Improvement of metallic artifacts in computed tomography in the absence of artifact reduction algorithms for spinal treatment planning applications

被引:1
|
作者
Alinejad, Marjan [1 ]
Pourmoghaddas, Amir [2 ]
Amouheidari, Alireza [3 ]
Shokrani, Parvaneh [4 ]
机构
[1] Guilan Univ Med Sci, Dept Med Phys, Rasht, Iran
[2] St Jude Childrens Res Hosp, Dept Radiat Oncol, 332 N Lauderdale St, Memphis, TN 38105 USA
[3] Isfahan Univ Med Sci, Dept Oncol, Milad Hosp, Esfahan, Iran
[4] Isfahan Univ Med Sci, Dept Med Phys, Esfahan, Iran
关键词
Computed tomography image artifact; extended computed tomography scale; spinal metallic implant; treatment planning; CT; IMPLANTS;
D O I
10.4103/jcrt.JCRT_1446_16
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim of Study: The goal of this research was to investigate if application of optimized imaging parameters, recommended in literature, would be effective in producing the image quality required for treatment planning of spinal radiation fields with metallic implants. Materials and Methods: CT images from an anthropomorphic torso phantom with and without spinal implants were acquired using different imaging protocols: raising kVp and mAs, reducing the pitch and applying an extended CT scale (ECTS) technique. Profiles of CT number (CT#) were produced using DICOM data of each image. The effect of artifact on dose calculation accuracy was investigated using the image data in the absence of implant as a reference and the recommended electron density tolerance levels (Delta rho e). Results: Raising the kVp was the only method that produced improvement to some degree in CT# in artifact regions. Application of ECTS improved CT# values only for metal. Conclusions: Although raising the kVp was effective in reducing metallic artifact, the significance of this effect on Delta rho e values in corrected images depends on the required tolerance for treatment planning dose calculation accuracy. ECTS method was only successful in correcting the CT number range in the metal. Although, application of ECTS method did not have any effect on artifact regions, its use is necessary in order to improve delineation of metal and accuracy of attenuation calculations in metal, provided that the treatment planning system can use an extended CT# calibration curve. Also, for Monte Carlo calculations using patient's images, ECTS-post-processed-CT images improve dose calculation accuracy for impure metals.
引用
收藏
页码:878 / 883
页数:6
相关论文
共 46 条
  • [1] Evaluation of Metal Artifact Reduction Algorithms in Computed Tomography and Their Application to Radiation Therapy Treatment Planning
    Norris, H.
    Rangaraj, D.
    Kim, S.
    MEDICAL PHYSICS, 2016, 43 (06) : 3448 - 3448
  • [2] Iterative algorithms for artifact reduction in computed tomography
    Skornitzke, S.
    RADIOLOGE, 2018, 58 (03): : 202 - 210
  • [3] Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction
    Giantsoudi, Drosoula
    De Man, Bruno
    Verburg, Joost
    Trofimov, Alexei
    Jin, Yannan
    Wang, Ge
    Gjesteby, Lars
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08): : R49 - R80
  • [4] Reduction of metal artifacts caused by multiple metallic objects in computed tomography
    Wei Xing
    Yan Bin
    Zhang Feng
    Li Yong-Li
    Xi Xiao-Qi
    Li Lei
    ACTA PHYSICA SINICA, 2014, 63 (05) : 058702
  • [5] Metallic artifact reduction by evaluation of the additional value of iterative reconstruction algorithms in hip prosthesis computed tomography imaging
    Neroladaki, Angeliki
    Martin, Steve Philippe
    Bagetakos, Ilias
    Botsikas, Diomidis
    Hamard, Marion
    Montet, Xavier
    Boudabbous, Sana
    MEDICINE, 2019, 98 (06)
  • [6] Reduction of Artifacts Caused by Deep Brain Stimulating Electrodes in Cranial Computed Tomography Imaging by Means of Virtual Monoenergetic Images, Metal Artifact Reduction Algorithms, and Their Combination
    Hokamp, Nils Grosse
    Hellerbach, Alexandra
    Gierich, Andreas
    Jordan, David W.
    Visser-Vandewalle, Veerle
    Maintz, David
    Haneder, Stefan
    INVESTIGATIVE RADIOLOGY, 2018, 53 (07) : 424 - 431
  • [7] Computed tomography imaging of periocular metallic foreign bodies can be improved by artifact reduction software
    Willett, Keirnan
    Sheng, Mike
    Woo, John
    Gorry, Thomasine
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [8] Clinical applications of positron emission tomography/computed tomography treatment planning
    Macapinlac, Homer A.
    SEMINARS IN NUCLEAR MEDICINE, 2008, 38 (02) : 137 - 140
  • [9] An adaptive approach to metal artifact reduction in helical computed tomography for radiation therapy treatment planning: Experimental and clinical studies
    Yazdia, M
    Gingras, L
    Beaulieu, L
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04): : 1224 - 1231
  • [10] Applications of Dual-Energy Computed Tomography for Artifact Reduction in the Head, Neck, and Spine
    Liao, Eric
    Srinivasan, Ashok
    NEUROIMAGING CLINICS OF NORTH AMERICA, 2017, 27 (03) : 489 - +