DETERMINATION OF ACYCLOVIR IN RABBIT PLASMA BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC (HPLC) TECHNIQUE

被引:3
|
作者
Malik, Nadia Shamshad [1 ]
Ahmads, Mahmood [2 ]
Minhas, Muhammad Usman [3 ]
Khalid, Qandeel [3 ]
机构
[1] Capital Univ Sci & Technol, Dept Pharm, Islamabad, Pakistan
[2] Univ Cent Punjab, Fac Pharm, Lahore, Pakistan
[3] Islamia Univ Bahawalpur, Fac Pharm & Alternat Med, Bahawalpur, Pakistan
来源
ACTA POLONIAE PHARMACEUTICA | 2019年 / 76卷 / 03期
关键词
reversed-phase high-performance liquid chromatography; acyclovir; plasma; SOLID-PHASE EXTRACTION; GANCICLOVIR; ASSAY; VALIDATION; METABOLITE; COLUMN; SERUM;
D O I
10.32383/appdr/100388
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
A rapid, sensitive and simple reversed-phase high-performance liquid chromatographic (HPLC) method has been developed and validated for the determination of acyclovir (ACV) in rabbit plasma. BDS C18 column was used to conduct analysis using ammonium dihydrogen phosphate buffer (50 mM) and methanol as mobile phase (98 : 2), with pH adjusted to 2.5 using orthophosphoric acid. The flow rate was kept at 1 mL/min. Selective precipitation of plasma proteins was done by adding 5% perchloric acid. Precipitated plasma proteins were separated by centrifugation. ACV moves in a supernatant, which was snapped and passed through a syringe filtration assembly. Direct injection of supernatant was given into a BDS C18 column and ACV was detected at 256 nm. The limit of detection for ACV in plasma was estimated as 15 ng/mL whereas the limit of quantitation was calculated as 25 ng/mL. Moreover, the developed method has been found to be selective and linear into concentration range of 25-2000 ng/mL. The present method could be successfully applied to samples from bioavailability and bioequivalence studies.
引用
收藏
页码:421 / 429
页数:9
相关论文
共 50 条
  • [1] AUTOMATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF ACYCLOVIR IN PLASMA
    SWART, KJ
    HUNDT, HKL
    GROENEWALD, AM
    JOURNAL OF CHROMATOGRAPHY A, 1994, 663 (01) : 65 - 69
  • [2] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF ACYCLOVIR IN SERUM
    SMITH, RL
    WALKER, DD
    JOURNAL OF CHROMATOGRAPHY, 1985, 343 (01): : 203 - 207
  • [3] Robust and Sensitive High-Performance Liquid Chromatographic-UV Detection Technique for the Determination of Tigecycline in Rabbit Plasma
    Zorpas, Kostas M.
    Valsami, Georgia N.
    Vryonis, Evangelos V.
    Skoutelis, Athanasios T.
    Archontaki, Helen A.
    JOURNAL OF AOAC INTERNATIONAL, 2011, 94 (03) : 847 - 856
  • [4] Simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma using fluorescence detection
    Peh, KK
    Yuen, KH
    JOURNAL OF CHROMATOGRAPHY B, 1997, 693 (01): : 241 - 244
  • [5] A simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma and application to a pharmacokinetic study
    Yu, Liyan
    Xiang, Bingren
    Zhan, Ying
    ARZNEIMITTEL-FORSCHUNG-DRUG RESEARCH, 2008, 58 (04): : 199 - 202
  • [6] High Performance Liquid Chromatographic Determination of Cinnamic Acid in Rabbit Plasma and Application in Study of Pharmacokinetics
    Wang Rui
    Sun Tianhui
    Jing Dan
    Chen Xiaohui
    Yu Zhiguo
    Bi Kaishun
    CHINESE JOURNAL OF CHROMATOGRAPHY, 2005, 23 (03) : 273 - 275
  • [7] IMPROVED LIQUID-CHROMATOGRAPHIC METHOD FOR ACYCLOVIR DETERMINATION IN PLASMA
    MOLOKHIA, AM
    NIAZY, EM
    ELHOOFY, SA
    ELDARDARI, ME
    JOURNAL OF LIQUID CHROMATOGRAPHY, 1990, 13 (05): : 981 - 989
  • [8] Determination of acyclovir in human plasma by high-performance liquid chromatography
    Boulieu, R
    Gallant, C
    Silberstein, N
    JOURNAL OF CHROMATOGRAPHY B, 1997, 693 (01): : 233 - 236
  • [9] Technique validation by liquid chromatography for the determination of acyclovir in plasma
    Fernández, M
    Sepúlveda, J
    Aránguiz, T
    von Plessing, C
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 791 (1-2): : 357 - 363
  • [10] A SIMPLE ISOCRATIC HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC (HPLC) DETERMINATION OF NAPROXEN IN HUMAN PLASMA USING A MICROBORE COLUMN TECHNIQUE
    WANWIMOLRUK, S
    JOURNAL OF LIQUID CHROMATOGRAPHY, 1990, 13 (08): : 1611 - 1625