MIM, FEM and experimental investigations of the thermal drift in an ultra-high precision set-up for dimensional metrology at the nanometre accuracy level

被引:3
|
作者
Bouderbala, Kamelia [1 ,2 ]
Nouira, Hichem [1 ]
Videcoq, Etienne [2 ]
Girault, Manuel [2 ]
Petit, Daniel [2 ]
机构
[1] Lab Natl Metrol & Essais LNE, Lab Commun Metrol LNE CNAM, 1 Rue Gaston Boissier, F-75015 Paris, France
[2] CNRS ENSMA Univ Poitiers, Inst Pprime, Dept Fluides, Therm,Combust ENSMA, Teleport 2,1 Ave Clement Ader,BP 40109, F-86961 Chasseneuil, France
关键词
Modal Identification Method (MIM); Finite Element Model (FEM); Thermal behaviour and drift; Dimensional metrology; Measurement; Evaluation;
D O I
10.1016/j.applthermaleng.2015.09.092
中图分类号
O414.1 [热力学];
学科分类号
摘要
Thermal drifts caused by the power dissipated by the mechanical guiding systems constitute the main limit to enhance the accuracy of an ultra-high precision cylindricity measuring machine. For this reason, a high precision compact prototype has been designed to simulate the behaviour of the instrument. It ensures in situ calibration of four capacitive displacement probes by comparison with four laser interferometers. The set-up includes three heating wires for simulating the powers dissipated by the mechanical guiding systems, four additional heating wires located between each laser interferometer head and its respective holder, 19 Platinum resistance thermometers (Pt100) to observe the temperature evolution inside the set-up and four Pt100 sensors to monitor the ambient temperature. Both a Reduced Model (RM), based on the Modal Identification Method (MIM), and a Finite Element Model (FEM) using COMSOL software were developed and optimized by comparison with the experimental results. Thereafter, time dependent tests were performed under several conditions to measure the temperature variation at 19 fixed positions in the system and compared to the simulated FEM and calculated RM results. Even though the numerical results show good agreement with experiment, the RM results reproduce the temperature variations far better than those of the FEM, revealing the importance of the RM proposed for the evaluation of the thermal behaviour of the system. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:491 / 504
页数:14
相关论文
共 5 条
  • [1] FEM and experimental investigation of the thermal drift in ultra-high precision measuring machines for dimensional metrology
    Bouderbala, K.
    Nouira, H.
    [J]. MEASUREMENT, 2016, 90 : 250 - 264
  • [2] Implementation of capacitive probes for ultra-high precision machine for cylindricity measurement with nanometre level of accuracy
    Alain, Vissiere
    Hichem, Nouira
    Mohamed, Damak
    Olivier, Gibaru
    [J]. INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, 2015, 16 (05) : 883 - 893
  • [3] Implementation of capacitive probes for ultra-high precision machine for cylindricity measurement with nanometre level of accuracy
    Vissiere Alain
    Nouira Hichem
    Damak Mohamed
    Gibaru Olivier
    [J]. International Journal of Precision Engineering and Manufacturing, 2015, 16 : 883 - 893
  • [4] Experimental thermal regulation of an ultra-high precision metrology system by combining Modal Identification Method and Model Predictive Control
    Bouderbala, K.
    Nouira, H.
    Girault, M.
    Videcoq, E.
    [J]. APPLIED THERMAL ENGINEERING, 2016, 104 : 504 - 515
  • [5] Experimental set-up for testing MJ photovoltaic cells under ultra-high irradiance levels with temperature and spectrum control
    Nieto-Nieto, Luis M.
    Ferrer-Rodriguez, Juan P.
    Munoz-Ceron, Emilio
    Perez-Higueras, Pedro
    [J]. MEASUREMENT, 2020, 165 (165)