The effect of force on laser fiber burnback during lithotripsy

被引:1
|
作者
Aryaei, Ashkan [1 ]
Chia, Ray [1 ]
Peng, Steven [1 ]
机构
[1] Boston Sci Corp, 150 Baytech Dr, San Jose, CA 95134 USA
关键词
Holmium; Ho:YAG; laser; optical fiber; force; lithotripsy; burnback; fragmentation; stone ablation; CALCULI;
D O I
10.1117/12.2286285
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Optical fibers for lithotripsy are designed to deliver the maximum energy precisely to the treatment site without a decrease in performance and without increasing the risks to patients and users. One of the obstacles to constant energy delivery is burnback of the optical fiber tip. So far, researchers identified mechanical, thermal, and optical factors as mechanisms in burnback phenomena. Among mechanical factors, the force applied by urologists against a stone is expected to play a dominant role in burnback. In this study, we introduce a novel technique to measure accurately the stone depth and volume ablation under varying force. Our results show varying burnback lengths on the optical fibers and varying stone depth and volume ablation depending on the optical fiber core size. For instance, the slope of the burnback as a function of the applied force for 273 mu m fibers was more than two times higher than for the 550 mu m fibers. The slope of the total volume of stone ablated as function of force for 550 mu m fibers was almost twice as much as for the 273 mu m fibers. The data suggest urologists can maximize the stone ablation rate and minimize fiber tip burnback by controlling the applied force on the optical fiber during a lithotripsy procedure.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Fiber optic muzzle brake tip for reducing fiber burnback and stone retropulsion during thulium fiber laser lithotripsy
    Hutchens, Thomas C.
    Gonzalez, David A.
    Irby, Pierce B.
    Fried, Nathaniel M.
    JOURNAL OF BIOMEDICAL OPTICS, 2017, 22 (01)
  • [2] Burnback: the role of pulse duration and energy on fiber-tip degradation during high-power laser lithotripsy
    Brandon A. Levin
    Ali H. Aldoukhi
    Kristian M. Black
    Timothy L. Hall
    William W. Roberts
    Khurshid R. Ghani
    Lasers in Medical Science, 2021, 36 : 1817 - 1822
  • [3] Burnback: the role of pulse duration and energy on fiber-tip degradation during high-power laser lithotripsy
    Levin, Brandon A.
    Aldoukhi, Ali H.
    Black, Kristian M.
    Hall, Timothy L.
    Roberts, William W.
    Ghani, Khurshid R.
    LASERS IN MEDICAL SCIENCE, 2021, 36 (09) : 1817 - 1822
  • [4] Cavitation Bubble Dynamics during Thulium Fiber Laser Lithotripsy
    Hardy, Luke A.
    Kennedy, Joshua D.
    Wilson, Christopher R.
    Irby, Pierce B.
    Fried, Nathaniel M.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS XII, 2016, 9689
  • [5] MECHANISMS OF FIBER-TIP DEGRADATION DURING LASER LITHOTRIPSY
    Pishchalnikov, Yuri
    Behnke-Parks, William
    Stoller, Marshall
    JOURNAL OF UROLOGY, 2023, 209 : E473 - E473
  • [6] Thulium Fiber Laser Lithotripsy
    Scott, Nicholas J.
    Cilip, Christopher M.
    Fried, Nathaniel M.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS V, 2009, 7161
  • [7] Fiber-based Fluorescence and Reflection Measurements during Laser Lithotripsy
    Krumm, L. S.
    Lange, B.
    Ozimek, T.
    Wiessmeyer, J. R.
    Kramer, M. W.
    Merseburger, A. S.
    Brinkmann, R.
    TRANSLATIONAL BIOPHOTONICS: DIAGNOSTICS AND THERAPEUTICS, 2021, 11919
  • [8] Fiber Optic Suctioning of Urinary Stone Phantoms During Laser Lithotripsy
    Blackmon, Richard L.
    Case, Jason R.
    Trammell, Susan R.
    Irby, Pierce B.
    Fried, Nathaniel M.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS IX, 2013, 8565
  • [9] THULIUM FIBER LASER RESULTS IN SHORTER OPERATING TIMES DURING URETEROSCOPY AND LASER LITHOTRIPSY
    Ryan, James R.
    Nguyen, Mitchell H.
    Linscott, Joshua A.
    Nowicki, Samuel W.
    Jumper, Brian M.
    Ingimarsson, Johann P.
    JOURNAL OF UROLOGY, 2021, 206 : E925 - E925
  • [10] Thermal effect of holmium laser during ureteroscopic lithotripsy
    Hui Liang
    Lijian Liang
    Yin Yu
    Bin Huang
    Jia’nan Chen
    Chaoguo Wang
    Zhangguo Zhu
    Xiaozhong Liang
    BMC Urology, 20