Monte Carlo simulation of TrueBeam flattening-filter-free beams using Varian phase-space files: Comparison with experimental data

被引:30
|
作者
Belosi, Maria F. [1 ]
Rodriguez, Miguel [2 ]
Fogliata, Antonella [1 ]
Cozzi, Luca [1 ]
Sempau, Josep [2 ,3 ]
Clivio, Alessandro [1 ]
Nicolini, Giorgia [1 ]
Vanetti, Eugenio [1 ]
Krauss, Harald [4 ]
Khamphan, Catherine [5 ]
Fenoglietto, Pascal [6 ]
Puxeu, Josep [7 ]
Fedele, David [8 ]
Mancosu, Pietro [9 ]
Brualla, Lorenzo [10 ]
机构
[1] Oncol Inst Southern Switzerland, Med Phys Unit, CH-6500 Bellinzona, Switzerland
[2] Univ Politecn Cataluna, Inst Tecn Energet, E-08028 Barcelona, Spain
[3] CIBER BBN, Spanish Networking Res Ctr, E-08028 Barcelona, Spain
[4] Inst Radioonkol, A-1100 Vienna, Austria
[5] Inst St Catherine, Med Phys Unit, F-84000 Avignon, France
[6] CRLC Val Aurelle Paul Lamarque, Dept Cancerol Radiotherapie, F-34090 Montpellier, France
[7] Inst Catala Oncol, Dept Med Phys, E-08028 Barcelona, Spain
[8] Casa Cura San Rossore, Radio Oncol Dept, I-56100 Pisa, Italy
[9] Humanitas Clin & Res Ctr, Dept Radiat Oncol, I-20089 Rozzano Milan, Italy
[10] Univ Klinikum Essen, Strahlenklin, NCTeam, D-45122 Essen, Germany
关键词
phase-space files; FFF beams; Monte Carlo vs measurements; DOSE CALCULATIONS; PHOTON; ELECTRON; PENELOPE; ALGORITHM; GEANT4;
D O I
10.1118/1.4871041
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Phase-space files for Monte Carlo simulation of the Varian TrueBeam beams have been made available by Varian. The aim of this study is to evaluate the accuracy of the distributed phase-space files for flattening filter free (FFF) beams, against experimental measurements from ten TrueBeam Linacs. Methods: The phase-space files have been used as input in PRIMO, a recently released Monte Carlo program based on the PENELOPE code. Simulations of 6 and 10 MV FFF were computed in a virtual water phantom for field sizes 3 Chi 3, 6 Chi 6, and 10 Chi 10 cm(2) using 1 Chi 1 Chi Chi 1 mm(3) voxels and for 20 Chi 20 and 40 Chi 40 cm(2) with 2 Chi 2 Chi 2 mm(3) voxels. The particles contained in the initial phase-space files were transported downstream to a plane just above the phantom surface, where a subsequent phase-space file was tallied. Particles were transported downstream this second phase-space file to the water phantom. Experimental data consisted of depth doses and profiles at five different depths acquired at SSD = 100 cm (seven datasets) and SSD = 90 cm (three datasets). Simulations and experimental data were compared in terms of dose difference. Gamma analysis was also performed using 1%, 1 mm and 2%, 2 mm criteria of dose-difference and distance-to-agreement, respectively. Additionally, the parameters characterizing the dose profiles of unflattened beams were evaluated for both measurements and simulations. Results: Analysis of depth dose curves showed that dose differences increased with increasing field size and depth; this effect might be partly motivated due to an underestimation of the primary beam energy used to compute the phase-space files. Average dose differences reached 1% for the largest field size. Lateral profiles presented dose differences well within 1% for fields up to 20 Chi 20 cm(2), while the discrepancy increased toward 2% in the 40 Chi 40 cm(2) cases. Gamma analysis resulted in an agreement of 100% when a 2%, 2 mm criterion was used, with the only exception of the 40 Chi 40 cm(2) field (similar to 95% agreement). With the more stringent criteria of 1%, 1 mm, the agreement reduced to almost 95% for field sizes up to 10 Chi 10 cm(2), worse for larger fields. Unflatness and slope FFF-specific parameters are in line with the possible energy underestimation of the simulated results relative to experimental data. Conclusions: The agreement between Monte Carlo simulations and experimental data proved that the evaluated Varian phase-space files for FFF beams from TrueBeam can be used as radiation sources for accurate Monte Carlo dose estimation, especially for field sizes up to 10 Chi 10 cm(2), that is the range of field sizes mostly used in combination to the FFF, high dose rate beams. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:11
相关论文
共 34 条
  • [1] Monte Carlo Simulation Using Varian Phase-Space Files and VirtuaLinac: A Comparison with TrueBeam Experimental Data
    Shine, N. S.
    Paramu, Raghukumar
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E619 - E619
  • [2] Flattening Filter Free phase space files: Comparison between Monte Carlo simulations and experimental data
    Belosi, M. F.
    Fogliata, A.
    Brualla, L.
    Sempau, J.
    Rodriguez, M.
    Clivio, A.
    Nicolini, G.
    Vanetti, E.
    Cozzi, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S147 - S147
  • [3] Monte Carlo simulation of beam characteristics from small fields based on TrueBeam flattening-filter-free mode
    Feng, Zhongsu
    Yue, Haizhen
    Zhang, Yibao
    Wu, Hao
    Cheng, Jinsheng
    Su, Xu
    [J]. RADIATION ONCOLOGY, 2016, 11
  • [4] Monte Carlo simulation of beam characteristics from small fields based on TrueBeam flattening-filter-free mode
    Zhongsu Feng
    Haizhen Yue
    Yibao Zhang
    Hao Wu
    Jinsheng Cheng
    Xu Su
    [J]. Radiation Oncology, 11
  • [5] A Monte Carlo simulation framework for electron beam dose calculations using Varian phase space files for TrueBeam Linacs
    Rodrigues, Anna
    Sawkey, Daren
    Yin, Fang-Fang
    Wu, Qiuwen
    [J]. MEDICAL PHYSICS, 2015, 42 (05) : 2389 - 2403
  • [6] Monte Carlo Simulation of a 6MV Varian Truebeam Without Flattening Filter Linac
    Song, T.
    Zhou, L.
    Jiang, S.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3819 - 3819
  • [7] Creation of a Monte Carlo TrueBeam Model by Reproducing Varian Phase Space Data
    O'Grady, K.
    Davis, S.
    Seuntjens, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3730 - 3730
  • [8] Flattening-Filter-Free Beam Quality Correction Factor Determination Using Experimental and Monte Carlo Methods
    Wood, M.
    Desai, V.
    Simiele, E.
    Taneja, S.
    DeWerd, L.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 469 - 469
  • [9] Validation of the Varian Generic Phase Space Files for Monte Carlo Calculations of Dose Distributions for the TrueBeam Linac Head
    Wang, L.
    Xing, L.
    Sawkey, D.
    Constantin, M.
    Svatos, M.
    Mok, E.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3820 - 3820
  • [10] Monitor unit verification for Varian TrueBeam VMAT plans using Monte Carlo calculations and phase space data
    Pant, Ankit
    Miri, Narges
    Bhagroo, Stephen
    Mathews, Joshua A.
    Nazareth, Daryl P.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (10):