Radiation dose of cone-beam computed tomography compared to conventional radiographs in orthodontics

被引:2
|
作者
Signorelli, Luca [1 ]
Patcas, Raphael [1 ]
Peltomaki, Timo [2 ]
Schaetzle, Marc [1 ]
机构
[1] Univ Zurich, Ctr Dent Med, Clin Orthodont & Pediat Dent, Plattenstr 11, CH-8032 Zurich, Switzerland
[2] Tampere Univ, Hosp & Dept Otolaryngol, Dept Eye Ear & Oral Dis, Dent & Oral Dis Outpatient Clin, FIN-33101 Tampere, Finland
关键词
Radiation dosage; Diagnostic techniques; Organs at risk; Thyroid; Phantoms; imaging; HEAD FILM MEASUREMENTS; IMAGING TECHNIQUE; RELIABILITY; ACCURACY; CT; DOSIMETRY; IMAGES; CHOICE;
D O I
10.1007/s00056-015-0002-4
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives The aim of this study was to determine radiation doses of different cone-beam computed tomography (CBCT) scan modes in comparison to a conventional set of orthodontic radiographs (COR) by means of phantom dosimetry. Materials and methods Thermoluminescent dosimeter (TLD) chips (3 x 1 x 1 mm) were used on an adult male tissue-equivalent phantom to record the distribution of the absorbed radiation dose. Three different scanning modes (i.e., portrait, normal landscape, and fast scan landscape) were compared to CORs [i.e., conventional lateral (LC) and posteroanterior (PA) cephalograms and digital panoramic radiograph (OPG)]. Results The following radiation levels were measured: 131.7, 91, and 77 mu Sv in the portrait, normal landscape, and fast landscape modes, respectively. The overall effective dose for a COR was 35.81 mu Sv (PA: 8.90 mu Sv; OPG: 21.87 mu Sv; LC: 5.03 mu Sv). Disucssion Although one CBCT scan may replace all CORs, one set of CORs still entails 2-4 times less radiation than one CBCT. Depending on the scan mode, the radiation dose of a CBCT is about 3-6 times an OPG, 8-14 times a PA, and 15-26 times a lateral LC. Finally, in order to fully reconstruct cephalograms including the cranial base and other important structures, the CBCT portrait mode must be chosen, rendering the difference in radiation exposure even clearer (131.7 vs. 35.81 mu Sv). Shielding radiation-sensitive organs can reduce the effective dose considerably. Conclsion CBCT should not be recommended for use in all orthodontic patients as a substitute for a conventional set of radiographs. In CBCT, reducing the height of the field of view and shielding the thyroid are advisable methods and must be implemented to lower the exposure dose.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 50 条
  • [1] Cone-Beam Computed Tomography in Orthodontics
    Abdelkarim, Ahmad
    [J]. DENTISTRY JOURNAL, 2019, 7 (03)
  • [2] Cone-Beam Computed Tomography and Radiographs in Dentistry: Aspects Related to Radiation Dose
    Coelho Lorenzoni, Diego
    Maria Bolognese, Ana
    Gamba Garib, Daniela
    Ribeiro Guedes, Fabio
    Franzotti Sant'Anna, Eduardo
    [J]. INTERNATIONAL JOURNAL OF DENTISTRY, 2012, 2012
  • [3] Radiation dose of cone-beam computed tomography compared to conventional radiographs in orthodontics; [Strahlenbelastung beim digitalen Volumentomogramm im Vergleich zu konventionellen kieferorthopädischen radiologischen Unterlagen]
    Signorelli L.
    Patcas R.
    Peltomäki T.
    Schätzle M.
    [J]. Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie, 2016, 77 (1): : 9 - 15
  • [4] Dose and Image Quality of Cone-Beam Computed Tomography as Compared With Conventional Multislice Computed Tomography in Abdominal Imaging
    Schegerer, Alexander A.
    Lechel, Ursula
    Ritter, Manuel
    Weisser, Gerald
    Fink, Christian
    Brix, Gunnar
    [J]. INVESTIGATIVE RADIOLOGY, 2014, 49 (10) : 675 - 684
  • [5] Comparison of radiation dose for implant imaging using conventional spiral tomography, computed tomography, and cone-beam computed tomography
    Chau, Anson C. M.
    Fung, Karl
    [J]. ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 107 (04): : 559 - 565
  • [6] Practical applications of cone-beam computed tomography in orthodontics
    Mah, James K.
    Huang, John C.
    Choo, HyeRan
    [J]. JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2010, 141 (10): : 7S - 13S
  • [7] Comparison of Cephalometric Radiographs Obtained From Cone-Beam Computed Tomography Scans and Conventional Radiographs
    van Vlijmen, Olivier J. C.
    Berge, Stefaan J.
    Swennen, Gwen R. J.
    Bronkhorst, Ewald M.
    Katsaros, Christos
    Kuijpers-Jagtman, Anne M.
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 67 (01) : 92 - 97
  • [8] Accuracy of cone-beam computed tomography for syndesmosis injury diagnosis compared to conventional computed tomography
    Hamard, Marion
    Neroladaki, Angeliki
    Bagetakos, Ilias
    Dubois-Ferriere, Victor
    Montet, Xavier
    Boudabbous, Sana
    [J]. FOOT AND ANKLE SURGERY, 2020, 26 (04) : 265 - 272
  • [9] Conventional frontal radiographs compared with frontal radiographs obtained from cone beam computed tomography
    Nur, Metin
    Kayipmaz, Saadettin
    Bayram, Mehmet
    Celikoglu, Mevlut
    Kilkis, Dogan
    Sezgin, Omer Said
    [J]. ANGLE ORTHODONTIST, 2012, 82 (04) : 579 - 584
  • [10] Evidence supporting the use of cone-beam computed tomography in orthodontics
    van Vlijmen, Olivier J. C.
    Kuijpers, Mette A. R.
    Berge, Stefaan J.
    Schols, Jan G. J. H.
    Maal, Thomas J. J.
    Breuning, Hero
    Kuijpers-Jagtman, Anne Marie
    [J]. JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2012, 143 (03): : 241 - 252