A comparison of electric field-strength standards for the frequency range of 30-1000 MHz

被引:0
|
作者
Garn, H
机构
关键词
antenna calibration; electric field-strength standards; electromagnetic compatibility; TEM-cell;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Published results about site attenuation and transmission loss measurements as well as antenna factor calculations and calibrations are summarized, The antenna types discussed are precision dipoles, a microstrip patch antenna, open-ended waveguides, and horn antennas as well as TEM-cells, The best values for agreement between theory and experiments that were reported for the transmission losses were in the range of 0.25-0.4 dB, An experimental comparison between precision reference dipoles and a TEM-cell that was performed in the Seibersdorf calibration laboratory is described with deviations of no more than 0.15 dB in the antenna factor or field-strength.