Pacemaker-mediated tachycardia: Management by pacemaker interrogation/reprogramming in the ED

被引:4
|
作者
Sobel, RM
Donaldson, PR
Dhruva, N
机构
[1] Emory Univ, Sch Med, Dept Emergency Med, Atlanta, GA USA
[2] Cardiol Dis Specialists, Atlanta, GA USA
来源
AMERICAN JOURNAL OF EMERGENCY MEDICINE | 2002年 / 20卷 / 04期
关键词
pacemaker malfunction; pacemaker-mediated tachycardia; pacemaker interrogation and programming; atrial flutter; overdrive pacing; emergency department;
D O I
10.1053/ajem.2002.33780
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We report a case of a patient with symptomatic pacemaker-mediated tachycardia (PMT). PMT should be suspected when asynchronous pacing induced by magnet application causes temporary resolution of an unknown tachycardia in a patient with a pacemaker. The underlying mechanism of PMT is either reentrant tachycardia or atrial triggering. PMT is better discerned with an intracardiac electrocardiogram (ECG) obtained through pacemaker interrogation (PI) than the standard 12-lead ECG. In the case reported, the interrogator was used to demonstrate that the initiating dysrhythmia was atrial flutter. Next, the programmer was used to overdrive the atrium restoring normal sinus rhythm. After pacemaker reprogramming, the patient was discharged from the emergency department (ED) in improved condition. We believe this to be the first reported case of a patient with PMT diagnosed, treated, and discharged effectively after PI, overdrive pacing and reprogramming from the ED. The pacemaker interrogator/programmer is a rapid noninvasive device that has practical applications in the ED. Copyright 2002, Elsevier Science (USA). All rights reserved.
引用
收藏
页码:336 / 339
页数:4
相关论文
共 50 条
  • [1] Pacemaker tachycardia: Is it pacemaker-mediated tachycardia or sinus tachycardia?
    Barold, S. Serge
    Leonelli, Fabio
    Khan, Nadim
    Cutro, Raymond
    Herweg, Bengt
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 (02): : 256 - 257
  • [2] Pacemaker-mediated tachycardia
    Ullah, W.
    Stewart, A.
    HEART, 2010, 96 (13) : 1062 - 1062
  • [3] PACEMAKER-MEDIATED TACHYCARDIA - ENGINEERING SOLUTIONS
    CALFEE, RV
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11): : 1917 - 1928
  • [4] Pacemaker-Mediated Tachycardia Interpolated Into Ventricular Tachycardia
    Milicevic, Goran
    Galic, Edvard
    Udiljak, Nikola
    TEXAS HEART INSTITUTE JOURNAL, 2022, 49 (06)
  • [5] Intermittent Irregular Pacemaker-Mediated Tachycardia
    Gilmore, Mark
    Lim, Sern
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2015, 38 (09): : 1117 - 1120
  • [6] TERMINATION OF PACEMAKER-MEDIATED TACHYCARDIA BY ADENOSINE
    CONTI, JB
    CURTIS, AB
    HILL, JA
    RAYMENANTS, ER
    CLINICAL CARDIOLOGY, 1994, 17 (01) : 47 - 48
  • [7] Pacemaker-mediated polymorphic ventricular tachycardia
    Goldman, DS
    Levine, PA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10): : 1993 - 1995
  • [8] Pacemaker-Mediated Tachycardia: A Case Report
    Gill, Rajwinder
    Meghrajani, Vineet
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (07)
  • [9] Pacemaker-mediated tachycardia: What is the mechanism?
    Jastrzebski, Marek
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (11): : 1549 - 1551
  • [10] SENSOR-INITIATED TERMINATION OF PACEMAKER-MEDIATED TACHYCARDIA IN A DDDR PACEMAKER
    LAU, CP
    LI, JPS
    CHENG, CH
    WONG, CK
    CHUNG, FLW
    AMERICAN HEART JOURNAL, 1991, 121 (02) : 595 - 597