High speed low coherence interferometer for optical coherence tomography

被引:3
|
作者
Hauger, C [1 ]
Wörz, M [1 ]
Hellmuth, T [1 ]
机构
[1] Carl Zeiss, D-73446 Oberkochen, Germany
关键词
OCT; white-light-interferometry; young's two-pinhole interference experiment; CCD-array;
D O I
10.1117/12.491305
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Optical coherence tomography (OCT) is a noninvasive imaging technology, which provides subsurface imaging of biological tissue with a resolution in the micrometer range. OCT sensors either work in the time or Fourier domain. We present a new interferometer setup based on a fiber double pinhole arrangement. Two fibers are placed in parallel similar to Young's two-pinhole interference experiment with spatial coherent and temporal incoherent light. The interference pattern is observed on a linear CCD-array. A complete A-scan can be derived from a single readout of the CCD-array. The experimental setup is described in detail. The main parameters of the setup are derived theoretically and compared with experiments. First images of technical and biological samples are presented.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 50 条
  • [1] Interferometer for optical coherence tomography
    Hauger, C
    Wörz, M
    Hellmuth, T
    APPLIED OPTICS, 2003, 42 (19) : 3896 - 3902
  • [2] Interferometer for optical coherence tomography
    Hauger, Christoph
    Wörz, Marco
    Hellmuth, Thomas
    Applied Optics, 2003, 42 (19): : 3896 - 3902
  • [3] Theoretical and experimental characterization of a stationary low coherence interferometer for optical coherence tomography
    Hauger, C
    Wang, L
    Wörz, M
    Hellmuth, T
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE TECHNIQUES, 2003, 5140 : 60 - 68
  • [4] High speed scanning optical coherence tomography
    Song, GJ
    Jacques, SL
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS: DIAGNOSTICS AND TREATMENT, 2002, 4916 : 259 - 265
  • [5] High speed, ultrahigh resolution optical coherence tomography
    Fujimoto, James G.
    2007 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, VOLS 1-4, 2007, : 247 - 247
  • [7] Optimal interferometer designs for optical coherence tomography
    Rollins, AM
    Izatt, JA
    OPTICS LETTERS, 1999, 24 (21) : 1484 - 1486
  • [8] Sagnac interferometer based optical coherence tomography
    Shalaby, Abdulrahman M.
    Shalaby, Mohamed
    AIN SHAMS ENGINEERING JOURNAL, 2024, 15 (03)
  • [9] Interferometer-in-Spectrometer for High-Resolution Optical Coherence Tomography
    Liu, Lishuang
    Xie, Jun
    Liu, Linbo
    Chen, Si
    JOURNAL OF SENSORS, 2020, 2020
  • [10] Optical coherence tomography with a nonlinear interferometer in the high parametric gain regime
    Machado, Gerard J.
    Frascella, Gaetano
    Torres, Juan P.
    Chekhova, Maria V.
    APPLIED PHYSICS LETTERS, 2020, 117 (09)