Closed-Loop Control of Humidification for Artifact Reduction in Capacitive ECG Measurements

被引:19
|
作者
Leicht, Lennart [1 ]
Eilebrecht, Benjamin [2 ]
Weyer, Soren [1 ]
Leonhardt, Steffen [1 ]
Teichmann, Daniel [1 ]
机构
[1] Rhein Westfal TH Aachen, Philips Chair Med Informat Technol, D-52074 Aachen, Germany
[2] Wilo, Dortmund, Germany
关键词
Active humidification; automotive seat; bioinstrumentation; capacitive ECG; non-contact monitoring; triboelectricity; ELECTRODE; SYSTEM; INTERFERENCE; SENSOR;
D O I
10.1109/TBCAS.2016.2613097
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recording biosignals without the need for direct skin contact offers new opportunities for ubiquitous health monitoring. Electrodes with capacitive coupling have been shown to be suitable for the monitoring of electrical potentials on the body surface, in particular ECG. However, due to triboelectric charge generation and motion artifacts, signal and thus diagnostic quality is inferior to galvanic coupling. Active closed-loop humidification of capacitive electrodes is proposed in this work as a new concept to improve signal quality. A capacitive ECG recording system integrated into a common car seat is presented. It can regulate the micro climate at the interface of electrode and patient by actively dispensing water vapour and monitoring humidity in a closed-loop approach. As a regenerative water reservoir, silica gel is used. The system was evaluated with respect to subjective and objective ECG signal quality. Active humidification was found to have a significant positive effect in case of previously poor quality. Also, it had no diminishing effect in case of already good signal quality.
引用
收藏
页码:300 / 313
页数:14
相关论文
共 50 条
  • [1] Electrode Humidification Design for Artifact Reduction in Capacitive ECG Measurements
    Tang, Yue
    Chang, Ronghui
    Zhang, Limin
    Yan, Feng
    Ma, Haowen
    Bu, Xiaofeng
    SENSORS, 2020, 20 (12) : 1 - 17
  • [2] Closed-loop Control of a Dynamic Capacitive Wireless Power Transfer System
    Sinha, Sreyam
    Afridi, Khurram K.
    2019 IEEE 20TH WORKSHOP ON CONTROL AND MODELING FOR POWER ELECTRONICS (COMPEL), 2019,
  • [3] A BI-AXIAL CAPACITIVE SCANNING MIRROR WITH CLOSED-LOOP CONTROL
    Lin, Heng
    Lin, Ta-Wei
    Hung, Andrew C. -L.
    Lu, Michael S. -C.
    2018 IEEE MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2018, : 567 - 570
  • [4] A closed-loop ΣΔ modulator for micromechanical capacitive sensors
    Lv, Risheng
    Chen, Weiping
    Yin, Liang
    Fu, Qiang
    Liu, Xiaowei
    Yan, Jingmin
    IEICE ELECTRONICS EXPRESS, 2018, 15 (04):
  • [5] Capacitive Sensing Circuit for Closed-Loop Control of Wide Tuning Range Microspectrometers
    Susli, Mohamad
    Silva, K. K. M. B. Dilusha
    Boussaid, Farid
    Faraone, Lorenzo
    Dell, John M.
    NEXT-GENERATION SPECTROSCOPIC TECHNOLOGIES III, 2010, 7680
  • [7] Order reduction and closed-loop vibration control in helicopter fuselages
    Mathews, A
    Sule, VR
    Venkatesan, C
    JOURNAL OF GUIDANCE CONTROL AND DYNAMICS, 2002, 25 (02) : 316 - 323
  • [8] CLOSED-LOOP CONTROL IS COMING
    GILMAN, HD
    GOODMAN, MY
    DEMOYER, R
    RADZIUL, JV
    WATER & WASTES ENGINEERING, 1973, 10 (06): : 42 - 44
  • [9] CLOSED-LOOP CONTROL OF ENERGY
    HAINES, RW
    HEATING-PIPING-AIR CONDITIONING, 1983, 55 (12): : 96 - 97
  • [10] Closed-loop system of a capacitive micro-accelerometer
    Chen, Wei-Ping
    Wang, Tian-Yang
    Yin, Liang
    Chen, Xiao-Liang
    Guo, Yu-Gang
    Liu, Xiao-Wei
    Harbin Gongye Daxue Xuebao/Journal of Harbin Institute of Technology, 2010, 42 (11): : 1720 - 1723