Survival for up to six months in calves supported with an implantable axial flow ventricular assist device

被引:18
|
作者
Macha, M
Litwak, P
Yamazaki, K
Kameneva, M
Kerrigan, JP
Antaki, JF
Butler, KC
Thomas, DC
Taylor, LP
Kormos, RL
Griffith, BP
Borovetz, HS
机构
[1] UNIV PITTSBURGH,MED CTR,MCGOWAN CTR,DIV CARDIOTHORAC SURG,ARTIFICIAL HEART & LUNG PROGRAM,PITTSBURGH,PA 15219
[2] NIMBUS INC,RANCHO CORDOVA,CA
关键词
D O I
10.1097/00002480-199707000-00009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper summarizes the authors' in vivo experience to date with an implantable axial flow blood pump designed for longterm ventricular support. This small, valveless pump with blood-lubricated bearings has been implanted in six calves (83 +/- 6 kg) as a left ventricular assist device (LVAS). The left ventricle and descending thoracic aorta were cannulated by left thoracotomy, and the pump was placed in a subcutaneous pocket below the costal margin. Animals remained hemodynamically stable throughout the course of support during partial left ventricular bypass. Five animals were killed after 15, 27, 52, 57, and 181 days. The longest survivor (181 days) demonstrated normal pump function at the time death. Pump speed was maintained at 10,100 +/- 100 rpm, with an average pump flow rate of 4.9 +/- 0.5 L/min under resting physiologic conditions. Average plasma free hemoglobin was 17.4 +/- 7.5 mg/dl. Renal, hepatic, and hematologic indices remained within physiologic range in all of these animals, except during the immediate postoperative period. Histopathologic analyses of major organs after death revealed small renal cortical infarcts in five of six animals; the remaining organs were normal. These animal studies support the feasibility of this small implanted axial flow pump for long-term ventricular assistance.
引用
收藏
页码:311 / 315
页数:5
相关论文
共 50 条
  • [1] A miniature implantable axial flow ventricular assist device
    DeBakey, ME
    [J]. ANNALS OF THORACIC SURGERY, 1999, 68 (02): : 637 - 640
  • [2] Long-term animal survival with an implantable axial flow pump as a left ventricular assist device
    Konishi, H
    Antaki, JF
    Litwak, P
    Kameneva, M
    Yamazaki, K
    Macha, M
    Kerrigan, TP
    Mandarino, WA
    Butler, KC
    Borovetz, TS
    Kormos, RL
    [J]. ARTIFICIAL ORGANS, 1996, 20 (02) : 124 - 127
  • [3] Systemic right ventricle supported by implantable axial-flow assist device
    Netuka, Ivan
    Maly, Jiri
    Szarszoi, Ondrej
    Novotny, Jiri
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2009, 36 (02) : 403 - 403
  • [4] Flow visualization for the implantable ventricular assist device EVAHEART®
    Takashi Yamane
    Masahiro Nishida
    Hiroshi Kawamura
    Takayuki Miyakoshi
    Kenji Yamazaki
    [J]. Journal of Artificial Organs, 2013, 16 : 42 - 48
  • [5] Flow visualization for the implantable ventricular assist device EVAHEARTA®
    Yamane, Takashi
    Nishida, Masahiro
    Kawamura, Hiroshi
    Miyakoshi, Takayuki
    Yamazaki, Kenji
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2013, 16 (01) : 42 - 48
  • [6] Chronic survival of calves implanted with the DeBakey ventricular assist device
    Fossum, TW
    Morley, D
    Benkowski, R
    Tayama, E
    Olsen, DB
    Burns, G
    Miller, MW
    Franks, J
    Martinez, E
    Carroll, G
    Edwards, J
    Vinnerqvist, A
    Lynch, B
    Stein, F
    Noon, GP
    DeBakey, ME
    [J]. ARTIFICIAL ORGANS, 1999, 23 (08) : 802 - 806
  • [7] Implantable Cardioverter Defibrillators and Survival in Continuous-Flow Left Ventricular Assist Device Patients
    Kutyifa, Valentina
    Fernandez, Genaro
    Sherazi, Saadia
    Aktas, Mehmet
    Huang, David
    McNitt, Scott
    Papernov, Anna
    Wang, Meng
    Massey, H. Todd
    Chen, Leway
    Alexis, Jeffrey D.
    [J]. ASAIO JOURNAL, 2019, 65 (01) : 49 - 53
  • [8] Durability testing of components for the Jarvik 2000 completely implantable axial flow left ventricular assist device
    Marlinski, E
    Jacobs, G
    Deirmengian, C
    Jarvik, R
    [J]. ASAIO JOURNAL, 1998, 44 (05) : M741 - M744
  • [9] Research and development of an implantable, axial-flow left ventricular assist device: The Jarvik 2000 heart
    Frazier, OH
    Myers, TJ
    Jarvik, RK
    Westaby, S
    Pigott, DW
    Gregoric, ID
    Khan, T
    Tamez, DW
    Conger, JL
    Macris, MP
    [J]. ANNALS OF THORACIC SURGERY, 2001, 71 (03): : S125 - S132
  • [10] Continuous axial-flow left ventricular assist device (Jarvik 2000) maintains kidney and liver perfusion for up to 6 months
    Letsou, GV
    Myers, TJ
    Gregoric, ID
    Delgado, R
    Shah, N
    Robertson, K
    Radovancevic, B
    Frazier, OH
    [J]. ANNALS OF THORACIC SURGERY, 2003, 76 (04): : 1167 - 1170