We present a case of an unusual complication after a resurfacing total knee arthroplasty. Fracture of the uncemented porous-coated femoral component occurred 4 years after its implantation. The mechanical axis was restored and collateral ligament balance was achieved at the primary procedure. At revision, the femoral component was found fractured at the junction of the trochlea with the medial condyle, anteriorly to the medial peg. A thin layer of fibrous tissue was interposed between bone and metal under the fracture area. Metallurgical analysis of the fractured component revealed fatigue failure but no structural defect. Lack of bony support and excessive cyclic loading led to fracture of the implant.
引用
下载
收藏
页码:1068 / 1071
页数:4
相关论文
共 50 条
  • [1] FRACTURE OF THE FEMORAL COMPONENT IN CEMENTLESS TOTAL KNEE ARTHROPLASTY
    WHITESIDE, LA
    FOSCO, DR
    BROOKS, JG
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1993, (286) : 71 - 77
  • [2] Atraumatic Fracture of Cemented Vanguard Total Knee Arthroplasty Femoral Component
    Vovos, Tyler J.
    Watters, Tyler S.
    Hughes, Erin M.
    McNabb, David C.
    ORTHOPEDICS, 2020, 43 (05) : E476 - E479
  • [3] Fracture of the oxidized zirconium femoral component after total knee arthroplasty
    Ichimura, Ryuji
    Minamikawa, Tomohiko
    Nakagawa, Hiroshi
    Mori, Akira
    Midorikawa, Koji
    Sakuragi, Koji
    Minamikawa, Hiromichi
    KNEE, 2022, 36 : 27 - 32
  • [4] Component fracture in total knee arthroplasty
    Clarke, HD
    Trousdale, RT
    KNEE, 1999, 6 (04): : 261 - 267
  • [5] Fracture of the stem of the femoral component after resurfacing arthroplasty of the hip
    Bowman, N. K.
    Bucher, T. A.
    Bassily, A. A.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (12): : 1652 - 1653
  • [6] Femoral Component Fracture in a Total Knee Arthroplasty Patient With a Persistent Flexion Contracture
    Tate, Jackson P.
    Schneider, Andrew M.
    Brown, Nicholas M.
    ARTHROPLASTY TODAY, 2023, 22
  • [7] FEMORAL COMPONENT LOOSENING IN TOTAL KNEE ARTHROPLASTY
    KING, TV
    SCOTT, RD
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1985, (194) : 285 - 290
  • [8] Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty
    Huang, CH
    Yang, CY
    Cheng, CK
    JOURNAL OF ARTHROPLASTY, 1999, 14 (03): : 375 - 379
  • [9] Fracture of titanium nitride-coated femoral component after total knee arthroplasty
    Park, Se-Wook
    Kim, Hyungsuk
    In, Yong
    KNEE, 2014, 21 (04): : 871 - 874
  • [10] Rotational malalignment of the femoral component in total knee arthroplasty
    Fehring, TK
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2000, (380) : 72 - 79