Voltammetric behavior and square-wave voltammetric determination of cefotaxime in urine

被引:31
|
作者
Aleksic, Mara M.
Kapetanovic, Vera
机构
[1] Univ Belgrade, Fac Pharm, Inst Phys Chem, Belgrade 11000, Serbia
[2] Univ Belgrade, Fac Pharm, Inst Analyt Chem, Belgrade 11000, Serbia
来源
JOURNAL OF ELECTROANALYTICAL CHEMISTRY | 2006年 / 593卷 / 1-2期
关键词
cefotaxime; square-wave voltammetry; urine;
D O I
10.1016/j.jelechem.2006.06.011
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The voltammetric behavior of cefotaxime (CFX) is investigated by cyclic and square-wave voltammetry in BR buffers (pH 1.80-12.0). Based on the cathodic reduction peak at approximately -0.5 V in BR buffer (pH 2.8 and 9.25), a robust, highly reliable square-wave voltammetric method (SW) was developed for determination of CFX. The linearity was achieved by SW voltammetry in 7.5 x 10(-9) M (3.52 ng ml(-1)) to 1.0 X 10(-7) M (47 ng ml(-1)) and from 12 x 10(-7) M (56.4 ng ml(-1)) to 1.2 x 10(-6) M (568.8 ng ml(-1)) range at pH 2.8 with detection and quantification limits of 0.813 ng ml(-1) and 2.71 ng ml(-1), and 12.6 ng ml(-1) and 42.2 ng ml(-1), respectively. At pH 9.25 linear range was obtained from 4 x 10(-8) M (18.8 ng ml(-1)) to 1.6 x 10(-7) M (75.2 ng ml(-1)) with detection and quantification limits of 2.947 ng ml(-1) and 9.82 ng ml(-1). The method was applied for CFX determination in spiked humane urine sample. A detection and quantification limits of 7.2 ng ml(-1) and 23.9 ng ml(-1) were achieved for determination of CFX in urine at pH 2.8 and 13.6 ng ml(-1) and 45.1 ng ml(-1) at pH 9.25. The comparative reference method was DPV. The proposed method was checked for determination of CFX in real human urine, and selectivity of the method over the metabolites was found to be quite satisfactory. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:258 / 266
页数:9
相关论文
共 50 条
  • [1] Determination of furazolidone in urine by square-wave voltammetric method
    Parham, H.
    Esfahani, B. Aibaghi
    [J]. JOURNAL OF THE IRANIAN CHEMICAL SOCIETY, 2008, 5 (03) : 453 - 457
  • [2] Determination of furazolidone in urine by square-wave voltammetric method
    H. Parham
    B. Aibaghi Esfahani
    [J]. Journal of the Iranian Chemical Society, 2008, 5 : 453 - 457
  • [3] Voltammetric behavior and square-wave voltammetric determination of trepibutone at a pencil graphite electrode
    Gao, W
    Song, JF
    Wu, NY
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2005, 576 (01) : 1 - 7
  • [4] SQUARE-WAVE VOLTAMMETRIC DETERMINATION OF DAMINOZIDE
    IANNIELLO, RM
    [J]. ANALYTICA CHIMICA ACTA, 1987, 193 : 81 - 86
  • [5] Square-wave Voltammetric Determination of Ezetimibe
    Yola, Mehmet Lutfi
    Ozaltin, Nuran
    [J]. REVISTA DE CHIMIE, 2011, 62 (04): : 420 - 426
  • [6] SQUARE-WAVE VOLTAMMETRIC DETERMINATION OF ACETALDEHYDE IN POVIDONE
    IANNIELLO, RM
    COLONNESE, R
    MACHNICKI, N
    [J]. JOURNAL OF THE ASSOCIATION OF OFFICIAL ANALYTICAL CHEMISTS, 1987, 70 (03): : 566 - 568
  • [7] Square-wave voltammetric method for determination of aconitic acid
    Spirevska, I
    Soptrajanova, L
    Gulaboski, R
    [J]. ANALYTICAL LETTERS, 2000, 33 (05) : 919 - 928
  • [8] Electrochemical properties and square-wave voltammetric determination of pravastatin
    Biljana Nigović
    [J]. Analytical and Bioanalytical Chemistry, 2006, 384 : 431 - 437
  • [9] Adsorptive stripping square-wave voltammetric behavior of insulin
    Nevado, JJB
    Flores, JR
    Peñalvo, GC
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 364 (08): : 753 - 757
  • [10] Electrochemical properties and square-wave voltammetric determination of pravastatin
    Nigovic, B
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 384 (02) : 431 - 437