Performance evaluation of the automated NucliSens easyMAG nucleic acid extraction platform in comparison with QIAamp Mini kit from clinical specimens

被引:19
|
作者
Francesca, Perandin [1 ]
Patrizia, Pollara C. [2 ]
Franco, Gargiulo [2 ]
Carlo, Bonfanti
Manca, Nino
机构
[1] Univ Brescia, Dept Appl & Expt Med, Dept Microbiol, Microbiol Sect, I-25123 Brescia, Italy
[2] Spedali Civil Brescia, UO Microbiol, I-25123 Brescia, Italy
关键词
CMV; EBV; HIV; Automated extraction; Realtime PCR; RNA EXTRACTION; DNA; PCR; VIRUS;
D O I
10.1016/j.diagmicrobio.2009.02.013
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The performance of the NucliSens easyMAG platform for the extraction of nucleic acid from different clinical specimens was compared with a manual procedure. A total of 308 specimens were analyzed: 209 plasma samples collected for virus detection and quantification of cytomegalovirus (CMV) and Epstein-Barr virus (EBV) (n = 70), and 29 for HIV genotyping for drug resistance. Linearity of extraction was tested on dilution series of CMV and EBV the correlation coefficient (R-2) for standard curves based on repeated extraction runs was 0.99 for CMV and EBV. Inter- and intrarun variability was in accordance with previous studies, and the correlation between automated and manual extraction was very high. The concordant results were 95.7% for CMV and 100% for EBV The results of sequence analysis for HIV drug resistance showed a concordance in 24 of 29 specimens. The NucliSens easyMAG is extremely easy to perform, is automated, and resulted in a strong reduction of hands-on time compared with manual protocol. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:158 / 165
页数:8
相关论文
共 50 条
  • [1] Evaluation of NucliSens easyMAG for automated nucleic acid extraction from various clinical specimens
    Loens, K.
    Bergs, K.
    Ursi, D.
    Goossens, H.
    Ieven, M.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (02) : 421 - 425
  • [2] Automated sample extraction on the NucliSens easyMAG platform for nucleic acid testing of sexually transmitted diseases
    Stelzl, E.
    Scherlen, J.
    Burghardt, G.
    Noppen, C.
    Kessler, H. H.
    Viollier, E.
    Schaefer, C.
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2006, 36 : S50 - S50
  • [3] Evaluation of the new NucliSens easyMAG® nucleic acid extraction system
    Bossart, W.
    Burkhardt, F.
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2006, 36 : S25 - S25
  • [4] Automated nucleic acid extraction by QIAsymphony SP and nuclisens easyMAG for nucleic acid testing in microbiology
    Raymaekers, M.
    Martens, K.
    Lommelen, T.
    Vandebroek, K.
    Oris, E.
    Cartuyvels, R.
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2009, 46 : S50 - S50
  • [5] Comparison of eMAG™ versus NucliSENS® EasyMAG® performance on clinical specimens
    Garcia, Magali
    Chessa, Celine
    Bourgoin, Anne
    Giraudeau, Genevieve
    Plouzeau, Chloe
    Agius, Gerard
    Leveque, Nicolas
    Beby-Defaux, Agnes
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2017, 88 : 52 - 57
  • [6] Comparative evaluation of two automated systems for nucleic acid extraction of BK virus: NucliSens easyMAG versus BioRobot MDx
    Kim, Sollip
    Park, Sook-Ja
    Namgoong, Seung
    Sung, Heungsup
    Kim, Mi-Na
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2009, 162 (1-2) : 208 - 212
  • [7] Comparison of NucliSens easyMAG and Qiagen nucleic acid extraction using screening broths for the detection of MRSA
    Ieven, M.
    Michiels, M.
    Bergs, K.
    Ursi, D.
    Goossens, H.
    [J]. INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 2007, 29 : S232 - S232
  • [8] Comparison of NucliSens easyMAG and Qiagen nucleic acid extraction using screening broths for the detection of MRSA
    Michiels, M.
    Jansens, H.
    Bergs, K.
    Goossens, H.
    Leven, M.
    [J]. ACTA CLINICA BELGICA, 2007, 62 (06) : 454 - 454
  • [9] Optimization of Toxoplasma gondii DNA extraction from amniotic fluid using NucliSENS easyMAG and comparison with QIAamp DNA minikit
    Faucher, B.
    Miermont, F.
    Ranque, S.
    Franck, J.
    Piarroux, R.
    [J]. EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2012, 31 (06) : 1035 - 1039
  • [10] Improvement in fetal DNA extraction from maternal plasma. Evaluation of the NucliSens magnetic extraction system and the QlAamp DSP virus kit in comparison with the QIAamp DNA blood mini kit
    Clausen, Frederik Banch
    Krog, Grethe Risum
    Rieneck, Klaus
    Dziegiel, Morten Hanefeld
    [J]. PRENATAL DIAGNOSIS, 2007, 27 (01) : 6 - 10