A simulation study of a dual-plate in-room PET system for dose verification in carbon ion therapy

被引:2
|
作者
Chen Ze [1 ,2 ]
Hu Zheng-Guo [1 ]
Chen Jin-Da [1 ]
Zhang Xiu-Ling [1 ]
Guo Zhong-Yan [1 ]
Xiao Guo-Qing [1 ]
Sun Zhi-Yu [1 ]
Huang Wen-Xue [1 ]
Wang Jian-Song [1 ]
机构
[1] Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
基金
中国科学院西部之光基金; 中国国家自然科学基金;
关键词
hadron therapy; PET; GATE; simulation; RADIOTHERAPY; PROTON;
D O I
10.1088/1674-1137/38/8/088202
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
During carbon ion therapy, lots of positron emitters such as C-11, O-15, C-15 are generated in irradiated tissues by nuclear reactions, and can be used to track the carbon beam in the tissue by a positron emission tomography (PET) scanner. In this study, an dual-plate in-room PET scanner has been designed and evaluated based on the GATE simulation platform to monitor patient dose in carbon ion therapy. The dual-plate PET is designed to avoid interference with the carbon beamline and with patient positioning. Its performance was compared with that of four-head and full-ring PET scanners. The dual-plate, four-head and full-ring PET scanners consisted of 30, 60, 60 detector modules, respectively, with a 36 cm distance between directly opposite detector modules for dose deposition measurements. Each detector module consisted of a 24 x 24 array of 2 mm x 2 mm x 18 mm LYSO pixels coupled to a Hamamatsu H8500 PMT. To estimate the production yield of positron emitters, a 10 cm x 15 cm x 15 cm cuboid PMMA phantom was irradiated with 172, 200, 250 MeV/u C-12 beams. 3D images of the activity distribution measured by the three types of scanner are produced by an iterative reconstruction algorithm. By comparing the longitudinal profile of positron emitters along the carbon beam path, it is indicated that use of the dual-plate PET scanner is feasible for monitoring the dose distribution in carbon ion therapy.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] A simulation study of a dual-plate in-room PET system for dose verification in carbon ion therapy
    陈泽
    胡正国
    陈金达
    张秀玲
    郭忠言
    肖国青
    孙志宇
    黄文学
    王建松
    Chinese Physics C, 2014, 38 (08) : 118 - 123
  • [2] A simulation study of a dual-plate in-room PET system for dose verification in carbon ion therapy
    陈泽
    胡正国
    陈金达
    张秀玲
    郭忠言
    肖国青
    孙志宇
    黄文学
    王建松
    Chinese Physics C, 2014, (08) : 118 - 123
  • [3] Precision of the measurements of distal fall-off of dose profiles in carbon ion therapy with a dedicated in-room PET system
    Belcari, N.
    Sportelli, G.
    Camarlinghi, N.
    Straub, K.
    Ferretti, S.
    Marino, N.
    Cirrone, G. A. P.
    Cuttone, G.
    Romano, F.
    Rosso, V.
    Del Guerra, A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S499 - S499
  • [4] A simulation study of a C-shaped in-beam PET system for dose verification in carbon ion therapy
    An, Su Jung
    Beak, Cheol-Ha
    Lee, Kisung
    Chung, Yong Hyun
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 698 : 37 - 43
  • [5] First in-room PET studies for range verification of proton therapy
    Zhu, Xuping
    Espana, Samuel
    Paganetti, Harald
    Bortfeld, Thomas
    Liebsch, Norbert
    Shih, Helen
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [6] Monte Carlo simulation study of a dual-plate PET camera dedicated to breast cancer Imaging
    Zhang, Jin
    Olcott, Peter D.
    Foudray, Angela M. K.
    Chinn, Garry
    Levin, Craig S.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1667 - 1671
  • [7] Evaluation of Distal Dose Surface with In-Room PET for Proton Therapy Monitoring
    Min, C.
    Zhu, X.
    Grogg, K.
    Winey, B.
    Fakhri, G.
    Bortfeld, T.
    Shih, H.
    Paganetti, H.
    MEDICAL PHYSICS, 2012, 39 (06) : 3860 - 3860
  • [8] Feasibility of Using Distal Endpoints for In-Room PET Range Verification of Proton Therapy
    Grogg, Kira
    Zhu, Xuping
    Min, Chul Hee
    Winey, Brian
    Bortfeld, Thomas
    Paganetti, Harald
    Shih, Helen A.
    El Fakhri, Georges
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (05) : 3290 - 3297
  • [9] Feasibility of using distal endpoints for In-room PET Range Verification of Proton Therapy
    Grogg, K.
    Zhu, X.
    Min, C. H.
    Winey, B.
    Bortfeld, T.
    Paganetti, H.
    Shih, H.
    El Fakhri, G.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3890 - 3894
  • [10] Novel in-room proton range verification using mobile PET: Potential for proton therapy monitoring
    Zhu, Xuping
    Knopf, Antje
    Bonab, Ali A.
    Engelsman, Martijn
    Paganetti, Harald
    Brady, Thomas
    Bortfeld, Thomas
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50