Feasibility and safety of using local anaesthesia with conscious sedation during complex cardiac implantable electronic device procedures

被引:12
|
作者
Kaya, Elif [1 ]
Suedkamp, Hendrik [1 ]
Lortz, Julia [1 ]
Rassaf, Tienush [1 ]
Janosi, Rolf Alexander [1 ]
机构
[1] Univ Duisburg Essen, West German Heart & Vasc Ctr Essen, Dept Cardiol & Vasc Med, Essen, Germany
来源
SCIENTIFIC REPORTS | 2018年 / 8卷
关键词
GENERAL-ANESTHESIA; HEART-FAILURE; ANALGESIA; MIDAZOLAM;
D O I
10.1038/s41598-018-25457-x
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
We assessed the feasibility and safety of using local anaesthesia with conscious sedation as an alternative to general anaesthesia during complex and noncomplex cardiac implantable device procedures. We enrolled 279 consecutive patients who underwent cardiac device implantation/replacement at our tertiary/quaternary cardiac specialist hospital during a 17-month study period. Continuous combined intravenous conscious sedation with propofol and midazolam plus fentanyl and local anaesthesia were used for all procedures. Among the patients, 113, 59, 43, and 64 patients underwent pacemaker implantation, implantable cardiac defibrillator implantation, cardiac resynchronisation therapy device implantation, and generator exchange, respectively. The procedural success rate was 100%, with no apnoea or hypoxia episodes requiring therapeutic intervention. None of the patients required conversion to general anaesthesia. The mean surgical duration was longer for complex vs. noncomplex procedures (p = 0.003). The minimum mean arterial pressure during complex procedures was slightly lower than that during noncomplex procedures (p = 0.03). The perioperative (<24 h) mortality rate was 0%, and neither complexity group required tracheal intubation. Only two patients (0.7%) required unplanned intensive care unit admission for further surveillance. Our findings suggest that local anaesthesia with conscious sedation is a safe and feasible option for cardiac device implantation procedures, including complex procedures.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Feasibility and safety of using local anaesthesia with conscious sedation during complex cardiac implantable electronic device procedures
    Elif Kaya
    Hendrik Südkamp
    Julia Lortz
    Tienush Rassaf
    Rolf Alexander Jánosi
    [J]. Scientific Reports, 8
  • [2] Safety and feasibility study of bronchoscopic cryobiopsy under conscious sedation and local anaesthesia without fluoroscopy guidance
    Guliani, Abhinav
    Nawani, Sumeet
    Basu, Arup
    Dhamija, Amit
    [J]. RESPIROLOGY, 2023, 28 : 376 - 376
  • [3] The Feasibility, Safety, and Value of a Virtual Cardiac Implantable Electronic Device Wound Check
    Marzec, Lucas N.
    Guigli, Rebecca
    Hamer, Mika
    Sauer, William H.
    Meyer, Martha
    Baringer, Connor
    Gritz, Mark
    Ho, P. Michael
    Sandy, Lisa
    Matlock, Daniel
    Allen, Larry A.
    [J]. CIRCULATION, 2018, 138
  • [4] High flow nasal oxygen during procedural sedation for cardiac implantable electronic device procedures A randomised controlled trial
    Conway, Aaron
    Collins, Peter
    Chang, Kristina
    Kamboj, Navpreet
    Filici, Ana Lopez
    Lam, Phoebe
    Parotto, Matteo
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2021, 38 (08) : 839 - 849
  • [5] Cardiac Implantable Electronic Device Safety during Magnetic Resonance Imaging
    Hwang, You Mi
    Kim, Jun
    Lee, Ji Hyun
    Kim, Minsu
    Nam, Gi-Byoung
    Choi, Kee-Joon
    Kim, You-Ho
    [J]. KOREAN CIRCULATION JOURNAL, 2016, 46 (06) : 804 - 810
  • [6] Mortality Following Electrophysiology and Cardiac Implantable Electronic Device Procedures
    Lee, Justin Z.
    Ling, Jayna
    Diehl, Nancy N.
    Hodge, David O.
    Padmanabhan, Deepak
    Killu, Ammar M.
    Madhavan, Malini
    Noseworthy, Peter A.
    Kapa, Suraj
    McLeod, Christopher J.
    Cha, Yong-Mei
    Deshmukh, Abhishek J.
    Srivathsan, Komandoor
    Kusumoto, Fred M.
    Shen, Win K.
    Friedman, Paul A.
    Munger, Thomas M.
    Asirvatham, Samuel J.
    Packer, Douglas L.
    Mulpuru, Siva K.
    [J]. CIRCULATION, 2017, 136
  • [7] Safety of propofol for conscious sedation during endoscopic procedures: A prospective study
    Heuss, LT
    Schnieper, P
    Drewe, J
    Pflimlin, E
    Beglinger, C
    [J]. GASTROINTESTINAL ENDOSCOPY, 2002, 55 (05) : AB144 - AB144
  • [8] Remifentanil conscious sedation during regional anaesthesia for carotid endarterectomy: rationale and safety
    Marrocco-Trischitta, MM
    Bandiera, G
    Camilli, S
    Stillo, F
    Cirielli, C
    Guerrini, P
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2001, 22 (05) : 405 - 409
  • [9] A randomised, double-blind, controlled trial on patients undergoing cardiac implantable electronic device implantation with local anaesthesia and sedation with and without pectoral nerve block
    Wang, Y.
    Chandran, R.
    Woon, K. L.
    Chang, S. S.
    Quek, H. Y.
    Singh, P. A.
    Tan, P. T.
    Yeo, C.
    Wong, K.
    Tong, K. L.
    Tan, V. H.
    [J]. EUROPEAN HEART JOURNAL, 2020, 41 : 776 - 776
  • [10] Conscious Sedation and Local Anesthesia for Patients Undergoing Neurotologic and Complex Otologic Procedures
    Svrakic, Maja
    Pollack, Aron
    Huncke, T. Kate
    Roland, J. Thomas, Jr.
    [J]. OTOLOGY & NEUROTOLOGY, 2014, 35 (10) : E277 - E285