Radial forces of stents used in thoracic endovascular aortic repair and bare self-expanding nitinol stents measured ex vivo - Rapid rescue for obstruction of the innominate artery using bare self-expanding nitinol stents

被引:10
|
作者
Matsumoto, Takuya [1 ]
Inoue, Kentaro [1 ]
Tanaka, Shinichi [1 ]
Aoyagi, Yukihiko [1 ]
Matsubara, Yutaka [1 ]
Matsuda, Daisuke [1 ]
Yoshiya, Keiji [1 ]
Yoshiga, Ryosuke [1 ]
Ohkusa, Tomoko [2 ]
Maehara, Yoshihiko [1 ]
机构
[1] Kyushu Univ, Grad Sch Med Sci, Dept Surg & Sci, Fukuoka, Japan
[2] Kyushu Univ, Ctr Clin & Translat Res, Fukuoka, Japan
关键词
Radial force measurement; TEVAR endoprostheses; self-expanding bare nitinol stents; MIGRATION; STENOSES; OUTCOMES; DESIGN;
D O I
10.1177/1708538116640131
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Purpose Our objective was to compare the radial forces of several stents ex vivo to identify stents suitable for rescue of the unexpected coverage of aortic arch branches in thoracic endovascular aortic repair. Methods We measured the radial forces of two types of self-expanding bare nitinol stents (E-luminexx and Epic) used singly or as double-walled pairs, and of three endoprostheses used in thoracic endovascular aortic repair (TEVAR, Gore c-TAG, Relay, and Valiant) by compressing the stent using an MTS Instron universal testing machine (model #5582). We also examined the compressive effects of the TEVAR endoprostheses and the bare nitinol stents on each other. Results The radial force was greater in the center than at the edge of each stent. In all stents tested, the radial force decreased incrementally with increasing stent diameter. The radial force at the center was two times greater when using two stents than with a single stent. In the compression test, only E-luminexx used as a pair was not compressed after compressing a Relay endoprosthesis by 12mm. Conclusion Two E-luminexx stents are appropriate to restore the blood flow if a TEVAR endoprosthesis covers the innominate artery following innominate-carotid-left subclavian arterial bypass.
引用
收藏
页码:36 / 41
页数:6
相关论文
共 50 条
  • [1] Comparison of visibility for four self-expanding nitinol bare stents in vitro
    Wiesinger, Benjamin
    Stuetz, Alexander
    Schmehl, Joerg
    Groezinger, Gerd
    Claussen, Claus Detlef
    Wiskirchen, Jakub
    ACTA RADIOLOGICA, 2012, 53 (09) : 1020 - 1025
  • [2] Self-expanding Nitinol stents for endovascular peripheral applications: A review
    Hoseini, Farzaneh
    Bellelli, Alberto
    Mizzi, Luke
    Pecoraro, Felice
    Spaggiari, Andrea
    Materials Today Communications, 2024, 41
  • [3] Heterogeneous vascular response after implantation of bare nitinol self-expanding stents in the swine femoropopliteal artery
    Kawarada, Osami
    Otsuka, Fumiyuki
    Miki, Kojiro
    Ikutomi, Masayasu
    Okada, Kozo
    Ogata, Soshiro
    Nishimura, Kunihiro
    Fitzgerald, Peter J.
    Honda, Yasuhiro
    CARDIOVASCULAR INTERVENTION AND THERAPEUTICS, 2023, 38 (02) : 210 - 222
  • [4] Heterogeneous vascular response after implantation of bare nitinol self-expanding stents in the swine femoropopliteal artery
    Osami Kawarada
    Fumiyuki Otsuka
    Kojiro Miki
    Masayasu Ikutomi
    Kozo Okada
    Soshiro Ogata
    Kunihiro Nishimura
    Peter J. Fitzgerald
    Yasuhiro Honda
    Cardiovascular Intervention and Therapeutics, 2023, 38 : 210 - 222
  • [5] Thermoelastic stress analysis of nitinol self-expanding stents
    Eaton-Evans, J.
    Dulieu-Barton, J. M.
    Little, E.
    Brown, I.
    ADVANCES IN EXPERIMENTAL MECHANICS IV, 2005, 3-4 : 47 - 52
  • [6] Nitinol self-expanding metal stents—safe and effective?
    Rahul Pannala
    Andrew S. Ross
    Nature Reviews Gastroenterology & Hepatology, 2012, 9 : 10 - 11
  • [7] Self-expanding nitinol stents: material and design considerations
    Stoeckel, D
    Pelton, A
    Duerig, T
    EUROPEAN RADIOLOGY, 2004, 14 (02) : 292 - 301
  • [8] Bopcompatibility aspects of nitinol self-expanding esophageal stents
    Alach, E.
    Bonicatto, S.
    Iasi, R.
    Troncoso, E.
    Caravelli, V
    Toloy, H.
    Barbero, A.
    Lavezzaro, M.
    Mainella, A.
    ANNALS OF ONCOLOGY, 2006, 17 : 86 - 86
  • [9] Self-expanding Nitinol Stents in proximal tracheostenosis.
    Hafner, B
    Mann, W
    Heussel, CP
    Scherhag, A
    Schlegel, J
    Kauczor, HU
    LARYNGO-RHINO-OTOLOGIE, 2000, 79 (03) : 165 - 170
  • [10] Performance Characteristics of Self-Expanding Peripheral Nitinol Stents
    Wissgott, C.
    Schmidt, W.
    Behrens, P.
    Schmitz, K. -P.
    Andresen, R.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (06): : 579 - 586