Performance Characteristics of the Whole-Body Discovery IQ PET/CT System

被引:60
|
作者
Reynes-Llompart, Gabriel [1 ]
Gamez-Cenzano, Cristina [1 ]
Romero-Zayas, Inmaculada [1 ]
Rodriguez-Bel, Laura [1 ]
Vercher-Conejero, Jose L. [1 ]
Marti-Climent, Josep M. [2 ]
机构
[1] Hosp Univ Bellvitge, IDIBELL, IDI, PET Unit,Nucl Med Dept, Barcelona, Spain
[2] Clin Univ Navarra, Nucl Med Dept, Pamplona, Spain
关键词
PET/CT; Discovery IQ; NEMA NU 2-2012; Q.Clear;
D O I
10.2967/jnumed.116.185561
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to assess the physical performance of a new PET/CT system, the Discovery IQ with 5-ring detector blocks. Methods: Performance was measured using the National Electrical Manufacturers Association NU2-2012 methodology. Image quality was extended by accounting for different acquisition parameters (lesion-to-background ratios [8:1, 4:1, and 2:1] and acquisition times) and reconstruction algorithms (VUE-point HD [VPHD], VPHD with point-spread-function modeling [VPHD-S], and Q.Clear). Tomographic reconstruction was also assessed using a Jaszczak phantom. Additionally, 30 patient lesions were analyzed to account for differences in lesion volume and SUV quantification between reconstruction algorithms. Results: Spatial resolution ranged from 4.2 mm at 1 cm to 8.5 mm at 20 cm. Sensitivity measured at the center and at 10 cm was 22.8 and 20.4 kps/kBq, respectively. The noise-equivalent counting rate peak was 124 kcps at 9.1 kBq/cm(3). The scatter fraction was 36.2%. The accuracy of correction for count losses and randoms was 3.9%. In the image quality test, contrast recovery for VPHD, VPHD-S, and Q.Clear ranged from 18%, 18%, and 13%, respectively (hot contrast; 10-mm sphere diameter; ratio, 2:1), to 68%, 67%, and 81%, respectively (cold contrast; 37-mm sphere diameter; ratio, 8:1). Background variability ranged from 3.4%, 3.0%, and 2.1%, respectively (ratio, 2: 1), to 5.5%, 4.8%, and 3.7%, respectively (ratio, 8: 1). On Q.Clear reconstruction, the decrease in the penalty term (b) increased the contrast recovery coefficients and background variability. With the Jaszczak phantom, image quality increased overall when a reconstruction algorithm modeling the point-spread function was used, and use of Q.Clear increased the signal-to-noise ratio. Lesions nalyzed using VPHD-S and Q.Clear had an SUVmean of 6.5 +/- 3 and 7 6 3, respectively (P<0.01), and an SUVmax of 11 +/- 4.8 and 12 +/- 4, respectively (P<0.01). No significant difference in mean lesion volume was found between algorithms. Conclusion: Among the various Discovery bismuth germanium oxide-based PET/CT scanners, the IQ with 5-ring detector blocks has the highest overall performance, with improved sensitivity and counting rate performance. Q.Clear reconstruction improves the PET image quality, with higher recovery coefficients and lower background variability.
引用
收藏
页码:1155 / 1161
页数:7
相关论文
共 50 条
  • [1] Performance evaluation of the new whole-body PET/CT scanner: Discovery ST
    Bettinardi, V
    Danna, M
    Savi, A
    Lecchi, M
    Castiglioni, I
    Gilardi, MC
    Bammer, H
    Lucignani, G
    Fazio, F
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (06) : 867 - 881
  • [2] Performance evaluation of the new whole-body PET/CT scanner: Discovery ST
    Valentino Bettinardi
    Massimo Danna
    Annarita Savi
    Michela Lecchi
    Isabella Castiglioni
    Maria Carla Gilardi
    Helmut Bammer
    Giovanni Lucignani
    Ferruccio Fazio
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2004, 31 : 867 - 881
  • [3] Performance of the new generation of whole-body PET/CT scanners: Discovery STE and Discovery VCT
    M. Teräs
    T. Tolvanen
    J. J. Johansson
    J. J. Williams
    J. Knuuti
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2007, 34 : 1683 - 1692
  • [4] Performance of the new generation of whole-body PET/CT scanners:: Discovery STE and discovery VCT
    Teraes, M.
    Tolvanen, T.
    Johansson, J. J.
    Williams, J. J.
    Knuuti, J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (10) : 1683 - 1692
  • [5] PERFORMANCE-CHARACTERISTICS OF A WHOLE-BODY PET SCANNER
    DEGRADO, TR
    TURKINGTON, TG
    WILLIAMS, JJ
    STEARNS, CW
    HOFFMAN, JM
    COLEMAN, RE
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1994, 35 (08) : 1398 - 1406
  • [6] Whole-body PET angiography on semiconductor PET/CT
    Norikane, Takashi
    Yamamoto, Yuka
    Takami, Yasukage
    Mitamura, Katsuya
    Matsumoto, Keisuke
    Maeda, Yukito
    Yamashita, Yoichi
    Nishiyama, Yoshihiro
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2022, 29 (02) : 885 - 888
  • [7] Whole-body PET angiography on semiconductor PET/CT
    Takashi Norikane
    Yuka Yamamoto
    Yasukage Takami
    Katsuya Mitamura
    Keisuke Matsumoto
    Yukito Maeda
    Yoichi Yamashita
    Yoshihiro Nishiyama
    [J]. Journal of Nuclear Cardiology, 2022, 29 : 885 - 888
  • [8] Diagnostic accuracy of whole-body PET/MRI and whole-body PET/CT for TNM staging in oncology
    Philipp Heusch
    Felix Nensa
    Benedikt Schaarschmidt
    Rupika Sivanesapillai
    Karsten Beiderwellen
    Benedikt Gomez
    Jens Köhler
    Henning Reis
    Verena Ruhlmann
    Christian Buchbender
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2015, 42 : 42 - 48
  • [9] Diagnostic accuracy of whole-body PET/MRI and whole-body PET/CT for TNM staging in oncology
    Heusch, Philipp
    Nensa, Felix
    Schaarschmidt, Benedikt
    Sivanesapillai, Rupika
    Beiderwellen, Karsten
    Gomez, Benedikt
    Koehler, Jens
    Reis, Henning
    Ruhlmann, Verena
    Buchbender, Christian
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (01) : 42 - 48
  • [10] Clinical comparison of a whole-body PET-MR system with PET-CT
    Machac, Josef
    Izquierdo-Garcia, David
    Knesaurek, Karin
    Rafique, Ash
    Davies, Olufunke
    Fuster, Valentin
    Fayad, Zahi
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2012, 53