Potentiometric flow injection determination of serum bromide in patients with epilepsy

被引:13
|
作者
Katsu, T [1 ]
Mori, Y [1 ]
Matsuka, N [1 ]
Gomita, Y [1 ]
机构
[1] OKAYAMA UNIV,SCH MED,DEPT HOSP PHARM,OKAYAMA 700,JAPAN
关键词
flow injection analysis; ion-selective electrode; bromide; therapeutic drug monitoring; epilepsy;
D O I
10.1016/S0731-7085(96)02017-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection system was constructed using a bromide-selective electrode and used to determine serum bromide in patients with epilepsy. A 10-mu l serum sample was injected into a carrier stream flowing at 0.12 mi min(-1). Potential changes and bromide concentrations were linearly related in the range 3-50 mM. The lower limit of detection for serum bromide was 1 mM and this electrode sensitivity spanned the entire concentration range required for bromide therapy (9-24 mM). The results compared favourably with those obtained by colorimetry. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:1829 / 1832
页数:4