Optimization of dose and image quality for computed radiography and digital radiography

被引:48
|
作者
Aldrich, JE
Duran, E
Dunlop, P
Mayo, JR
机构
[1] Vancouver Gen Hosp, Dept Radiol, Vancouver, BC V5Z 1M9, Canada
[2] Univ British Columbia, Vancouver, BC V5Z 1M9, Canada
[3] Vancouver Gen Hosp, Dept Occupat Hlth, Vancouver, BC, Canada
关键词
computed radiography; digital radiography; optimization; patient dose;
D O I
10.1007/s10278-006-9944-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The surface doses to patients during chest, abdomen and pelvis radiography were measured over a period of 3 years, during which time computed radiography (CR) and digital radiography (DR) systems were introduced to replace film-screen systems. For film-screen and CR the surface doses were measured with thermoluminescent dosimeters. For DR the surface doses were calculated from the dose-area product (DAP) meter readings. Measurements were made for each type of examination and detector type on 10 average-size patients. Measurements were made immediately after the new systems were introduced, and subsequently as adjustments were made to optimize dose and image quality. Published diagnostic reference levels were used as target values in this optimization. Initially, CR doses were the same as or higher than for film-screen, and the doses were lower for DR compared to film-screen. Subsequent clinical experience with the systems led to changes in the technique used for chest examinations both for CR and for DR. For CR, it was possible to change the algorithm and decrease the dose to one quarter of the initial value with acceptable image quality. For DR, it was decided to reduce noise by increasing the dose by a factor of two. No changes were made to abdomen or pelvic imaging techniques for either CR or DR. The final patient surface doses using CR were similar to published diagnostic reference doses; for DR, all patient doses were less than published reference levels.
引用
收藏
页码:126 / 131
页数:6
相关论文
共 50 条
  • [1] Optimization of Dose and Image Quality for Computed Radiography and Digital Radiography
    John E. Aldrich
    Emerenciana Duran
    Pat Dunlop
    John R. Mayo
    [J]. Journal of Digital Imaging, 2006, 19 : 126 - 131
  • [2] Optimization of dose and image quality in pediatrics chest digital radiography
    Deevband, Mohammad Reza
    Rahmani, Fereshteh
    Bardar, Meisam Ebrahimi
    Ghorbani, Mahdi
    [J]. JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2021, 52 (03) : 443 - 449
  • [3] Dose Optimization and Image Quality in Digital Radiography: A Review Article
    Barai, Juhi
    Tivaskar, Suhas
    Sable, Apeksha
    Luharia, Anurag
    [J]. JOURNAL OF PHARMACEUTICAL NEGATIVE RESULTS, 2022, 13 (03) : 1076 - 1080
  • [4] Dose optimization and image quality measurement in digital abdominal radiography
    Karal, Omer
    Tokgoz, Nazime
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2023, 205
  • [5] Optimization of image quality and minimization of radiation dose for chest computed radiography
    Kong, X.
    Liu, H.
    Rong, X.
    Sweet, C.
    Yang, Z.
    Edwards, S.
    Thai, T.
    Thomas, J.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2016 - 2016
  • [6] Quality and dose optimization in hand computed radiography
    Pavan, A. L. M.
    Alves, A. F. F.
    Duarte, S. B.
    Giacomini, G.
    Sardenberg, T.
    Miranda, J. R. A.
    Pina, D. R.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (08): : 1065 - 1069
  • [7] DIGITAL RADIOGRAPHY: IMAGE QUALITY AND RADIATION DOSE
    Seibert, J. Anthony
    [J]. HEALTH PHYSICS, 2008, 95 (05): : 586 - 598
  • [8] Evaluation of dose-image-quality optimization in digital chest radiography
    Lee, S. C.
    Wang, J. N.
    Liu, S. C.
    Jiang, S. H.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 544 - 547
  • [9] Dose and image quality optimization in neonatal radiography
    Dougeni, E. D.
    Delis, H. B.
    Karatza, A. A.
    Kalogeropoulou, C. P.
    Skiadopoulos, S. G.
    Mantagos, S. P.
    Panayiotakis, G. S.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (958): : 807 - 815
  • [10] Quality and dose optimization in canine chest radiography using a digital radiography system
    Santana de Souza, Sergio Augusto
    Fattori Alves, Allan Felipe
    Mamprim, Maria Jaqueline
    Pina, Diana Rodrigues
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2022, 195