Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube

被引:7
|
作者
Shin, Sunghwan [1 ]
Oh, Jongwon [1 ]
Park, Hyung-Doo [1 ]
机构
[1] Sungkyunkwan Univ, Samsung Med Ctr, Dept Lab Med & Genet, Sch Med, 81 Irwon Ro, Seoul 06351, South Korea
关键词
Barricor tube; Bias; Turnaround time; Analytes; Serum separating tube; Plasma separating tube; STABILITY; HEMOLYSIS; QUALITY; IMPACT; GEL;
D O I
10.3343/alm.2021.41.1.114
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of -10.29% and -9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit.
引用
收藏
页码:114 / 119
页数:6
相关论文
共 11 条
  • [1] Evaluation of BD Barricor™ Plasma Blood Collection Tube for Biochemical Tests
    Kosem, Arzu
    Topcuoglu, Canan
    Sezer, Sevilay
    Cevher, Simal K.
    Yenigun, Ezgi C.
    Dede, Fatih
    Turhan, Turan
    CLINICAL LABORATORY, 2019, 65 (12) : 2315 - 2327
  • [2] Comparison of Becton Dickinson Vacutainer rapid serum tube with the serum separator tube for routine chemistry and immunoassay tests
    Yan, Ronald
    Lou, Amy
    Watts, Gail
    Tarr, Heather
    Smith, Hilary
    Kinney, Lois
    Chen, Yu
    JOURNAL OF CLINICAL PATHOLOGY, 2014, 67 (07) : 599 - 604
  • [3] Verification of BD Vacutainer® Barricor™ plasma blood collection tube
    Smolcic, V. Supak
    Ongaro, L.
    Bilic-Zulle, L.
    CLINICA CHIMICA ACTA, 2019, 493 : S524 - S524
  • [4] Evaluation of the Becton-Dickinson rapid serum tube: does it provide a suitable alternative to lithium heparin plasma tubes?
    Dimeski, Goce
    Masci, Paul P.
    Trabi, Manuela
    Lavin, Martin F.
    de Jersey, John
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (05) : 651 - 657
  • [5] A comparison of stability of chemical analytes in plasma from the BD Vacutainer® Barricor™ tube with mechanical separator versus tubes containing gel separator
    Gawria, Ghassaan
    Tillmar, Linda
    Landberg, Eva
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2020, 34 (02)
  • [6] COMPARISON of the BD Barricor Tube with BD serum separator tube(SST) for selected chemistry analytes on Beckman-Coulter AU 5800 and DXI.
    Paul, Tan
    Magne, Fabien
    Hetu, Pierre-Olivier
    Marie-Claire, Belanger
    CLINICAL BIOCHEMISTRY, 2018, 61 : 51 - 51
  • [7] Suitability of Becton Dickinson Vacutainer rapid serum tube for collecting and storing blood samples for antibiotic and anticonvulsant drug monitoring
    Yan, Ronald
    Colantonio, David
    Wong, Pui-Yuen
    Chen, Yu
    JOURNAL OF CLINICAL PATHOLOGY, 2014, 67 (09) : 807 - 810
  • [8] Suitability of Becton Dickinson Vacutainer® Rapid Serum Tube (RST) for Collecting and Storing Blood Samples for Anticonvulsant Drug Monitoring
    Chen, Y.
    Walsh, W.
    Wang, Y.
    Wong, P.
    Colantonio, D.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A248 - A249
  • [9] Suitability of Becton Dickinson Vacutainer® Rapid Serum Tube (RST) for Collecting and Storing Blood Samples for Anticonvulsant and Antibiotic Therapeutic Drug Monitoring
    Chen, Y.
    Walsh, W.
    Wang, Y.
    Wong, P. Y.
    Colantonio, D.
    THERAPEUTIC DRUG MONITORING, 2009, 31 (05) : 638 - 638
  • [10] COMPARISON OF THE BACT ALERT PEDIATRIC BLOOD CULTURE SYSTEM, PEDI-BACT, WITH CONVENTIONAL CULTURE USING THE 20-MILLILITER BECTON-DICKINSON SUPPLEMENTED PEPTONE BROTH TUBE
    KRISHER, KK
    WHYBURN, DR
    KOEPNICK, FE
    JOURNAL OF CLINICAL MICROBIOLOGY, 1993, 31 (04) : 793 - 797