Monte Carlo simulated correction factors for output factor measurement with the CyberKnife system-results for new detectors and correction factor dependence on measurement distance and detector orientation

被引:62
|
作者
Francescon, P. [1 ]
Kilby, W. [2 ]
Satariano, N. [1 ]
机构
[1] ULSS6, Dept Med Phys, I-36100 Vicenza, Italy
[2] Accuray Inc, Sunnyvale, CA 94089 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 06期
关键词
small field dosimetry; Monte Carlo simulation; CyberKnife system; FIELDS;
D O I
10.1088/0031-9155/59/6/N11
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A previous study of the corrections needed for output factor measurements with the CyberKnife system has been extended to include new diode detectors (IBA SFD and Exradin D1V), an air filled microchamber (Exradin CC01) and a scintillation detector (Exradin W1). The dependence of the corrections on detector orientation (detector long axis parallel versus perpendicular to the beam axis) and source to detector distance (SDD) was evaluated for these new detectors and for those in our previous study. The new diodes are found to over-respond at the smallest (5 mm) field size by 2.5% (D1V) and 3.3% (SFD) at 800 mm SDD, while the CC01 under-responds by 7.4% at the same distance when oriented parallel to the beam. Corrections for all detectors tend to unity as field size increases. The W1 corrections are <0.5% at all field sizes. Microchamber correction factors increase substantially if the detector is oriented perpendicular to the beam (by up to 23% for the PTW 31014). Corrections also vary with SDD, with the largest variations seen for microchambers in the perpendicular orientation (up to 13% change at 650 mm SDD versus 800 mm) and smallest for diodes (similar to 1% change at 650 mm versus 800 mm). The smallest and most stable corrections are found for diodes, liquid filled microchambers and scintillation detectors, therefore these should be preferred for small field output factor measurements. If air filled microchambers are used, then the parallel orientation should be preferred to the perpendicular, and care should be taken to use corrections appropriate to the measurement SDD.
引用
收藏
页码:N11 / N17
页数:7
相关论文
共 20 条
  • [1] Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife system
    Francescon, P.
    Kilby, W.
    Satariano, N.
    Cora, S.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12): : 3741 - 3758
  • [2] Profile and Output Factor Measurement for a Cyberknife System Using Different Detectors
    Yang, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S971 - S971
  • [3] Small Fields Output Factors Measurement and Correction Factor Determination for Several Active and Passive Detectors
    Sharma, D.
    Sharma, S.
    Chowdhary, R.
    Pilakkal, S.
    Rashal, S.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3595 - 3595
  • [4] Uncertainty Evaluation of Colour Correction Factor Using Monte Carlo Method in LED Photometry Measurement
    Zhao Weiqiang
    Liu Hui
    Yan Jinyun
    Su Ying
    Jiang Lin
    [J]. OPTICAL METROLOGY AND INSPECTION FOR INDUSTRIAL APPLICATIONS VI, 2019, 11189
  • [5] Determination of small field output factors and correction factors using a Monte Carlo method for a 1000 MU/min CyberKnife® system equipped with fixed collimators
    Moignier, C.
    Huet, C.
    Barraux, V.
    Bassinet, C.
    Baumann, M.
    Sebe-Mercier, K.
    Loiseau, C.
    Batalla, A.
    Makovicka, L.
    [J]. RADIATION MEASUREMENTS, 2014, 71 : 287 - 292
  • [6] On the output factor measurements of the CyberKnife iris collimator small fields: Experimental determination of the kQclin,Qmsrfclin,fmsr correction factors for microchamber and diode detectors
    Pantelis, E.
    Moutsatsos, A.
    Zourari, K.
    Petrokokkinos, L.
    Sakelliou, L.
    Kilby, W.
    Antypas, C.
    Papagiannis, P.
    Karaiskos, P.
    Georgiou, E.
    Seimenis, I.
    [J]. MEDICAL PHYSICS, 2012, 39 (08) : 4875 - 4885
  • [7] Measurement and Monte Carlo calculation of the backscatter correction factor at the exit of high energy photon beams as a function of different parameters
    Koukorava, C
    Dimitriadis, A
    Theodorou, K
    Kappas, C
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S88 - S89
  • [8] Field Size Dependence of Output Factor for Uniform Scanning Proton Beams: A Comparison of TPS Calculation, Measurement and Monte Carlo Simulation
    Zheng, Y.
    Singh, H.
    Islam, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 362 - 362
  • [9] Longitudinal measurement of chromatic dispersion along an optical fiber transmission system with a new correction factor
    Abbasi, Madiha
    Baig, Mirza Imran
    Shaikh, Muhammad Shafique
    [J]. 1ST INTERNATIONAL CONFERENCE ON SENSING FOR INDUSTRY, CONTROL, COMMUNICATIONS, & SECURITY TECHNOLOGIES (ICSICCST 2013), 2013, 51
  • [10] Small field output factor measurement and verification for CyberKnife robotic radiotherapy and radiosurgery system using 3D polymer gel, ionization chamber, diode, diamond and scintillator detectors, Gafchromic film and Monte Carlo simulation
    Kafi, Md Abdullah Al
    Arib, Mehenna
    Al Moussa, Akram
    Alzorkany, Faisal
    Shehadeh, Mamoun
    Yusof, Mohd Fahmi Mohd
    Moftah, Belal
    [J]. APPLIED RADIATION AND ISOTOPES, 2023, 192