Effects of poor bone window on the assessment of cerebral autoregulation with transcranial Doppler sonography - A source of systematic bias and strategies to avoid it

被引:6
|
作者
Lorenz, Matthias W. [1 ]
Loesel, Nadine [1 ]
Thoelen, Nina [1 ]
Gonzalez, Marilen [1 ]
Lienerth, Christian [2 ]
Dvorak, Florian [1 ]
Roelz, Waltraud [1 ]
Humpich, Marek [1 ]
Sitzer, Matthias [1 ]
机构
[1] Univ Frankfurt, Dept Neurol, D-60528 Frankfurt, Germany
[2] Univ Frankfurt, Brain Imaging Ctr, D-60528 Frankfurt, Germany
关键词
Vascular dementia; Cerebral autoregulation; Transcranial Doppler sonography; Insonation quality; Temporal bone window; Ultrasound contrast agent; CAROTID-ARTERY STENOSIS; WHITE-MATTER LESIONS; COLLATERAL FLOW PATTERNS; VASCULAR DEMENTIA; VASOMOTOR REACTIVITY; BLOOD-PRESSURE; ACETAZOLAMIDE VASOREACTIVITY; CEREBROVASCULAR REACTIVITY; HYPERTENSIVE PATIENTS; OCCLUSIVE DISEASE;
D O I
10.1016/j.jns.2009.02.329
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: The consequences of poor insonation conditions on autoregulation parameters assessed with transcranial Doppler (TCD) are unclear. Methods: We present two new complementary methods to assess the quality of a TCD signal. Inserting a thin aluminium foil between TCD probe and skin makes a simple model to artificially worsen a good insonation window. Validation studies are presented. We assessed insonation quality and cerebral autoregulation parameters with transfer function analysis and cross correlation in 46 healthy volunteers with and without the aluminium foil model. The same studies were operated on 45 patients with good insonation windows, naive. after worsening the bone window and during constant infusion of an ultrasound contrast agent. For studying reproducibility, we assessed autoregulation twice in 30 patients with poor bone windows, with and without constant contrast infusion. Results: Both methods to measure insonation quality are valid and reproducible. The aluminium foil model realistically simulates a natural poor bone window, reducing the signal quality (e.g. energy of the signal spectrum from 33.4 +/- 3.5 to 26.2 +/- 2.5 dB, p<0.001). Thereby, the autoregulation parameters are systematically biased (e.g. phase difference from 37.3 +/- 10.1 degrees to 25.9 +/- 15.1 degrees, p<0.001); while with the use of an ultrasound contrast agent this can be largely compensated (phase difference 35.7 +/- 10.7 degrees, p<0.001). The reproducibility is significantly improved (ICC from 0.76 to 0.90, p<0.05). Conclusions: Poor bone windows can cause considerable bias in TCD autoregulation parameters. This bias might be avoided by the use of ultrasound contrast agents, which may greatly improve the credibility of TCD autoregulation assessment in elderly patients. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 6 条
  • [1] Influence of temporal insonation window quality on the assessment of cerebral autoregulation with transcranial Doppler sonography
    Lorenz, Matthias W.
    Gonzalez, Marilen
    Lienerth, Christian
    Loesel, Nadine
    Thoelen, Nina
    Sitzer, Matthias
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2007, 33 (10): : 1540 - 1545
  • [2] Assessment of cerebral autoregulation with transcranial Doppler sonography in poor bone windows using constant infusion of an ultrasound contrast agent
    Lorenz, Matthias W.
    Thoelen, Nina
    Loesel, Nadine
    Lienerth, Christian
    Gonzalez, Marilen
    Humpich, Marek
    Roelz, Waltraud
    Dvorak, Florian
    Sitzer, Matthias
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2008, 34 (03): : 345 - 353
  • [3] Mechanisms of cerebral autoregulation, assessment and interpretation by means of transcranial Doppler sonography
    Hilz, MJ
    Stemper, B
    Heckmann, JG
    Neundörfer, B
    FORTSCHRITTE DER NEUROLOGIE PSYCHIATRIE, 2000, 68 (09) : 398 - 412
  • [4] Transcranial Doppler sonography during acute 80° head-down tilt (HDT) for the assessment of cerebral autoregulation in humans
    Heckmann, JG
    Hilz, MJ
    Hagler, H
    Mück-Weymann, M
    Neundörfer, B
    NEUROLOGICAL RESEARCH, 1999, 21 (05) : 457 - 462
  • [5] Transcranial Doppler sonography during acute 80-degree head-down tilt for the assessment of cerebral autoregulation humans
    Hilz, MJ
    Russo, H
    Hagler, H
    Mück-Weymann, M
    Neundörfer, B
    ANNALS OF NEUROLOGY, 1998, 44 (03) : 482 - 483
  • [6] Cerebral blood flow autoregulation assessment by correlation analysis between mean arterial blood pressure and transcranial doppler sonography or near infrared spectroscopy is different: A pilot study
    Thudium, Marcus
    Moestl, Stefan
    Hoffmann, Fabian
    Hoff, Alex
    Kornilov, Evgeniya
    Heusser, Karsten
    Tank, Jens
    Soehle, Martin
    PLOS ONE, 2023, 18 (06):