Determination of methotrexate in serum by high-performance liquid chromatography

被引:34
|
作者
Aboleneen, H
Simpson, J
Backes, D
机构
[1] Diagnostics Division, Abbott Laboratories, Abbott Park
来源
关键词
methotrexate;
D O I
10.1016/0378-4347(95)00580-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A sensitive high-performance Liquid chromatographic method with ultraviolet detection was developed to quantitate methotrexate in serum-based calibrators, controls and patient samples. Sample clean-up was achieved with C-18 Sep-Pak Classic cartridges. The chromatographic separation was accomplished on a 5-mu m Ultrasphere ODS Beekman column. 8-Chlorotheophylline was used as an internal standard. The method was validated by recovery, linearity, accuracy and precision studies. Two standard curves were constructed to cover the high and the low ends of the calibrator range (0.05-1.0 mu mol/l). Response was found linear over the whole range of the calibrator set with a correlation coefficient of 0.999 and 1.00 for the low-level and the high-level curves, respectively. Accuracy varied from 12% at the lowest level to 1.2% at the highest level. The precision study showed a C.V. of 14.4% at the lowest level and 3.3% at the highest level.
引用
收藏
页码:317 / 322
页数:6