Shifting scintillator prototype large pixel wavelength-shifting fiber detector for the POWGEN3 powder diffractometer

被引:34
|
作者
Crow, ML [1 ]
Hodges, JP [1 ]
Cooper, RG [1 ]
机构
[1] Oak Ridge Natl Lab, Spallat Neutron Source, Oak Ridge, TN 37831 USA
关键词
thermal-neutron scintillator; neutron detector; neutron diffraction;
D O I
10.1016/j.nima.2004.04.167
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A prototype neutron scintillation detector has been developed, specifically aimed at the needs of the powder diffractometer POWGEN3 at the Spallation Neutron Source at Oak Ridge National Laboratory. This instrument requires a detector array with large (6 mm x 40 mm) pixels and an area of several square meters. The prototype uses a two-dimensional grid of wavelength shifting fibers, with the fiber axes parallel to the scintillator screen, to collect the scintillation photons. The fiber ends for each pixel go to a specific set of four photomultiplier tubes, so that the position of each event can be determined by a 4-tube C-2(n) X C-2(n) coded coincidence. The observed maximum light yield with a (LiF)-Li-6/ZnS:Ag neutron scintillation screen, summed over four tubes, is greater than 200 photons/neutron. This is about 0.13% of the similar to150,000 photons/neutron produced in the scintillator. The light yield is sufficient to allow pulse discrimination between neutron signals and gamma-ray background. Further light collection gains should be achievable using double-clad fiber and green-enhanced photomultiplier tubes. Currently, the shape, structure, and specific composition of the scintillator are being investigated, on the assumption that (LiF)-Li-6/ZnS:Ag will be the chosen scintillator material. Published by Elsevier B.V.
引用
收藏
页码:287 / 292
页数:6
相关论文
共 33 条
  • [1] A liquid scintillator detector with wavelength-shifting fibre readout
    Doucet, M
    Fabre, JP
    Grégoire, G
    Panman, J
    Zucchelli, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 453 (03): : 545 - 552
  • [2] SCINTILLATOR ANTICOINCIDENCE DETECTOR USING WAVELENGTH-SHIFTING LIGHTGUIDES
    KOTOV, YD
    KUROCHKIN, AV
    SUSLOV, AY
    YUROV, VN
    GALOCHKIN, KE
    MALINOVSKAYA, SA
    KORNILONVSKAYA, LD
    LEBEDEVA, LP
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1993, 36 (04) : 537 - 540
  • [3] High efficiency plastic scintillator detector with wavelength-shifting fiber readout for the GLAST Large Area Telescope
    Moiseev, A. A.
    Deering, P. L.
    Hartman, R. C.
    Johnson, T. E.
    Nebel, T. R.
    Ormes, J. F.
    Thompson, D. J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 583 (2-3): : 372 - 381
  • [4] Performance study of a prototype PET scanner using CsI(Na) scintillator with wavelength-shifting fiber readout
    Zavarzin, V
    Johnson, O
    Kudrolli, H
    Worstell, W
    SCIFI 97: CONFERENCE ON SCINTILLATING FIBER DETECTORS, 1998, 450 : 489 - 494
  • [5] The Nonuniformity of the Light Yield in Scintillator Strips with Wavelength-Shifting Fibers of the DANSS Detector
    Alekseev, I. G.
    Kalinkin, D. V.
    Machikhiliyan, I. V.
    Nesterov, V. M.
    Pogorelov, N. A.
    Rusinov, V. Yu.
    Svirida, D. N.
    Starostin, A. S.
    Tarkovsky, E. I.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2018, 61 (03) : 328 - 331
  • [6] Light yield measurements in a liquid scintillator detector with wavelength-shifting fibre readout
    Doucet, M
    Fabre, JP
    Grégoire, G
    Panman, J
    Zucchelli, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 459 (03): : 459 - 468
  • [7] The Nonuniformity of the Light Yield in Scintillator Strips with Wavelength-Shifting Fibers of the DANSS Detector
    I. G. Alekseev
    D. V. Kalinkin
    I. V. Machikhiliyan
    V. M. Nesterov
    N. A. Pogorelov
    V. Yu. Rusinov
    D. N. Svirida
    A. S. Starostin
    E. I. Tarkovsky
    Instruments and Experimental Techniques, 2018, 61 : 328 - 331
  • [8] Design of a muon tomography system with a plastic scintillator and wavelength-shifting fiber arrays
    Jo, Woo Jin
    Kim, Hyun-Il
    An, Su Jung
    Lee, Chae Young
    Baek, Cheol-Ha
    Chung, Yong Hyun
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 732 : 568 - 572
  • [9] Development of a wavelength-shifting fiber-based photon detector for LBNE
    Wasserman, R.
    Buchanan, N.
    JOURNAL OF INSTRUMENTATION, 2013, 8
  • [10] A large-area two-dimensional scintillator detector with a wavelength-shifting fibre readout for a time-of-flight single-crystal neutron diffractometer
    Nakamura, T.
    Kawasaki, T.
    Hosoya, T.
    Toh, K.
    Oikawa, K.
    Sakasai, K.
    Ebine, M.
    Birumachi, A.
    Soyama, K.
    Katagiri, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 686 : 64 - 70