Determination of relative ion chamber calibration coefficients from depth-ionization measurements in clinical electron beams

被引:13
|
作者
Muir, B. R. [1 ]
McEwen, M. R. [1 ]
Rogers, D. W. O. [2 ]
机构
[1] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
[2] Carleton Univ, Carleton Lab Radiotherapy Phys, Dept Phys, Ottawa, ON K1S 5B6, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 19期
基金
加拿大自然科学与工程研究理事会;
关键词
electron beams; calibration coefficient; reference dosimetry; REPLACEMENT CORRECTION FACTORS; WALL CORRECTION FACTORS; HIGH-ENERGY PHOTON; RADIOTHERAPY BEAMS; INITIAL ENERGY; K(Q) FACTORS; MEAN ENERGY; P-WALL; DOSIMETRY; CODE;
D O I
10.1088/0031-9155/59/19/5953
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A method is presented to obtain ion chamber calibration coefficients relative to secondary standard reference chambers in electron beams using depth-ionization measurements. Results are obtained as a function of depth and average electron energy at depth in 4, 8, 12 and 18 MeV electron beams from the NRC Elekta Precise linac. The PTW Roos, Scanditronix NACP-02, PTW Advanced Markus and NE 2571 ion chambers are investigated. The challenges and limitations of the method are discussed. The proposed method produces useful data at shallow depths. At depths past the reference depth, small shifts in positioning or drifts in the incident beam energy affect the results, thereby providing a built-in test of incident electron energy drifts and/or chamber set-up. Polarity corrections for ion chambers as a function of average electron energy at depth agree with literature data. The proposed method produces results consistent with those obtained using the conventional calibration procedure while gaining much more information about the behavior of the ion chamber with similar data acquisition time. Measurement uncertainties in calibration coefficients obtained with this method are estimated to be less than 0.5%. These results open up the possibility of using depth-ionization measurements to yield chamber ratios which may be suitable for primary standards-level dissemination.
引用
收藏
页码:5953 / 5969
页数:17
相关论文
共 50 条
  • [1] Ionization chamber shift correction and surface dose measurements in electron beams
    Das, IJ
    McNeeley, SW
    Cheng, CW
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (11): : 3419 - 3424
  • [2] Characterization of a multilayer ionization chamber for relative depth-dose curves in particle beams
    Vai, A.
    Maestri, D.
    Magro, G.
    Mairani, A.
    Mastella, E.
    Mirandola, A.
    Molinelli, S.
    Russo, S.
    Togno, M.
    La Civita, S.
    Ciocca, M.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S473 - S473
  • [3] Possibility of using cylindrical ionization chambers for percent depth-dose measurements in clinical electron beams
    Ono, Takeshi
    Araki, Fujio
    Yoshiyama, Fumiaki
    MEDICAL PHYSICS, 2011, 38 (08) : 4647 - 4654
  • [4] Dosimetric impact of calibration coefficients determined using linear accelerator photon and electron beams for ionization chamber in an on-site dosimetry audit
    Tani, Kensuke
    Wakita, Akihisa
    Tohyama, Naoki
    Fujita, Yukio
    JOURNAL OF RADIATION RESEARCH, 2024, 65 (05) : 619 - 627
  • [5] Determination of the effective point of measurement of a ionization chamber in light-ion beams
    Barna, S.
    Resch, A.
    Puchalska, M.
    Georg, D.
    Palmans, H.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S230 - S231
  • [6] CALIBRATION OF PARALLEL-PLATE IONIZATION-CHAMBER FOR DOSE MEASUREMENT IN ELECTRON-BEAMS
    GROENEWALD, WA
    SOUTH AFRICAN JOURNAL OF SCIENCE, 1987, 83 (09) : 572 - 572
  • [7] Determination of ion recombination correction factors for a liquid ionization chamber in megavoltage photon beams
    Choi, Sang Hyoun
    Kim, Kum-Bae
    Ji, Young Hoon
    Kim, Chan Hyeong
    Kim, Seonghoon
    Huh, Hyun Do
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (09) : 1439 - 1447
  • [8] Determination of ion recombination correction factors for a liquid ionization chamber in megavoltage photon beams
    Sang Hyoun Choi
    Kum-Bae Kim
    Young Hoon Ji
    Chan Hyeong Kim
    Seonghoon Kim
    Hyun Do Huh
    Journal of the Korean Physical Society, 2015, 66 : 1439 - 1447
  • [9] Performance assessment of ionization chambers in standard diagnostic radiology beams through the determination of calibration coefficients
    Akerele, Olumide Olaife
    Oyeyemi, Samuel Mofolorunsho
    Olaniyi, David Olakanmi
    Fashogbon, Oluwadamilare Timothy
    Agada, Francis Adole
    Kelani, Sherif Olaniyi
    Ladapo, Akinkunmi Emmanuel
    Shiyanbade, Ahmed Mohammed
    Adeniran, Bamidele Musbau
    Owoade, Latifat Ronke
    RADIATION MEASUREMENTS, 2025, 181
  • [10] Determination of penumbral widths from ion chamber measurements
    Herrup, D
    Chu, J
    Cheung, H
    Pankuch, M