A physiologic in vitro model of the inferior vena cava with a computer-controlled flow system for testing of inferior vena cava filters

被引:12
|
作者
Jaeger, HJ
Mair, T
Geller, M
Kinne, RK
Christmann, A
Mathias, KD
机构
[1] SOC BIOMED RES DORTMUND,DORTMUND,GERMANY
[2] FACHHSCH DORTMUND,INST ENGN MECH,DORTMUND,GERMANY
[3] MAX PLANCK INST MOL PHYSIOL,D-44139 DORTMUND,GERMANY
[4] UNIV DORTMUND,CTR COMP,DORTMUND,GERMANY
关键词
inferior vena cava; in vitro model; inferior vena cava filter;
D O I
10.1097/00004424-199709000-00003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. The authors develop a physiologic in vitro model of the inferior vena cava (IVC) for testing of filters. METHODS. The model is driven by a centrifugal pump. A computer-controlled electromagnetic valve is used for generation of different flow patterns, Limitation of the pressure increase in case of IVC occlusion is achieved by a bypass circuit. A glycerin solution is used for perfusion. Artificial clots are made from polyacrylamide gel. Data acquisition includes continuous monitoring of flow and difference pressure over the filter and video recording of the testing events. RESULTS. The model can generate constant and pulsatile hows, The pressure increase can be limited to 70 mm Hg in case of occlusion. Calculation of the flow velocities in the IVC is possible. A classification of thrombus capturing is presented. The testing of most of the results are reproducible. CONCLUSIONS. The in vitro model simulates the physiologic conditions in the IVC. It can be used for comparative testing of different filters and the evaluation of new filter designs.
引用
收藏
页码:511 / 522
页数:12
相关论文
共 50 条
  • [1] In vitro model of the vena cava with a computer-controlled flow system for testing of vena cava filters
    Jaeger, HJ
    Geller, M
    Mair, T
    Bongert, M
    Kinne, R
    Mathias, KD
    RADIOLOGY, 1996, 201 : 324 - 324
  • [2] Dual inferior vena cava: Two inferior vena cava filters
    Hashmi, Zubair A.
    Smaroff, Gregory G.
    ANNALS OF THORACIC SURGERY, 2007, 84 (02): : 661 - 663
  • [3] Vena cava filters and inferior vena cava thrombosis
    Corriere, Matthew A.
    Suave, Kenneth J.
    Ayerdi, Juan
    Craven, Brandon L.
    Stafford, Jeanette M.
    Geary, Randolph L.
    Edwards, Matthew S.
    JOURNAL OF VASCULAR SURGERY, 2007, 45 (04) : 789 - 794
  • [4] Inferior Vena Cava Filters
    Nicolaides, A.
    Fareed, J.
    Kakkar, A. K.
    Comerota, A. J.
    Goldhaber, S. Z.
    Hull, R.
    Myers, K.
    Samama, M.
    Fletcher, J.
    Kalodiki, E.
    Bergqvist, D.
    Bonnar, J.
    Caprini, J. A.
    Carter, C.
    Conard, J.
    Eklof, B.
    Elalamy, I.
    Gerotziafas, G.
    Geroulakos, G.
    Giannoukas, A.
    Greer, I.
    Griffin, M.
    Kakkos, S.
    Lassen, M. R.
    Lowe, G. D. O.
    Markel, A.
    Prandoni, P.
    Raskob, G.
    Spyropoulos, A. C.
    Turpie, A. G.
    Walenga, J. M.
    Warwick, D.
    CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2013, 19 (02) : 204 - 205
  • [5] Inferior vena cava filters
    Kennedy, Sean A.
    Baerlocher, Mark O.
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 2015, 187 (16) : 1234 - 1234
  • [6] Inferior vena cava filters
    Bergqvist, David
    WORLD JOURNAL OF SURGERY, 2007, 31 (02) : 265 - 266
  • [7] Inferior Vena Cava Filters
    David Bergqvist
    World Journal of Surgery, 2007, 31 : 265 - 266
  • [8] INFERIOR VENA CAVA FILTERS
    不详
    INTERNATIONAL ANGIOLOGY, 2013, 32 (02) : 223 - 224
  • [9] Inferior vena cava filters
    Duffett, L.
    Carrier, M.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2017, 15 (01) : 3 - 12
  • [10] Inferior Vena Cava Filters
    Weinberg, Ido
    Kaufman, John
    Jaff, Michael R.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2013, 6 (06) : 539 - 547