Internal breast dosimetry in mammography: Experimental methods and Monte Carlo validation with a monoenergetic x-ray beam

被引:14
|
作者
Fedon, Christian [1 ,2 ]
Caballo, Marco [1 ]
Longo, Renata [2 ,3 ]
Trianni, Annalisa [4 ]
Sechopoulos, Ioannis [1 ,5 ]
机构
[1] Radboud Univ Nijmegen, Dept Radiol & Nucl Med, Med Ctr, POB 9101, NL-6500 HB Nijmegen, Netherlands
[2] Ist Nazl Fis Nucl, Sez Trieste, I-34127 Trieste, Italy
[3] Univ Trieste, Dipartimento Fis, I-34127 Trieste, Italy
[4] Presidio Osped S Maria della Misericordia, Dept Med Phys, Azienda Sanit Univ Integrata ASUIUD, Ple S Maria della Misericordia, I-33100 Udine, Italy
[5] Dutch Expert Ctr Screening LRCB, POB 6873, NL-6503 GJ Nijmegen, Netherlands
关键词
breast dosimetry; GafChromic film; GEANT4; MOSFET; TLD; LIF-MG; CU; P THERMOLUMINESCENCE DOSIMETERS; EFFECT TRANSISTOR DOSIMETERS; SYNCHROTRON-RADIATION; DIGITAL TOMOSYNTHESIS; COMPUTED-TOMOGRAPHY; DOSE-RESPONSE; FILMS; ENERGY; SIMULATION; DEPENDENCE;
D O I
10.1002/mp.12792
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo investigate the performance, such as energy dependence and sensitivity, of thermoluminescent dosimeters (TLD), metal oxide semiconductor field-effect transistor dosimeters (MOSFET), and GafChromic films, and to validate the estimates of local dose deposition of a Monte Carlo (MC) simulation for breast dosimetry applications. MethodsExperimental measurements were performed using a monoenergetic beam at the ELETTRA synchrotron radiation light source (Trieste, Italy). The three types of dosimeters were irradiated in a plane transversal to the beam axis and calibrated in terms of air kerma. The sensitivity of MOSFET dosimeters and GafChromic films was evaluated in the range of 18-28 keV. Three different calibration curves for the GafChromic films were tested (logarithmic, rational, and exponential functions) to evaluate the best-fit curve in the dose range of 1-20 mGy. Internal phantom dose measurements were performed at 20 keV for four different depths (range 0-3 cm, with 1 cm steps) using a homogeneous 50% glandular breast phantom. A GEANT4 MC simulation was modified to match the experimental setup. Thirty sensitive volumes, on the axial-phantom plane were included at each depth in the simulation to characterize the internal dose variation and compare it to the experimental TLD and MOSFET measurements. Experimental 2D dose maps were obtained with the GafChromic films and compared to the simulated 2D dose distributions estimated with the MC simulations. ResultsThe sensitivity of the MOSFET dosimeters and GafChromic films increased with x-ray energy, by up to 37% and 48%, respectively. Dose-response curves for the GafChromic film result in an uncertainty lower than 5% above 6 mGy, when a logarithmic relationship is used in the dose range of 1-10 mGy. All experimental values fall within the experimental uncertainty and a good agreement (within 5%) is found against the MC simulation. The dose decreased with increasing phantom depth, with the reduction being similar to 80% after 3 cm. The uncertainty of the empirical measurements makes the experimental values compatible with a flat behavior across the phantom slab for all the investigated depths, while the MC points to a dose profile with a maximum toward the center of the phantom. ConclusionsThe calibration procedures and the experimental methodologies proposed lead to good accuracy for internal breast dose estimation. In addition, these procedures can be successfully applied to validate MC codes for breast dosimetry at the local dose level. The agreement among the experimental and MC results not only shows the correctness of the empirical procedures used but also of the simulation parameters.
引用
收藏
页码:1724 / 1737
页数:14
相关论文
共 50 条
  • [1] Internal breast dosimetry in mammography: Monte Carlo validation in homogeneous and anthropomorphic breast phantoms with a clinical mammography system
    Fedon, Christian
    Caballo, Marco
    Sechopoulos, Ioannis
    [J]. MEDICAL PHYSICS, 2018, 45 (08) : 3950 - 3961
  • [2] Validation of X-Ray Spectra in Screening Mammography Simulated with Monte Carlo
    Mohammed Talbi
    M’hamed El Mansouri
    Aziz Oustous
    Oussama Nhila
    Rajaa Sebihi
    Mohammed Khalis
    [J]. Moscow University Physics Bulletin, 2021, 76 : S30 - S35
  • [3] Validation of X-Ray Spectra in Screening Mammography Simulated with Monte Carlo
    Talbi, Mohammed
    El Mansouri, M'hamed
    Oustous, Aziz
    Nhila, Oussama
    Sebihi, Rajaa
    Khalis, Mohammed
    [J]. MOSCOW UNIVERSITY PHYSICS BULLETIN, 2021, 76 (SUPPL 1) : S30 - S35
  • [4] A Monte Carlo simulation of x-ray mammography
    Spyrou, G
    Panayiotakis, G
    Tzanakos, G
    [J]. PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 842 - 843
  • [5] X-ray dosimetry: comparing Monte Carlo simulations and experimental data
    Stichelbaut, F
    Bol, JL
    Lundhal, B
    Martin, F
    Rose, G
    Schlecht, J
    Smith, R
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2004, 71 (1-2) : 345 - 349
  • [6] Monoenergetic X-ray beam in digital full field mammography
    Diekmann, F
    Diekmann, SD
    Fischer, T
    Arkadiev, V
    Lawaczeck, R
    Hamm, BK
    [J]. RADIOLOGY, 2002, 225 : 417 - 417
  • [7] Mammography X-ray spectra simulated with Monte Carlo
    Vega-Carrillo, H. R.
    Gonzalez, J. Ramirez
    Manzanares-Acuna, E.
    Hernandez-Davila, V. M.
    Villasana, R. Hernandez
    Mercado, G. A.
    [J]. MEDICAL PHYSICS, 2008, 1032 : 208 - 211
  • [8] Monte Carlo simulation of x-ray spectra in mammography
    Ng, KP
    Kwok, CS
    Tang, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05): : 1309 - 1318
  • [9] Monte Carlo simulations for X-ray breast dosimetry using homogeneous and heterogeneous phantoms
    Tucciariello, R. M.
    Barca, P.
    Del Sarto, D.
    Lamastra, R.
    Traino, A. C.
    Fantacci, M. E.
    [J]. NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2021, 44 (4-5):
  • [10] Monte Carlo and 60Co-based kilovoltage x-ray dosimetry methods
    Lawless, Michael J.
    Dimaso, Lianna
    Palmer, Benjamin
    Micka, John
    Culberson, Wesley S.
    DeWerd, Larry A.
    [J]. MEDICAL PHYSICS, 2018, 45 (12) : 5564 - 5576