Axial strain calculation using a low-pass digital differentiator in ultrasound elastography

被引:102
|
作者
Luo, JW
Bai, J [1 ]
He, P
Ying, K
机构
[1] Tsing Hua Univ, Dept Biomed Engn, Beijing 100084, Peoples R China
[2] Wright State Univ, Dept Biomed Ind & Human Factors Engn, Dayton, OH 45435 USA
基金
中国国家自然科学基金;
关键词
D O I
10.1109/TUFFC.2004.1334844
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In ultrasound elastography, tissue axial strains are calculated from the gradient of the estimated axial displacements. However, the common differentiation operation amplifies the noises in the displacement estimation, especially at high frequencies. In this paper, a low-pass digital differentiator (LPDD) is proposed to calculate the axial strain from the estimated tissue displacement. Several LPDDs that have been well developed in the field of digital signal processing are presented. The corresponding performances are compared qualitatively and quantitatively in computer simulations and in preliminary phantom and in vitro experiments. The results are consistent with the theoretical analysis of the LPDDs.
引用
收藏
页码:1119 / 1127
页数:9
相关论文
共 50 条