Influence of digital clubbing on oxygen saturation measurements by pulse-oximetry in cystic fibrosis patients

被引:11
|
作者
Van Ginderdeuren, Filip
Van Cauwelaert, Karlien
Malfroot, Anne
机构
[1] Free Univ Brussels, Akad Ziekenhuis, Dept Physiotherapy, B-1090 Brussels, Belgium
[2] Vrije Univ Brussels, Acad Ziekenhuis, Dept Paediat, Paediat Resp Med & Cyst Fibrosis Clin, Brussels, Belgium
关键词
digital clubbing; arterial oxygen saturation; pulse-oximetry (SpO(2)); cystic fibrosis;
D O I
10.1016/j.jcf.2006.01.007
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Objective: To determine the influence of digital clubbing on oxygen saturation by pulse oximetry measurements (SpO(2)) in Cystic Fibrosis patients. Background: Measuring the arterial oxygen saturation at the fingertip by pulse-oximetry is commonly used in the management of CF patients. In these patients, clinical signs of hyperoxia are often observed with oxygen supplies based on digital oximetry readings. This suggests inaccuracies in the digital measurement method, which in its turn may be caused by digital clubbing. In order to study the influence of digital clubbing, measurements between fingertip and forehead sensor were compared in a clubbing and non-clubbing CF-population. The ear sensor measurements are used as a reference variable. Methods: Two groups were examined. Group I consisted of 50 CF patients without digital clubbing (DPD/IPD ratio< 1.00). Group 2 consisted of 50 CF patients with digital clubbing (DPD / IPD ratio> 1.00). Patients were measured at rest before any treatment and with their daily oxygen supply, if applicable. Saturation was simultaneously measured with three Criticare SpO(2) T pulse oximeters, using a fingertip sensor at the right index (transmission oximetry), a forehead sensor at the forehead (reflectance oximetry) and an ear sensor at the right ear. Results: Using the Bland and Altman method no clear difference was found between the saturation measurements of right ear versus forehead sensor in the two groups. When the measurements of right ear versus fingertip sensor are compared there is still no difference for the non-clubbing group. On the contrary, for the clubbing group lower saturation scores were measured by the fingertip probe compared to the right ear measurement. The differences in saturation became greater as the saturation value at fingertip was lower. Conclusion: Digital clubbing significantly influences the registrations of the SpO(2) measurements by means of a fingertip probe, underestimating the saturation. It can be advised to use the ear sensor as good alternative for these patients. (C) 2006 European Cystic Fibrosis Society. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:125 / 128
页数:4
相关论文
共 50 条
  • [1] PULSE OXIMETRY, DIGITAL CLUBBING, AND CYSTIC-FIBROSIS
    SCHONI, MH
    HURLIMANN, R
    BRUDERER, K
    LANCET, 1987, 1 (8546): : 1371 - 1371
  • [2] THE EFFECT OF CARBOXYHEMOGLOBIN ON PULSE-OXIMETRY OXYGEN-SATURATION MEASUREMENTS
    PANACEK, EA
    JOURILLES, NJ
    SHERMAN, BW
    RUTHERFORD, WF
    CLINICAL RESEARCH, 1991, 39 (02): : A421 - A421
  • [3] Evaluation of pulse-oximetry oxygen saturation taken through skin protective covering
    James J.
    Tiwari L.
    Upadhyay P.
    Sreenivas V.
    Bhambhani V.
    Puliyel J.M.
    BMC Pediatrics, 6 (1)
  • [4] Racial Disparity in Oxygen Saturation Measurements by Pulse Oximetry
    Jamali, Haya
    Castillo, Lauren T.
    Morgan, Chelsea Cosby
    Coult, Jason
    Muhammad, Janice L.
    Osobamiro, Oyinkansola O.
    Parsons, Elizabeth C.
    Adamson, Rosemary
    ANNALS OF THE AMERICAN THORACIC SOCIETY, 2022, 19 (12) : 1951 - 1964
  • [5] Pulse oximetry oxyhemoglobin saturation in patients with idiopathic pulmonary fibrosis
    Bennett, David
    Pieroni, Maria Grazia
    Lucrezia, Pianigiani
    Darnell, Paolo
    TireIli, Francesca
    Fossi, Antonella
    Rottoli, Paola
    EUROPEAN RESPIRATORY JOURNAL, 2014, 44
  • [6] Reversal of digital clubbing after lung transplantation in cystic fibrosis patients: A clue to the pathogenesis of clubbing
    Augarten, A
    Goldman, R
    Laufer, J
    Szeinberg, A
    Efrati, O
    Barak, A
    Miller, MS
    Yahav, Y
    PEDIATRIC PULMONOLOGY, 2002, 34 (05) : 378 - 380
  • [8] VALIDATION OF ARTERIAL OXYGEN-SATURATION MEASUREMENTS INUTERO USING PULSE OXIMETRY
    HARRIS, AP
    SENDAK, MJ
    CHUNG, DC
    RICHARDSON, CA
    AMERICAN JOURNAL OF PERINATOLOGY, 1993, 10 (03) : 250 - 254
  • [9] DAYTIME PULSE-OXIMETRY MEASUREMENTS DO NOT PREDICT NOCTURNAL DESATURATIONS IN ADULT SICKLE CELL PATIENTS
    Kadali, R. A.
    Chohan, H. S.
    Efird, J.
    Boettger, P.
    Liles, D.
    Knupp, C.
    Judy, J. S.
    Sharma, S.
    SLEEP, 2011, 34 : A240 - A240
  • [10] Pulse Oximetry Overestimates Oxygen Saturation in COPD
    Amalakanti, Sridhar
    Pentakota, Mohan Rao
    RESPIRATORY CARE, 2016, 61 (04) : 423 - 427