NITRATE DETERMINATION IN CHILEAN CALICHE SAMPLES BY UV-VISIBLE ABSORBANCE MEASUREMENTS AND MULTIVARIATE CALIBRATION

被引:0
|
作者
Bravo, Manuel [1 ]
Olivieri, Alejandro C. [3 ,4 ]
Oelckers, Beatriz [2 ]
机构
[1] P Univ Catolica Valparaiso, Inst Quim, Lab Quim Analt & Ambiental, Valparaiso, Chile
[2] SQM Ind, Antofagasta, Chile
[3] Univ Nacl Rosario, Fac Ciencias Bioquim & Farmaceut, Dept Quim Analit, RA-2000 Rosario, Santa Fe, Argentina
[4] Consejo Nacl Invest Cient & Tecn, Inst Quim, RA-2000 Rosario, Santa Fe, Argentina
关键词
Nitrate determination; UV-visible spectrophotometry; Multivariate calibration; Partial least-squares; PARTIAL LEAST-SQUARES; ANALYTE PREPROCESSING METHODS; WAVELENGTH SELECTION; SPECTROSCOPY; REGRESSION; MIXTURES; FIGURES; ERROR; MERIT; WATER;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Multivariate calibration of UV-visible spectral data using partial least-squares (PLS) has been applied to the determination of the nitrate content in Chilean Caliche samples, in the concentration range from 1 to 20% NaNO(3). The multivariate approach is required since the samples do also contain unknown interferences which are spectrally active in the useful wavelength region for nitrate quantitation (near 301 nm). A set of fifteen calibration samples was employed to build the multivariate model, selected using the Kennard-Stone methodology, starting from real Caliche samples whose nitrate content was previously determined using the reference Devarda method. The figures of merit of the multivariate model were satisfactory (the limit of detection and quantitation reached 0.04% and 0.12% of NaNO3, respectively, with ail average error of prediction of 0.3% of NaNO(3)). Then, the PLS model was then applied to a set of independent Caliche samples. The results were compared with a univariate UV approach, and with the nitrate content determined by the reference method, using the linear regression of predicted vs. reference concentration values, together with the elliptical joint confidence region test for the slope and intercept of the latter regression. The results indicate that the univariate method is unsuitable for analyzing the presently studied samples, unlike the multivariate model. Finally, the analytical methodology proposed appears as reliable and cheap alternative for routine analysis of a large number of caliche samples.
引用
收藏
页码:93 / 98
页数:6
相关论文
共 50 条
  • [1] Determination of Total Phenolic Content in Olive Oil Samples by UV-visible Spectrometry and Multivariate Calibration
    Fuentes, Edwar
    Baez, Maria E.
    Bravo, Manuel
    Cid, Camila
    Labra, Fabian
    FOOD ANALYTICAL METHODS, 2012, 5 (06) : 1311 - 1319
  • [2] Additive determination in an electrolytic zinc bath by UV-Visible spectroscopy and multivariate calibration
    Barriola, Ainara
    Garcia, Eva
    Ostra, Miren
    Ubide, Carlos
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2008, 155 (06) : D480 - D484
  • [3] Determination of the content and identity of lidocaine solutions with UV-visible spectroscopy and multivariate calibration
    Wiberg, K
    Hagman, A
    Burén, P
    Jacobsson, SP
    ANALYST, 2001, 126 (07) : 1142 - 1148
  • [4] Simultaneous determination of iron, cobalt, nickel and copper by UV-visible spectrophotometry with multivariate calibration
    Rodriguez, AMG
    de Torres, AG
    Pavon, JMC
    Ojeda, CB
    TALANTA, 1998, 47 (02) : 463 - 470
  • [5] AUTOMATIC ABSORBANCE CALIBRATION ROUTINE FOR A COMPUTERIZED UV-VISIBLE RAPID SCANNING SPECTROPHOTOMETER
    YACYNYCH, AM
    MARK, HB
    CHEMICAL INSTRUMENTATION, 1976, 7 (04): : 295 - 300
  • [6] Simultaneous determination of food colorants in liquid samples by UV-Visible spectroscopy and multivariate data analysis using a reduced calibration matrix
    Bordagaray, Ane
    Davila, Sergio
    Garcia-Arrona, Rosa
    Vidal, Maider
    Ostra, Miren
    JOURNAL OF CHEMOMETRICS, 2019, 33 (10)
  • [7] ELECTROCHEMICAL CLEANUP OF SOLUTIONS FOR UV-VISIBLE ASSAYS INVOLVING LOW ABSORBANCE OR DILUTE SAMPLES
    RUBINSON, KA
    MARK, HB
    ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1982, 15 (12): : 989 - 993
  • [8] Determination of Total Phenolic Content in Olive Oil Samples by UV–visible Spectrometry and Multivariate Calibration
    Edwar Fuentes
    María E. Báez
    Manuel Bravo
    Camila Cid
    Fabián Labra
    Food Analytical Methods, 2012, 5 : 1311 - 1319
  • [9] MULTIVARIATE OPTIMIZATION OF PARAMETERS FOR THE DETERMINATION OF THORIUM IN ROCK SAMPLES BY CLOUD POINT EXTRACTION COUPLED TO UV-VISIBLE SPECTROPHOTOMETRY
    Jalbani, Nusrat
    Soylak, Mustafa
    Munshi, Alia Bano
    Kazi, Tasneem Gul
    FRESENIUS ENVIRONMENTAL BULLETIN, 2014, 23 (9A): : 2304 - 2309
  • [10] Fast assessment of diene values in pyrolysis gasoline by UV-visible spectroscopy and multivariate calibration
    Hilgemann, Mauricio
    Nascimento, Paulo C.
    Dias, Daiane
    Guterres, Marcos V.
    Stringhini, Fabiane M.
    de Carvalho, Leandro M.
    Bohrer, Denise
    JOURNAL OF PETROLEUM SCIENCE AND ENGINEERING, 2011, 78 (02) : 283 - 287