inferior vena cava;
in vitro model;
inferior vena cava filter;
D O I:
10.1097/00004424-199709000-00003
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
RATIONALE AND OBJECTIVES. The authors develop a physiologic in vitro model of the inferior vena cava (IVC) for testing of filters. METHODS. The model is driven by a centrifugal pump. A computer-controlled electromagnetic valve is used for generation of different flow patterns, Limitation of the pressure increase in case of IVC occlusion is achieved by a bypass circuit. A glycerin solution is used for perfusion. Artificial clots are made from polyacrylamide gel. Data acquisition includes continuous monitoring of flow and difference pressure over the filter and video recording of the testing events. RESULTS. The model can generate constant and pulsatile hows, The pressure increase can be limited to 70 mm Hg in case of occlusion. Calculation of the flow velocities in the IVC is possible. A classification of thrombus capturing is presented. The testing of most of the results are reproducible. CONCLUSIONS. The in vitro model simulates the physiologic conditions in the IVC. It can be used for comparative testing of different filters and the evaluation of new filter designs.
机构:
Northside Med Ctr, Western Reserve Care Syst NEOUCOM, Youngstown, OH 44505 USANorthside Med Ctr, Western Reserve Care Syst NEOUCOM, Youngstown, OH 44505 USA
Hashmi, Zubair A.
Smaroff, Gregory G.
论文数: 0引用数: 0
h-index: 0
机构:
Northside Med Ctr, Western Reserve Care Syst NEOUCOM, Youngstown, OH 44505 USANorthside Med Ctr, Western Reserve Care Syst NEOUCOM, Youngstown, OH 44505 USA
机构:
Massachusetts Gen Hosp, Inst Heart Vasc & Stroke Care, Boston, MA 02114 USAMassachusetts Gen Hosp, Inst Heart Vasc & Stroke Care, Boston, MA 02114 USA
Weinberg, Ido
Kaufman, John
论文数: 0引用数: 0
h-index: 0
机构:
Oregon Hlth & Sci Univ, Dotter Intervent Inst, Portland, OR 97201 USAMassachusetts Gen Hosp, Inst Heart Vasc & Stroke Care, Boston, MA 02114 USA
Kaufman, John
Jaff, Michael R.
论文数: 0引用数: 0
h-index: 0
机构:
Massachusetts Gen Hosp, Inst Heart Vasc & Stroke Care, Boston, MA 02114 USAMassachusetts Gen Hosp, Inst Heart Vasc & Stroke Care, Boston, MA 02114 USA