Comparison of measurements of anterior chamber angle via anterior segment optical coherence tomography and ultrasound biomicroscopy

被引:2
|
作者
Yu, Zi-Yan [1 ]
Huang, Ting [2 ]
Lu, Lu [1 ]
Qu, Bo [1 ]
机构
[1] China Med Univ, Key Lab Lens Res Liaoning Prov, Dept Ophthalmol, Eye Hosp,Affiliated Hosp 4, 11 Xinhua Rd, Shenyang 110005, Liaoning, Peoples R China
[2] Aier Mega Ophthalm Hosp Chongqing, Dept Cataract, Chongqing 400060, Peoples R China
关键词
Anterior segment-optical coherence tomography; Primary angle closure glaucoma; Anterior chamber angle; Angel opening distance at 500 m; Trabeculo-iris space area at 500 m(2); QUANTIFICATION; GLAUCOMA; EYES;
D O I
10.12998/wjcc.v8.i15.3249
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Glaucoma is an irreversible optic neuropathy with the loss of visual field and decrease of vision. The variable clinical manifestations may result in differential diagnostic difficulties. The early screening and diagnosis of glaucoma are currently experiencing a demand for anterior segment analysis tools that can gather more information with one short measurement. Therefore, we analyzed the agreement, difference, and correlation of chamber angle parameters such as angel opening distance at 500 mu m (AOD500) and trabeculo-iris space area at 500 mu m(2) (TISA500) measured by anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM). AIM To compare the differences, correlation, and agreement in measuring AOD500 and TISA500 by AS-OCT and UBM. METHODS Both AS-OCT and UBM were performed to measure AOD500 and TISA500 in 45 subjects (72 eyes). All subjects without glaucoma were collected from October 2015 to August 2016 at the Ophthalmology Department of the Fourth Affiliated Hospital of China Medical University. Data of the two groups (AOD500 and TISA500) were compared by nonparametric tests. Pearson correlative analysis and Bland-Altman analysis were used to compare the correlation and agreement. RESULTS There were no significant differences between AS-OCT and UBM in measuring AOD500 (P1 = 0.110, P2 = 0.633, P3 = 0.078, and P4 = 0.474) and TISA500 (P1 = 0.584, P2 = 0.889, P3 = 0.297, and P4 = 0.550) of the four quadrants of the anterior chamber angle. There was a high correlation in measuring AOD500 (r1 = 0.562, r2 = 0.671, r3 = 0.635, and r4 = 0.720; P < 0.001) and TISA500 (r1 = 0.584, r2 = 0.889, r3 = 0.297, and r4 = 0.550; P < 0.001). There was a good agreement in measuring AOD500 and TISA500 by the two modalities. CONCLUSION There is a high correlation and agreement between AOD500 and TISA500 measurements by AS-OCT and UBM. They are interchangeable under some circumstances. AS-OCT proves to be a better early screening tool for glaucoma.
引用
收藏
页码:3249 / U143
页数:11
相关论文
共 50 条
  • [1] Comparison of measurements of anterior chamber angle via anterior segment optical coherence tomography and ultrasound biomicroscopy
    Zi-Yan Yu
    Ting Huang
    Lu Lu
    Bo Qu
    [J]. World Journal of Clinical Cases, 2020, (15) : 3249 - 3258
  • [2] Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment
    Dada, Tanuj
    Sihota, Ramanjit
    Gadia, Ritu
    Aggarwal, Anand
    Mandal, Subrata
    Gupta, Viney
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05): : 837 - 840
  • [3] Reproducibility of Anterior Chamber Angle Measurements with Anterior Segment Optical Coherence Tomography
    Tan, Annelie N.
    Sauren, Loes D. C.
    de Brabander, John
    Berendschot, Tos T. J. M.
    Passos, Valeria Lima
    Webers, Carroll A. B.
    Nuijts, Rudy M. M. A.
    Beckers, Henny J. M.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (05) : 2095 - 2099
  • [4] Anterior segment imaging: ultrasound biomicroscopy and anterior segment optical coherence tomography
    Nolan, Winifred
    [J]. CURRENT OPINION IN OPHTHALMOLOGY, 2008, 19 (02) : 115 - 121
  • [5] Comparison of Different Modes in Optical Coherence Tomography and Ultrasound Biomicroscopy in Anterior Chamber Angle Assessment
    Wang, Dandan
    Pekmezci, Melike
    Basham, Ryan P.
    He, Mingguang
    Seider, Michael I.
    Lin, Shan C.
    [J]. JOURNAL OF GLAUCOMA, 2009, 18 (06) : 472 - 478
  • [6] Prospective comparison of ultrasound biomicroscopy and anterior segment optical coherence tomography for evaluation of anterior chamber dimensions in European eyes with primary angle closure
    Mansouri, K.
    Sommerhalder, J.
    Shaarawy, T.
    [J]. EYE, 2010, 24 (02) : 233 - 239
  • [7] Imaging of Anterior Segment Tumours: A Comparison of Ultrasound Biomicroscopy Versus Anterior Segment Optical Coherence Tomography
    Kottaridou, Eleni
    Hatoum, Adam
    [J]. CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (01)
  • [8] Prospective comparison of ultrasound biomicroscopy and anterior segment optical coherence tomography for evaluation of anterior chamber dimensions in European eyes with primary angle closure
    K Mansouri
    J Sommerhalder
    T Shaarawy
    [J]. Eye, 2010, 24 : 233 - 239
  • [9] Comparison of Ultrasound Biomicroscopy versus Anterior Segment Optical Coherence Tomography for Imaging of Anterior Segment Tumors
    Bianciotto, C. G.
    Romanelli, M.
    Guzman, M.
    Mazzuca, D.
    Green, R.
    Shields, C. L.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [10] Assessment of Anterior Chamber by Ultrasound Biomicroscopy and Anterior Segment Optical Coherence Tomography in Patients with Inflammatory Glaucoma
    Qian, Yiyong
    Liu, Lin
    Shi, Yuehui
    Wang, Minli
    Li, Min
    Zou, Jun
    [J]. JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2019, 47 (12) : 5950 - 5956