
    LEMIEUX, AF
    BEEMAN, WW
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1946, 17 (04): : 130 - 132
  • [8] A high voltage, demountable x-ray tube
    Kliever, WH
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1939, 10 (04): : 126 - 128
  • [9] Relationship between X-ray attenuation coefficient and X-ray tube voltage
    Feng, CC
    Zhou, W
    Wu, CL
    [J]. TRENDS IN NDE SCIENCE AND TECHNOLOGY - PROCEEDINGS OF THE 14TH WORLD CONFERENCE ON NDT (14TH WCNDT), VOLS 1-5, 1996, : 1357 - 1360
  • [10] A design of a low voltage x-ray tube
    Dershem, E
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA AND REVIEW OF SCIENTIFIC INSTRUMENTS, 1929, 18 (02): : 127 - 130