共 50 条
Performance evaluation of the automated NucliSens easyMAG nucleic acid extraction platform in comparison with QIAamp Mini kit from clinical specimens
被引:19
|作者:
Francesca, Perandin
[1
]
Patrizia, Pollara C.
[2
]
Franco, Gargiulo
[2
]
Carlo, Bonfanti
Manca, Nino
机构:
[1] Univ Brescia, Dept Appl & Expt Med, Dept Microbiol, Microbiol Sect, I-25123 Brescia, Italy
[2] Spedali Civil Brescia, UO Microbiol, I-25123 Brescia, Italy
关键词:
CMV;
EBV;
HIV;
Automated extraction;
Realtime PCR;
RNA EXTRACTION;
DNA;
PCR;
VIRUS;
D O I:
10.1016/j.diagmicrobio.2009.02.013
中图分类号:
R51 [传染病];
学科分类号:
100401 ;
摘要:
The performance of the NucliSens easyMAG platform for the extraction of nucleic acid from different clinical specimens was compared with a manual procedure. A total of 308 specimens were analyzed: 209 plasma samples collected for virus detection and quantification of cytomegalovirus (CMV) and Epstein-Barr virus (EBV) (n = 70), and 29 for HIV genotyping for drug resistance. Linearity of extraction was tested on dilution series of CMV and EBV the correlation coefficient (R-2) for standard curves based on repeated extraction runs was 0.99 for CMV and EBV. Inter- and intrarun variability was in accordance with previous studies, and the correlation between automated and manual extraction was very high. The concordant results were 95.7% for CMV and 100% for EBV The results of sequence analysis for HIV drug resistance showed a concordance in 24 of 29 specimens. The NucliSens easyMAG is extremely easy to perform, is automated, and resulted in a strong reduction of hands-on time compared with manual protocol. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:158 / 165
页数:8
相关论文