Failure analysis of the tibial component baseplate after total knee arthroplasty

被引:4
|
作者
Oliveira, A. L. L. [1 ]
Cueva, E. G. [2 ]
Carvalho, R. T. [3 ]
机构
[1] Univ Sao Paulo, Dept Mech Engn, BR-2231 Sao Paulo, Brazil
[2] Univ Brasilia, UnB Gama Coll, Area Especial, Brasilia, DF, Brazil
[3] Orthopaed & Traumatol Hosp Publ Workers State Sao, Sao Paulo, Brazil
关键词
Total knee arthroplasty; Baseplate; Failure analysis; Fracture; Prostheses design; REPLACEMENT; FRACTURE; WEAR; TRAY; UHMWPE; DESIGN; JOINTS; PERFORMANCE; PROSTHESIS; WALKING;
D O I
10.1016/j.engfailanal.2013.10.012
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The present work was performed on a retrieved modular prosthetic tibial component, which presented a fracture in the medial portion of the baseplate. In order to understand the causes that led to the failure of the baseplate, the steel, the mechanical and structural characteristics and the fracture were analyzed and correlated with numerical studies related to the biomechanical loads and displacements transmitted to the knee joint during human gait. The results showed a set of adverse events that occurred before the fracture of the baseplate. The steel, used for the manufacture of tibial component, had a high ferrite content, low mechanical strength, and did not comply with standard specifications. Design errors of the baseplate and insert assembly were shown to be the cause for the damage to the polyethylene of the insert. This damage included wear, plastic deformation and plaque formation by delamination, which increased the gap between the fitting guides. These phenomena enabled the relative displacement between the components, and the misaligning of the mechanical axis, causing an increase of the bending moment in the medial region of the baseplate. The load peaks concentrated in the primary contact area with the medial femoral condyle exceeded the mechanical strength limits of the piece, causing the fracture of the metallic tibial baseplate. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:147 / 154
页数:8
相关论文
共 50 条
  • [1] Catastrophic Failure of the Tibial Component After Total Knee Arthroplasty: Fracture and Dissociation Between the Baseplate and Stem
    Worgul, Christopher A.
    Lentine, Brandon
    Dicken, Quinten G.
    Freccero, David M.
    ARTHROPLASTY TODAY, 2022, 14 : 194 - 198
  • [2] Retrieval study of tibial baseplate fracture after total knee arthroplasty
    Chatterji, U
    Ashworth, MJ
    Smith, AL
    Brewster, N
    Lewis, PL
    JOURNAL OF ARTHROPLASTY, 2005, 20 (01): : 101 - 107
  • [3] Tibial component failure mechanisms in total knee arthroplasty
    Berend, ME
    Ritter, MA
    Meding, JB
    Faris, PM
    Keating, EM
    Redelman, R
    Faris, GW
    Davis, KE
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) : 26 - 34
  • [4] Radiographic False Evidence of a Tibial Baseplate Fracture After Total Knee Arthroplasty
    Constantinescu, David S.
    Ross, Jeremy A.
    Patel, Nirav K.
    Strong, Benjamin M.
    Giambra, Laura A.
    Golladay, Gregory J.
    ARTHROPLASTY TODAY, 2021, 8 : 114 - 117
  • [5] AN ANALYSIS OF TIBIAL COMPONENT DESIGN IN TOTAL KNEE ARTHROPLASTY
    MURASE, K
    CROWNINSHIELD, RD
    PEDERSEN, DR
    CHANG, TS
    JOURNAL OF BIOMECHANICS, 1983, 16 (01) : 13 - &
  • [6] Tibial baseplate fracture associated with polyethylene wear and osteolysis after total knee arthroplasty
    Cankaya, Deniz
    Yilmaz, Serdar
    Deveci, Alper
    Ozkurt, Bulent
    INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2014, 5 (02): : 49 - 51
  • [7] Full Versus Surface Tibial Baseplate Cementation in Total Knee Arthroplasty
    Galasso, Olimpio
    Jenny, Jean-Yves
    Saragaglia, Dominique
    Miehlke, Rolf K.
    ORTHOPEDICS, 2013, 36 (02) : E151 - E158
  • [8] FAILURE OF TOTAL KNEE ARTHROPLASTY DUE TO LOOSENING AND DEFORMATION OF TIBIAL COMPONENT
    DUCHEYNE, P
    KAGAN, A
    LACEY, JA
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1978, 60 (03): : 384 - 391
  • [9] Medial proximal tibial resorption after total knee arthroplasty according to the design of the cobalt chrome tibial baseplate
    Sang Jun Song
    Hyun Woo Lee
    Cheol Hee Park
    Archives of Orthopaedic and Trauma Surgery, 2023, 143 : 3401 - 3407
  • [10] Medial proximal tibial resorption after total knee arthroplasty according to the design of the cobalt chrome tibial baseplate
    Song, Sang Jun
    Lee, Hyun Woo
    Park, Cheol Hee
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (06) : 3401 - 3407