Transvenous lead performance of implantable cardioverter-defibrillators and pacemakers

被引:7
|
作者
Mori, Hitoshi [1 ]
Kato, Ritsushi [1 ]
Ikeda, Yoshifumi [1 ]
Tsutsui, Kenta [1 ]
Hoya, Hiroki [1 ]
Tanaka, Sayaka [1 ]
Iwanaga, Shiro [1 ]
Nakano, Shintaro [1 ]
Muramatsu, Toshihiro [1 ]
Sumitomo, Naokata [2 ]
Matsumoto, Kazuo [1 ]
机构
[1] Saitama Med Univ, Dept Cardiol, Int Med Ctr, 1397-1 Yamane Hidaka, Saitama 3501298, Japan
[2] Saitama Med Univ, Int Med Ctr, Dept Pediat Cardiol, Saitama, Japan
来源
关键词
failure rate; implantable cardioverter defibrillator; lead failure; pacemaker; predictors;
D O I
10.1111/pace.14154
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: After the reports of recalled leads, several technological improvements have been introduced and the durability of implantable cardioverter defibrillator (ICD) leads has improved. The incidence of lead failures is now less than in the previous studies. However, there are few reports that have shown the long-term durability of ICD leads as compared to pacemaker (PM) leads. This study analyzed the medium to long-term performance of transvenous ICD leads as compared to PM leads. Methods: We retrospectively studied 1227 cases from April 2007 to December 2017 who underwent an initial transvenous ICD or PM implantation. The number of lead failures and patient background characteristics were analyzed. Results: During a median 3-3.5 years follow up period, 1 (0.3%) ICD lead and 18 (2.4%) PM leads failed. The incidence of lead failures was significantly higher in the PM group than ICD group (p = .019). Males were associated with a higher incidence of lead failures in the PM group. Conclusion: Since the era of recalled ICD leads, the durability of ICD leads has remarkably improved and the incidence of lead failures with non-recalled ICD leads has been less than that for PM leads.
引用
收藏
页码:481 / 489
页数:9
相关论文
共 50 条
  • [1] IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS
    BARDY, GH
    HOFER, B
    JOHNSON, G
    KUDENCHUK, PJ
    POOLE, JE
    DOLACK, GL
    GLEVA, M
    MITCHELL, R
    KELSO, D
    [J]. CIRCULATION, 1993, 87 (04) : 1152 - 1168
  • [2] Troubleshooting pacemakers and implantable cardioverter-defibrillators
    Scher, DL
    [J]. CURRENT OPINION IN CARDIOLOGY, 2004, 19 (01) : 36 - 46
  • [3] Deactivation of Pacemakers and Implantable Cardioverter-Defibrillators
    Kramer, Daniel B.
    Mitchell, Susan L.
    Brock, Dan W.
    [J]. PROGRESS IN CARDIOVASCULAR DISEASES, 2012, 55 (03) : 290 - 299
  • [4] Psychosocial impact of pacemakers and implantable cardioverter-defibrillators
    Candinas, R
    Buchi, S
    Klaghofer, R
    Sensky, T
    Buddeberg, C
    Duru, F
    [J]. XXI CONGRESS OF THE EUROPEAN SOCIETY OF CARDIOLOGY, 1999, : 857 - 862
  • [5] Pacemakers and implantable cardioverter-defibrillators in pediatric patients
    Silka, Michael J.
    Bar-Cohen, Yaniv
    [J]. HEART RHYTHM, 2006, 3 (11) : 1360 - 1366
  • [6] Management of infected pacemakers and implantable cardioverter-defibrillators
    Sohail, M. R.
    [J]. INTERNAL MEDICINE JOURNAL, 2007, 37 (07)
  • [7] Sensing lead-related complications in patients with transvenous implantable cardioverter-defibrillators
    Lawton, JS
    Ellenbogen, KA
    Wood, MA
    Stambler, BS
    Herre, JM
    Nath, S
    Bernstein, RC
    DiMarco, JP
    Haines, DE
    Szentpetery, S
    Baker, LD
    Damiano, RJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (06): : 647 - 651
  • [8] Management of infected pacemakers and implantable cardioverter-defibrillators - Reply
    Post, J. J.
    Alexopoulos, C.
    Jones, P. D.
    [J]. INTERNAL MEDICINE JOURNAL, 2007, 37 (07) : 510 - 510
  • [9] FAILURES OF EPICARDIAL AND TRANSVENOUS LEADS FOR IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS
    MATTKE, S
    MULLER, D
    MARKEWITZ, A
    KAULBACH, H
    SCHMOCKEL, M
    DORWARTH, U
    HOFFMANN, E
    STEINBECK, G
    [J]. AMERICAN HEART JOURNAL, 1995, 130 (05) : 1040 - 1044
  • [10] TRANSVENOUS AND SUBCUTANEOUS IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS - RADIOGRAPHIC ASSESSMENT
    DALY, BD
    CASCADE, PN
    HUMMEL, JD
    KALBFLEISCH, S
    STRICKBERGER, SA
    LANGBERG, J
    MORADY, F
    [J]. RADIOLOGY, 1994, 191 (01) : 273 - 278