Rapid and highly sensitive determination of low-molecular-weight carbonyl compounds in drinking water and natural water by preconcentration HPLC with 2,4-dinitrophenylhydrazine

被引:34
|
作者
Takeda, Kazuhiko [1 ]
Katoh, Shinya [1 ]
Nakatani, Nobutake [1 ]
Sakugawa, Hiroshi [1 ]
机构
[1] Hiroshima Univ, Grad Sch Biopsphere Sci, Hiroshima 7398521, Japan
关键词
D O I
10.2116/analsci.22.1509
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The aim of this research was to develop a simple procedure for a highly sensitive determination of low-molecular-weight (LMW) carbonyl compounds in drinking water and natural water. We employed a preconcentration HPLC system with 2,4-dinitrophenylhydrazine (DNPH) for the determination of LMW carbonyl compounds. A C-18 reverse-phase preconcentration column was used instead of a sample loop at the sample injection valve. A 0.1 - 5.0 mL portion of the derivatized sample solution was injected with a gas-tight syringe, and a 15% acetonitrile aqueous solution was pushed through the preconcentration column to remove the unreacted excess DNPH, which caused serious interference in the determination of formaldehyde. The detection limits were 1 - 3 nM with a relative standard deviation of 2 - 5% for 20 nM standard solutions (n = 5). The calibration curves were essentially unaffected by coexisting sea salts. Applications to commercial mineral water, tap water, river water, pond water and seawater are presented.
引用
收藏
页码:1509 / 1514
页数:6
相关论文