Tissue shrinkage in microwave thermal ablation: comparison of three commercial devices

被引:34
|
作者
Farina, Laura [1 ]
Nissenbaum, Yitzhak [2 ]
Cavagnaro, Marta [1 ]
Goldberg, S. Nahum [2 ,3 ]
机构
[1] Sapienza Univ Rome, Dept Informat Engn Elect & Telecommun, Via Eudossiana 18, I-00184 Rome, Italy
[2] Hadassah Hebrew Univ, Dept Radiol, Med Ctr, Jerusalem, Israel
[3] Beth Israel Deaconess Med Ctr, Dept Radiol, 330 Brookline Ave, Boston, MA 02215 USA
关键词
Microwave thermal ablation; tissue shrinkage; tissue characterisation; 2.45; GHZ; DIELECTRIC-PROPERTIES; TEMPERATURE; ANTENNA; LIVER; PERFORMANCE; DESIGN; MODELS; SYSTEM; MHZ;
D O I
10.1080/02656736.2017.1362115
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate, characterise and compare the extent of tissue shrinkage induced from three different commercial microwave ablation devices, and to elucidate the mechanism behind the distinctive performances obtained.Materials and methods: Microwave ablation (N=152) was conducted with three different commercial devices on cubes of ex vivo liver (10-402mm/side) embedded in agar phantoms. 50-60W was applied for 1-10min duration. Pre- and post-ablation dimensions of the samples, as well as the extent of carbonisation and coagulation were measured and correlated. ANOVA was performed to evaluate statistical significance.Results: For all devices, logarithmic correlations with time were observed for both tissue shrinkage (R-2=0.84-1.00) and induced carbonisation (R-2=0.73-0.99) radially to the antenna axis. Along the longitudinal axis of the antenna, for two of the devices shrinkage did not appreciably change with time (p>0.05), yet carbonisation increased linearly (R-2=0.57-0.94). For the third fully internally-cooled device, both carbonisation and shrinkage showed logarithmic trends (R-2=0.85-0.98 and R-2=0.78-0.94, respectively) based upon delayed carbonisation appearing only 5min into ablation and onward. For all devices, non-uniform shrinkage was noted within the coagulated area increasing from the boundary of the ablated area (14%) to the limit of carbonisation (39%) in a linear fashion (R-2=0.88)Conclusions: Microwave ablation device construction can alter the extent of post-ablation coagulation and tissue shrinkage. Given that tissue shrinkage in the coagulated area shows non-uniform behaviour, observed differences can be attributed in part to the applicator cooling system that alters the ablation temperature profile.
引用
收藏
页码:382 / 391
页数:10
相关论文
共 50 条
  • [1] Characterisation of tissue shrinkage during microwave thermal ablation
    Farina, Laura
    Weiss, Noam
    Nissenbaum, Yitzhak
    Cavagnaro, Marta
    Lopresto, Vanni
    Pinto, Rosanna
    Tosoratti, Nevio
    Amabile, Claudio
    Cassarino, Simone
    Goldberg, S. Nahum
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2014, 30 (07) : 419 - 428
  • [2] Thermal Ablation of Lung Tissue: In Vivo Experimental Comparison of Microwave and Radiofrequency
    Crocetti, Laura
    Bozzi, Elena
    Faviana, Pinuccia
    Cioni, Dania
    Della Pina, Clotilde
    Sbrana, Alberto
    Fontanini, Gabriella
    Lencioni, Riccardo
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 33 (04) : 818 - 827
  • [3] Thermal Ablation of Lung Tissue: In Vivo Experimental Comparison of Microwave and Radiofrequency
    Laura Crocetti
    Elena Bozzi
    Pinuccia Faviana
    Dania Cioni
    Clotilde Della Pina
    Alberto Sbrana
    Gabriella Fontanini
    Riccardo Lencioni
    CardioVascular and Interventional Radiology, 2010, 33 : 818 - 827
  • [4] Tissue shrinkage in microwave ablation: ex vivo predictive model validation
    Farina, L.
    Nissenbaum, Y.
    Goldberg, S. N.
    Cavagnaro, M.
    2018 2ND URSI ATLANTIC RADIO SCIENCE MEETING (AT-RASC), 2018,
  • [5] Tissue shrinkage in microwave ablation of liver: an ex vivo predictive model
    Amabile, Claudio
    Farina, Laura
    Lopresto, Vanni
    Pinto, Rosanna
    Cassarino, Simone
    Tosoratti, Nevio
    Goldberg, S. Nahum
    Cavagnaro, Marta
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2017, 33 (01) : 101 - 109
  • [6] "We Are Not The Same": Comparison of Three Radiofrequency Thermal Devices Used for Saphenous Vein Ablation
    Kontopodis, Nikolaos
    V. Ioannou, Christos
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2023, 66 (01) : 102 - 102
  • [7] Ex vivo tissue shrinking in microwave thermal ablation
    Farina, L.
    Amabile, C.
    Nissenbaum, Y.
    Cavagnaro, M.
    Lopresto, V.
    Pinto, R.
    Tosoratti, N.
    Cassarino, S.
    Weiss, N.
    Goldberg, S. N.
    2015 9TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2015,
  • [8] Planar Microwave Sensor for Thermal Ablation of Organic Tissue
    Puentes, Margarita
    Bashir, Fahed
    Maasch, Matthias
    Schuessler, Martin
    Jakoby, Rolf
    2013 EUROPEAN MICROWAVE CONFERENCE (EUMC), 2013, : 479 - 482
  • [9] Comparison of Thermal Behaviour of Commercial Packages for Power Devices
    Giraudi, Lorenzo
    Tranchero, Maurizio
    Romano, Claudio
    Santero, Paolo
    2023 29TH INTERNATIONAL WORKSHOP ON THERMAL INVESTIGATIONS OF ICS AND SYSTEMS, THERMINIC, 2023,
  • [10] Evolution of a Microwave Instrument for Analysis and Thermal Ablation of Organic Tissue
    Puentes, Margarita
    Schuessler, Martin
    Damm, Christian
    Jakoby, Rolf
    2014 44TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2014, : 283 - 286