Performance evaluation of the HAMAMATSU SHR22000 whole-body PET scanner using the IEC standard

被引:0
|
作者
Tian Yi [1 ]
Atsushi, Katabe
Liu Hua-Feng
Bao Chao
Takaji, Yamashita
机构
[1] Zhejiang Univ, State Ley Lab Modern Opt Instrumentat, Hangzhou 310027, Peoples R China
[2] Hamamatsu Photon KK, Cent Res Lab, Hamamatsu, Shizuoka 438601, Japan
关键词
PET; performance evaluation; GB/T; 18988.1; IEC; 61675-1;
D O I
暂无
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
SHR22000 is a whole-body PET scanner developed by Hamamatsu Photonics K.K. The scanner has a 600mm patient aperture and a 225mm axial FOV, which can, cover the whole human head and large portion of the body so that a common whole-body scan can be accomplished within 5 bed position with a 3.2mm average transverse resolution at 10mm from the axis of FOV. The paper gives a system performance evaluation of this scanner using GB/T 18988.1-2003/IEC 61675-1:1998 standard, which includes the spatial resolution, tomographic sensitivity, count rate characteristic, scatter fraction and recovery coefficient. At last a human body scanning result is given as an imaging example.