Super-Resolution in Digital Breast Tomosynthesis: Limitations of the Conventional System Design and Strategies for Optimization

被引:0
|
作者
Acciavatti, Raymond J. [1 ]
Vent, Trevor L. [1 ]
Barufaldi, Bruno [1 ]
Wileyto, E. Paul [2 ]
Noel, Peter B. [1 ]
Maidment, Andrew D. A. [1 ]
机构
[1] Univ Penn, Dept Radiol, 3400 Spruce St, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Epidemiol Biostat & Informat, 423 Guardian Dr, Philadelphia, PA 19104 USA
关键词
Digital Breast Tomosynthesis; Virtual Clinical Trials; Super-Resolution; Calcifications; Anthropomorphic Phantom; Fourier Transform; Image Quality; Image Reconstruction;
D O I
10.1117/12.2563839
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Our previous work explored the use of super-resolution as a way to improve the visibility of calcifications in digital breast tomosynthesis. This paper demonstrates that there are anisotropies in super-resolution throughout the reconstruction, and investigates new motion paths for the x-ray tube to suppress these anisotropies. We used a theoretical model of a sinusoidal test object to demonstrate the existence of the anisotropies. In addition, high-frequency test objects were simulated with virtual clinical trial (VCT) software developed for breast imaging. The simulated objects include a lead bar pattern phantom as well as punctate calcifications in a breast-like background. In a conventional acquisition geometry in which the source motion is directed laterally, we found that super-resolution is not achievable if the frequency is oriented in the perpendicular direction (posteroanteriorly). Also, there are positions, corresponding to various slices above the breast support, at which super-resolution is inherently not achievable. The existence of these anisotropies was validated with VCT simulations. At locations predicted by theoretical modeling, the bar pattern phantom showed aliasing, and the spacing between individual calcifications was not properly resolved. To show that super-resolution can be optimized by re-designing the acquisition geometry, we applied our theoretical model to the analysis of new motion paths for the x-ray tube; specifically, motions with more degrees of freedom and with more rapid pulsing (submillimeter spacing) between source positions. These two strategies can be used in combination to suppress the anisotropies in super-resolution.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Observation of super-resolution in digital breast tomosynthesis
    Acciavatti, Raymond J.
    Maidment, Andrew D. A.
    [J]. MEDICAL PHYSICS, 2012, 39 (12) : 7518 - 7539
  • [2] Investigating the Potential for Super-Resolution in Digital Breast Tomosynthesis
    Acciavatti, Raymond J.
    Maidment, Andrew D. A.
    [J]. MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [3] Modeling Acquisition Geometries with Improved Super-Resolution in Digital Breast Tomosynthesis
    Acciavatti, Raymond J.
    Wileyto, E. Paul
    Maidment, Andrew D. A.
    [J]. MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [4] Impact of super-resolution and image acquisition on the detection of calcifications in digital breast tomosynthesis
    Barufaldi, Bruno
    Acciavatti, Raymond J.
    Conant, Emily F.
    Maidment, Andrew D. A.
    [J]. EUROPEAN RADIOLOGY, 2024, 34 (01) : 193 - 203
  • [5] Impact of super-resolution and image acquisition on the detection of calcifications in digital breast tomosynthesis
    Bruno Barufaldi
    Raymond J. Acciavatti
    Emily F. Conant
    Andrew D. A. Maidment
    [J]. European Radiology, 2024, 34 : 193 - 203
  • [6] Proposing Rapid Source Pulsing for Improved Super-Resolution in Digital Breast Tomosynthesis
    Acciavatti, Raymond J.
    Vent, Trevor L.
    Barufaldi, Bruno
    Wileyto, E. Paul
    Noel, Peter B.
    Maidment, Andrew D. A.
    [J]. MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [7] Aberration effect and optimization design of super-resolution telescope optical system
    Shao Hong-yu
    Li Ying-chao
    Wang Chao
    Shi Hao-dong
    Liu Zhuang
    Li Guan-lin
    [J]. CHINESE OPTICS, 2020, 13 (01): : 106 - 120
  • [8] An Alternate Design for the Defrise Phantom To Quantify Resolution in Digital Breast Tomosynthesis
    Acciavatti, Raymond J.
    Mannherz, William
    Nolan, Margaret
    Maidment, Andrew D. A.
    [J]. MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [9] Limitations of Learning-Based Super-Resolution
    Shoji, Hiroki
    Gohshi, Seiichi
    [J]. 2015 INTERNATIONAL SYMPOSIUM ON INTELLIGENT SIGNAL PROCESSING AND COMMUNICATION SYSTEMS (ISPACS), 2015, : 646 - 651
  • [10] Super-resolution of turbulent video: Potentials and limitations
    Yaroslavsky, Leonid P.
    Shabat, Gil
    Fishbain, Barak
    Ideses, Ianir A.
    [J]. IMAGE PROCESSING: ALGORITHMS AND SYSTEMS VI, 2008, 6812