Risk of Impingement and Third-body Abrasion With 28-mm Metal-on-metal Bearings

被引:13
|
作者
Clarke, Ian C. [1 ,2 ]
Lazennec, Jean-Yves [3 ]
Brusson, Adrien [3 ]
Savisaar, Christina [4 ]
Bowsher, John G. [4 ]
Burgett, Michelle [2 ]
Donaldson, Thomas K. [5 ]
机构
[1] Loma Linda Univ, Dept Orthopaed, Loma Linda, CA 92350 USA
[2] DARF Ctr, Colton, CA 92324 USA
[3] UPMC Univ, Dept Orthopaed, La Pitie Hosp, Paris, France
[4] US FDA, Silver Spring, MD USA
[5] Empire Orthoped, Colton, CA USA
关键词
TOTAL HIP-REPLACEMENT; CERAMIC FEMORAL HEADS; LONG-TERM WEAR; ACETABULAR COMPONENT; ARTICULAR SURFACE; EARLY OSTEOLYSIS; ION LEVELS; ALUMINA; DISLOCATION; DEBRIS;
D O I
10.1007/s11999-013-3399-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Concerns have been raised about the sequelae of metal-on-metal (MoM) bearings in total hip arthroplasty (THA). However, retrieval studies, which offer the best insight into the clinically relevant mechanisms of MoM wear, have followed predictable trends to date such as indicting cobalt-chromium (CoCr) metallurgy, cup design, high conformity between the head and cup, "steep cups," "microseparation," and "edge wear." We wished to evaluate a set of retrieved 28-mm MoM THA for signs of (1) cup-to-stem impingement; (2) normal wear pattern and concomitant stripe damage on femoral heads that would signify adverse wear mechanics; and (3) well-defined evidence of third-body scratches on bearings that would indicate large abrasive particles had circulated the joint space. Ten 28-mm MOM retrievals were selected on the basis that femoral stems were included. Revision surgeries at 3 to 8 years were for pain, osteolysis, and cup loosening. CoCr stems and the MoM bearings were produced by one vendor and Ti6Al4V stems by a second vendor. All but two cases had been fixed with bone cement. We looked for patterns of normal wear and impingement signs on femoral necks and cup rims. We looked for adverse wear defined as stripe damage that was visually apparent on each bearing. Wear patterns were examined microscopically to determine the nature of abrasions and signs of metal transfer. Graphical models recreated femoral neck and cup designs to precisely correlate impingement sites on femoral necks to cup positions and head stripe patterns. The evidence revealed that all CoCr cup liners had impinged on either anterior or posterior facets of femoral necks. Liner impingement at the most proximal neck notch occurred with the head well located and impingement at the distal notch occurred with the head rotated 5 mm out of the cup. The hip gained 20A degrees motion by such a subluxation maneuver with this THA design. All heads had stripe wear, the basal and polar stripes coinciding with cup impingement sites. Analysis of stripe damage revealed 40 to 100-mu m wide scratches created by large particles ploughing across bearing surfaces. The association of stripe wear with evidence of neck notching implicated impingement as the root cause, the outcome being the aggressive third-body wear. We found consistent evidence of impingement, abnormal stripe damage, and evidence of third-body abrasive wear in a small sample of one type of 28-mm MoM design. Impingement models demonstrated that 28-mm heads could lever 20A degrees out of the liners. Although other studies continue to show good success with 28-mm MoM bearings, their use has been discontinued at La Pitie Hospital. Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
引用
收藏
页码:497 / 508
页数:12
相关论文
共 36 条
  • [1] Adverse loading effects on tribocorrosive degradation of 28 mm metal-on-metal hip replacement bearings *
    Beadling, Andrew R.
    Bryant, Michael G.
    Dowson, Duncan
    Neville, Anne
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2021, 235 (12) : 2664 - 2674
  • [2] Metal-on-metal hip resurfacing compared with 28-mm diameter metal-on-metal total hip replacement A RANDOMISED STUDY WITH SIX TO NINE YEARS' FOLLOW-UP
    Vendittoli, P-A.
    Riviere, C.
    Roy, A. G.
    Barry, J.
    Lusignan, D.
    Lavigne, M.
    [J]. BONE & JOINT JOURNAL, 2013, 95B (11): : 1464 - 1473
  • [3] A Minimum of 15-Year Results of Cementless Total Hip Arthroplasty Using a 28-mm Metal-On-Metal Articulation
    Park, Chan-Woo
    Kim, Joo-Hwan
    Lim, Seung-Jae
    Moon, Young-Wan
    Park, Youn-Soo
    [J]. JOURNAL OF ARTHROPLASTY, 2019, 34 (07): : 1387 - 1394
  • [4] Cemented metal-on-metal total hip replacement with 28-mm head: prospective, long-term, clinical, radiological and metal ions data
    Malek I.A.
    Rao S.P.P.S.
    Rath N.K.
    Mallya U.N.
    [J]. European Journal of Orthopaedic Surgery & Traumatology, 2015, 25 (4) : 749 - 755
  • [5] Hip Resurfacing Compared with 28-mm Metal-on-Metal Total Hip Replacement A Randomized Study with 15 Years of Follow-up*
    Vendittoli, Pascal-Andre
    Shahin, Maged
    Riviere, Charles
    Roy, Alain Guy
    Barry, Janie
    Lavigne, Martin
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2020, 102 (21S): : 80 - 90
  • [6] Particle release in metal-on-metal bearings. A risk analysis
    Heisel, C.
    Thomsen, M.
    Jakubowitz, E.
    Kretzer, J. P.
    [J]. ORTHOPADE, 2008, 37 (07): : 644 - 649
  • [7] Minimum ten-year results of a 28-mm metal-on-metal bearing in cementless total hip arthroplasty in patients fifty years of age and younger
    Innmann, Moritz M.
    Gotterbarm, Tobias
    Kretzer, Jan Philippe
    Merle, Christian
    Ewerbeck, Volker
    Weiss, Stefan
    Aldinger, Peter R.
    Streit, Marcus R.
    [J]. INTERNATIONAL ORTHOPAEDICS, 2014, 38 (05) : 929 - 934
  • [8] A comparison of friction in 28 mm conventional and 55 mm resurfacing metal-on-metal hip replacements
    Brockett, C. L.
    Williams, S.
    Jin, Z. M.
    Isaac, G.
    Fisher, J.
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART J-JOURNAL OF ENGINEERING TRIBOLOGY, 2007, 221 (J3) : 391 - 398
  • [9] Minimum 15-year results of metasul 28-mm metal-on-metal total hip arthroplasty in patients younger than 50 years of age
    Jin, Sheng-Yu
    Jin, Jing-Yao
    Kang, Joon-Kyoo
    Yoon, Taek-Rim
    Park, Kyung-Soon
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2021, 16 (01)
  • [10] Minimum ten-year results of a 28-mm metal-on-metal bearing in cementless total hip arthroplasty in patients fifty years of age and younger
    Moritz M. Innmann
    Tobias Gotterbarm
    Jan Philippe Kretzer
    Christian Merle
    Volker Ewerbeck
    Stefan Weiss
    Peter R. Aldinger
    Marcus R. Streit
    [J]. International Orthopaedics, 2014, 38 : 929 - 934