High-stability cryogenic scanning tunneling microscope based on a closed-cycle cryostat

被引:41
|
作者
Hackley, Jason D. [1 ]
Kislitsyn, Dmitry A. [1 ]
Beaman, Daniel K. [1 ]
Ulrich, Stefan [2 ]
Nazin, George V. [1 ]
机构
[1] 1253 Univ Oregon, Dept Chem & Biochem, Eugene, OR 97403 USA
[2] RHK Technol Inc, Troy, MI 48083 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2014年 / 85卷 / 10期
基金
美国国家科学基金会;
关键词
MOLECULE VIBRATIONAL SPECTROSCOPY; SINGLE-MOLECULE; SURFACES;
D O I
10.1063/1.4897139
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report on the design and operation of a cryogenic ultra-high vacuum (UHV) scanning tunneling microscope (STM) coupled to a closed-cycle cryostat (CCC). The STM is thermally linked to the CCC through helium exchange gas confined inside a volume enclosed by highly flexible rubber bellows. The STM is thus mechanically decoupled from the CCC, which results in a significant reduction of the mechanical noise transferred from the CCC to the STM. Noise analysis of the tunneling current shows current fluctuations up to 4% of the total current, which translates into tip-sample distance variations of up to 1.5 picometers. This noise level is sufficiently low for atomic-resolution imaging of a wide variety of surfaces. To demonstrate this, atomic-resolution images of Au(111) and NaCl(100)/Au(111) surfaces, as well as of carbon nanotubes deposited on Au(111), were obtained. Thermal drift analysis showed that under optimized conditions, the lateral stability of the STM scanner can be as low as 0.18 /h. Scanning Tunneling Spectroscopy measurements based on the lock-in technique were also carried out, and showed no detectable presence of noise from the closed-cycle cryostat. Using this cooling approach, temperatures as low as 16 K at the STM scanner have been achieved, with the complete cool-down of the system typically taking up to 12 h. These results demonstrate that the constructed CCC-coupled STM is a highly stable instrument capable of highly detailed spectroscopic investigations of materials and surfaces at the atomic scale. (C) 2014 AIP Publishing LLC.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] HIGH-STABILITY SCANNING TUNNELING MICROSCOPE
    VANDEWALLE, GFA
    GERRITSEN, JW
    VANKEMPEN, H
    WYDER, P
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1985, 56 (08): : 1573 - 1576
  • [2] HIGH-STABILITY BIMORPH SCANNING TUNNELING MICROSCOPE
    BLACKFORD, BL
    DAHN, DC
    JERICHO, MH
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1987, 58 (08): : 1343 - 1348
  • [3] APPLICATIONS OF A HIGH-STABILITY SCANNING TUNNELING MICROSCOPE
    VANKEMPEN, H
    VANDEWALLE, GFA
    IBM JOURNAL OF RESEARCH AND DEVELOPMENT, 1986, 30 (05) : 509 - 514
  • [4] Cool-down dynamics of a cryostat with a closed-cycle cryogenic refrigerator
    Zhang, BX
    ADVANCES IN CRYOGENIC ENGINEERING, VOL 43 PTS A AND B, 1998, 43 : 691 - 696
  • [5] COMPACT, HIGH-STABILITY, THIMBLE-SIZE SCANNING TUNNELING MICROSCOPE
    BESENBACHER, F
    LAEGSGAARD, E
    MORTENSEN, K
    NIELSEN, U
    STENSGAARD, I
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (07): : 1035 - 1038
  • [6] HTS scanning SQUID microscope cooled by a closed-cycle refrigerator
    Univ of Maryland at College Park, College Park, United States
    IEEE Trans Appl Supercond, 2 III (3704-3707):
  • [7] HTS scanning SQUID microscope cooled by a closed-cycle refrigerator
    Fleet, EF
    Chatraphorn, S
    Wellstood, FC
    Green, SM
    Knauss, LA
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) : 3704 - 3707
  • [8] A closed-cycle 1 K refrigeration cryostat
    Wang, Chao
    Lichtenwalter, Ben
    Friebel, Aaron
    Tang, Hong X.
    Cryogenics, 2014, 64 : 5 - 9
  • [9] Closed-cycle optical cryostat as an integrated system
    不详
    MRS BULLETIN, 2010, 35 (12) : 951 - 951
  • [10] A closed-cycle 1 K refrigeration cryostat
    Wang, Chao
    Ben Lichtenwalter
    Friebel, Aaron
    Tang, Hong X.
    CRYOGENICS, 2014, 64 : 5 - 9