Removal of the protruding retention wire via a femoral approach while leaving the active-fixation atrial ''J'' lead in situ: A technique for the management of class III atrial ''J'' leads

被引:0
|
作者
Ganz, LI
Meyerovitz, MF
Kandarpa, K
Mitchell, GF
机构
[1] BRIGHAM & WOMENS HOSP, DIV CARDIOL, BOSTON, MA 02115 USA
[2] BRIGHAM & WOMENS HOSP, DEPT RADIOL, BOSTON, MA 02115 USA
[3] HARVARD UNIV, SCH MED, BOSTON, MA USA
来源
关键词
retention wire fracture; lead extraction;
D O I
10.1111/j.1540-8159.1996.tb03166.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Telectronics Accufix Atrial ''J'' pacing lead poses a mechanical risk to patients of retention wire fracture and protrusion. Standard lead extraction techniques include percutaneous approaches, which are associated with significantly risk of morbidity and mortality, and open procedures, which necessitate thoracotomy. In nine patients referred with Class III retention wire fractures, attempts were made to snare the protruding retention wire from a femoral approach using snare devices and bioptomes, In six cases, the retention wire was successfully removed, leaving the lead body in place. In four patients with lead function that was able to be evaluated, the atrial lend remained functional after the procedure. There were no complications. Snare removal of the protruding retention wire via a femoral approach should be considered as an option in the management of patients with Class III Accufix leads.
引用
收藏
页码:1508 / 1512
页数:5
相关论文
共 1 条
  • [1] REMOVAL OF THE PROTRUDING RETENTION WIRE VIA A FEMORAL APPROACH WHILE LEAVING THE ACCUFIX ATRIAL LEAD IN-SITU
    GANZ, LI
    MEYEROVITZ, MF
    KANDARPA, K
    MITCHELL, GF
    CIRCULATION, 1995, 92 (08) : 713 - 713