Radiation exposure to patients and personnel during interventional ERCP at a teaching institution

被引:61
|
作者
Heyd, RL
Kopecky, KK
Sherman, S
Lehman, GA
Stockberger, SM
机构
[1] INDIANA UNIV,MED CTR,SCH MED,DEPT RADIOL,INDIANAPOLIS,IN 46202
[2] INDIANA UNIV,MED CTR,SCH MED,DEPT MED,DIV GASTROENTEROL,INDIANAPOLIS,IN 46202
关键词
D O I
10.1016/S0016-5107(96)70166-9
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background: This study was designed to determine the radiation dose to patients and personnel during ERCP procedures. Methods: Phantom studies were performed to compare dosimetry for two image capture systems and to determine the effectiveness of shielding in reducing stray radiation. Radiation techniques and dosimetry were recorded in 72 patients undergoing ERCP. Results: Phantom studies indicated that increasing fluoroscopy voltage from 75 to 96 kV decreased entrance dose by 50%. Image capture by digital radiography decreased radiation exposure by 66%. Shielding decreased stray radiation exposure by 93%. All patients underwent cholangiography (n = 71) and/or pancreatography (n = 53). The mean number of interventional ERCP procedures performed per patient was 1.8(range, 0 to 6). The mean measured patient entrance dose was 80 mGy (8 R; range, 0.2 to 73 R); however, the calculated mean entrance dose (based on measured intensifying screen doses) may have been as high as 3000 mGy (30 R; range, 0.8 to 300 R). Measured patient exposure increased with fluoroscopy time (r = 0.9) and with the number of interventions performed (r = 0.3). The mean dose to personnel was estimated at 0.04 mR. Conclusions: The patient radiation dose depended most on fluoroscopy time. The dose may be lowered by minimizing fluoroscopy time, using higher voltage and lower current for fluoroscopy, and using digital radiography for documentation. Personnel were adequately protected.
引用
收藏
页码:287 / 292
页数:6
相关论文
共 50 条
  • [1] Radiation exposure to personnel and patients during cardiac angiographic examinations and interventional procedures
    Zorzetto, M
    Padovani, R
    Bernardi, G
    Morocutti, G
    Fontanelli, A
    [J]. PHYSICA MEDICA, 1997, 13 : 177 - 177
  • [2] Radiation exposure to personnel performing ERCP
    Singhal, S
    Naidu, L
    Preece, DE
    Vohrah, A
    Loft, DE
    [J]. GUT, 2003, 52 : A5 - A5
  • [3] Correlation between fluoroscopy time and personnel radiation exposure during ERCP.
    Vandervoort, J
    Tham, TCK
    Wong, RCK
    Cialfi, SA
    CarrLocke, DL
    [J]. GASTROINTESTINAL ENDOSCOPY, 1996, 43 (04) : 127 - 127
  • [4] Experience of endoscopists and radiation exposure to patients during ERCP
    Li, A
    Lee, D
    Wong, S
    Chan, A
    [J]. GASTROINTESTINAL ENDOSCOPY, 2004, 59 (05) : AB122 - AB122
  • [5] RADIATION EXPOSURE OF PATIENTS AND OPERATORS DURING INTERVENTIONAL RADIOLOGY
    KRAHE, T
    EWEN, K
    LACKNER, K
    KOSTER, O
    NICOLAS, V
    [J]. FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NUKLEARMEDIZIN, 1986, 145 (02): : 217 - 220
  • [6] RADIATION EXPOSURE TO PATIENTS AND RADIOLOGISTS DURING INTERVENTIONAL PROCEDURES
    Molyvda-Athanasopoulou, E.
    Karlatira, M.
    Gotzamani-Psarrakou, A.
    Koulouris, Ch.
    Siountas, A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2011, 147 (1-2) : 86 - 89
  • [7] Optimisation of Radiation Exposure to Gastroenterologists and Patients during Therapeutic ERCP
    Alzimami, Khalid
    Sulieman, Abdelmoneim
    Paroutoglou, Georgios
    Potamianos, Spiros
    Vlychou, Marianna
    Theodorou, Kiki
    [J]. GASTROENTEROLOGY RESEARCH AND PRACTICE, 2013, 2013
  • [8] RADIATION EXPOSURE OF NONRADIOLOGIC HOSPITAL PERSONNEL DURING TREATMENT OF PATIENTS
    NURNBERGER, CE
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1960, 83 (03): : 507 - 513
  • [9] RADIATION EXPOSURE TO PERSONNEL DURING ANGIOGRAPHY
    GREENE, MW
    EKELUND, MJ
    PITMAN, RJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (03): : 587 - 587
  • [10] RADIATION EXPOSURE OF THE PERSONNEL DURING DECOMMISSIONING
    LORCHER, G
    PETRASCH, P