Cone beam X-ray luminescence computed tomography reconstruction with a priori anatomical information

被引:2
|
作者
Lo, Pei-An [1 ]
Lin, Meng-Lung [1 ]
Jin, Shih-Chun [2 ]
Chen, Jyh-Cheng [2 ]
Lin, Syue-Liang [2 ]
Chang, C. Allen [2 ]
Chiang, Huihua Kenny [1 ]
机构
[1] Natl Yang Ming Univ, Inst Biomed Engn, Taipei 112, Taiwan
[2] Natl Yang Ming Univ, Dept Biomed Imaging & Radiol Sci, Taipei 112, Taiwan
关键词
X-ray luminescence computed tomography; image reconstruction; fluorescence diffuse optical tomography; cone beam; prior anatomical information; hybrid imaging system;
D O I
10.1117/12.2061321
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
X-ray luminescence computed tomography (XLCT) is a novel molecular imaging modality that reconstructs the optical distribution of x-ray-excited phosphor particles with prior information of anatomical CT image. The prior information improves the accuracy of image reconstruction. The system can also present anatomical CT image. The optical system based on a highly sensitive charge coupled device (CCD) is perpendicular to a CT system. In the XLCT system, the x-ray was adopted to excite the phosphor of the sample and CCD camera was utilized to acquire luminescence emitted from the sample in 360-degree projections in free-space. In this study, the fluorescence diffuse optical tomography (FDOT)-like algorithm was used for image reconstruction, the structural prior information was incorporated in the reconstruction by adding a penalty term to the minimization function. The phosphor used in this study is Gd2O2S:Tb. For the simulation and experiments, the data was collected from 16 projections. The cylinder phantom was 40 mm in diameter and contains 8 mm diameter inclusion; the phosphor in the in vivo study was 5 mm in diameter at a depth of 3 mm. Both the estimation errors were no more than 5%. Based on the results from these simulation and experimental studies, the novel XLCT method has demonstrated the feasibility for in vivo animal studies.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Direct prior regularization from anatomical images for cone beam X-ray luminescence computed tomography reconstruction
    Gao, Peng
    Rong, Junyan
    Liu, Tianshuai
    Zhang, Wenli
    Pu, Huangsheng
    Liang, Zhengrong
    Lu, Hongbing
    [J]. MEDICAL IMAGING 2018: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2018, 10578
  • [2] Quantitative cone beam X-ray luminescence tomography/X-ray computed tomography imaging
    Chen, Dongmei
    Zhu, Shouping
    Chen, Xueli
    Chao, Tiantian
    Cao, Xu
    Zhao, Fengjun
    Huang, Liyu
    Liang, Jimin
    [J]. APPLIED PHYSICS LETTERS, 2014, 105 (19)
  • [3] Adaptive shrinking reconstruction framework for cone-beam X-ray luminescence computed tomography
    Zhang, Haibo
    Huang, Xiaodong
    Zhou, Mingquan
    Geng, Guohua
    He, Xiaowei
    [J]. BIOMEDICAL OPTICS EXPRESS, 2020, 11 (07) : 3717 - 3732
  • [4] Cone beam x-ray luminescence computed tomography: A feasibility study
    Chen, Dongmei
    Zhu, Shouping
    Yi, Huangjian
    Zhang, Xianghan
    Chen, Duofang
    Liang, Jimin
    Tian, Jie
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [5] MLEM reconstruction with specific initial image for cone-beam x-ray luminescence computed tomography
    Ruan, Jiabin
    Gao, Peng
    Liu, Tianshuai
    Zhan, Yonghua
    Lu, Hongbing
    Rong, Junyan
    [J]. MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [6] Scattering-compensated cone beam x-ray luminescence computed tomography
    Gao, Peng
    Rong, Junyan
    Pu, Huangsheng
    Liu, Wenlei
    Liao, Qimei
    Lu, Hongbing
    [J]. MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [7] Cone Beam X-ray Luminescence Computed Tomography Based on Bayesian Method
    Zhang, Guanglei
    Liu, Fei
    Liu, Jie
    Luo, Jianwen
    Xie, Yaoqin
    Bai, Jing
    Xing, Lei
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2017, 36 (01) : 225 - 235
  • [8] Cone-beam x-ray luminescence computed tomography based on x-ray absorption dosage
    Liu, Tianshuai
    Rong, Junyan
    Gao, Peng
    Zhang, Wenli
    Liu, Wenlei
    Zhang, Yuanke
    Lu, Hongbing
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2018, 23 (02)
  • [9] Dynamic cone beam X-ray luminescence computed tomography with principal component analysis
    Gao, Peng
    Rong, Junyan
    Liu, Tianshuai
    Zhang, Wenli
    Pu, Huangsheng
    Zhang, Yuanke
    Lu, Hongbing
    [J]. MEDICAL IMAGING 2018: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2018, 10578
  • [10] X-ray luminescence computed tomography using a focused x-ray beam
    Zhang, Wei
    Lun, Michael C.
    Nguyen, Alex Anh-Tu
    Li, Changqing
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2017, 22 (11)